This talk will familiarize MRI engineers and clinicians with some of the principal limitations on MRI scanner hardware performance, along with methods for characterizing and mitigating these imperfections. We will also discuss issues arising from the interaction of the scanner's static and radiofrequency fields with the human body.
MRI scanner components have interdependent requirements and must work in harmony. It is therefore becoming rare to design magnet, gradient and RF subsystems separately and expect the system to work well when bolted together. The interaction of the gradient coil with the magnet is an important example, on which there is little in the literature. Also, it is important to match the gradient linearity, RF coverage region and homogeneous magnet region to prevent detection of unencoded spins from folding into the FOV will be discussed.
The system imperfections discussed in this talk are briefly outlined below. We divide the imperfections into two categories: (1) those arising from limitations on the scanner hardware itself, and (2) issues arising from the interaction of scanner electromagnetic fields and the human body.
Interaction of scanner fields with the body
In addition to the above imperfections of the MRI scanner hardware itself, the introduction of the patient’s body to the bore causes important changes in both the static and radiofrequency fields generated by the scanner. Because these patient-specific perturbations lead to a large variety of image artifacts, it is important to understand how to measure these imperfections and what approaches are available to remedy them.
B0 INHOMOGENEITY
B1 INHOMOGENEITY
1. Jovicich J, Czanner S, Greve D, Haley E, Van Der Kouwe A, Gollub R, Kennedy D, Schmitt F, Brown G, MacFall J, Fischl B, Dale A. Reliability in multi-site structural MRI studies: Effects of gradient non-linearity correction on phantom and human data. Neuroimage. 2006;30(2):436-443. PMID: 16300968.
2. Kornak J, Marinelli L, Hylton N. Gradient Nonlinearity Correction to Improve Apparent Diffusion Coefficient Accuracy and Standardization in the American College of Radiology Imaging Network 6698 Breast Cancer Trial. J Magn Reson Imag. 2016;42(4):908-919.
3. Hedeen RA, Edelstein WA. Characterization and Prediction of Gradient Acoustic. Magn Reson Med. 1997;37(1):7-10.
4. Le Bihan D, Poupon C, Amadon A, Lethimonnier F. Artifacts and pitfalls in diffusion MRI. J Magn Reson Imaging. 2006. PMID: 16897692.
5. Setsompop K, Kimmlingen R, Eberlein E, Witzel T, Cohen-Adad J, McNab JA, Keil B, Tisdall MD, Hoecht P, Dietz P, Cauley SF, Tountcheva V, Matschl V, Lenz VH, Heberlein K, Potthast A, Thein H, Van Horn J, Toga A, Schmitt F, Lehne D, Rosen BR, Wedeen V, Wald LL. Pushing the limits of in vivo diffusion MRI for the Human Connectome Project. Neuroimage. 2013;80:220-233. PMID: 23707579.
6. Welz A, Zaitsev M, Jia F, Liu Z, Korvink J, Schmidt H, Lehr H, Post H, Dewney A, Hennig J. Development of a non-shielded PatLoc gradient insert for human head imaging. Proc 17th Sci Meet Int Soc Magn Reson Med. 2009;Honolulu(2007):3073.
7. De Zanche N, Barmet C, Nordmeyer-Massner JA, Pruessmann KP. NMR Probes for measuring magnetic fields and field dynamics in MR systems. Magn Reson Med. 2008;60(1):176-186. PMID: 18581363.
8. Barmet C, De Zanche N, Pruessmann KP. Spatiotemporal magnetic field monitoring for MR. Magn Reson Med. 2008;60(1):187-197. PMID: 18581361.
9. Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP. Gradient system characterization by impulse response measurements with a dynamic field camera. Magn Reson Med. 2013;69(2):583-593. PMID: 22499483.
10. Stockmann JP, Wald LL. In vivo B<inf>0</inf> field shimming methods for MRI at 7T. Neuroimage. 2017.
11. Rosen BR, Wedeen VJ, Brady TJ. Selective saturation in NMR. J Comp Assist Tomog. 1984;8(5):813-818.
12. De Graaf RA, Wiley InterScience (Online service). In Vivo NMR Spectroscopy : Principles and Techniques. John Wiley & Sons; 2007.
13. Hargreaves BA, Cunningham CH, Nishimura DG, Conolly SM. Variable-rate selective excitation for rapid MRI sequences. Magn Reson Med. 2004;52(3):590-597. PMID: 15334579.
14. Schneider R, Ritter D, Haueisen J, Pfeuffer J. B0-informed variable density trajectory design for enhanced correction of off-resonance effects in parallel transmission. Magn Reson Med. 2014;71(4):1381-1393. PMID: 23716347.
15. Davids M, Schad LR, Wald LL, Guerin B. Fast three-dimensional inner volume excitations using parallel transmission and optimized k-space trajectories. Magn Reson Med. 2016;76(4):1170-1182. PMID: 26527590.
16. Gore JC, Barry RL, Vannesjo SJ, Smith SA, By S. Spinal cord MRI at 7T. Neuroimage. 2017;168:437-451.
17. Stockmann JP, Wald LL. In vivo B0 field shimming methods for MRI at 7T. Neuroimage. 2017;(June):71-87. PMID: 28602943.
18. Hetherington HP, Chu WJ, Gonen O, Pan JW. Robust fully automated shimming of the human brain for high-field 1H spectroscopic imaging. Magn Reson Med. 2006;56(1):26-33. PMID: 16767750.
19. Jenkinson M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. Fsl. Neuroimage. 2012;62(2):782-790. PMID: 21979382.
20. Karsa A, Shmueli K. SEGUE: a Speedy rEgion-Growing algorithm for Unwrapping Estimated phase. IEEE Transactions on Medical Imaging. 2018.
21. Robinson S, Schödl H, Trattnig S. A method for unwrapping highly wrapped multi-echo phase images at very high field: UMPIRE. Magn Reson Med. 2014.
22. Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B(o) field variations. Magn Reson Med. 1995;34(1):65-73. PMID: 7674900.
23. Andersson JLR, Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo echo-planar images: Application to diffusion tensor imaging. Neuroimage. 2003;20(2):870-888. PMID: 14568458.
24. Griswold M a., Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202-1210. 25. Wald LL. The future of acquisition speed, coverage, sensitivity, and resolution. Neuroimage. 2012;62(2):1221-1229. PMID: 22421052.
26. Kim T, Lee Y, Zhao T, Hetherington HP, Pan JW. Gradient-echo EPI using a high-degree shim insert coil at 7 T: Implications for BOLD fMRI. Magn Reson Med. December 2016. PMID: 27910126.
27. Juchem C, Nixon TW, McIntyre S, Boer VO, Rothman DL, De Graaf RA. Dynamic multi-coil shimming of the human brain at 7 T. J Magn Reson. 2011;212(2):280-288. PMID: 21824794.
28. Vannesjo SJ, Miller KL, Clare S, Tracey I. Spatiotemporal analysis of breathing-induced fi elds in the cervical spinal cord at 7T. Proc Int Soc Magn Reson Med. 2017:0752.
29. Duerst Y, Wilm BJ, Dietrich BE, Vannesjo SJ, Barmet C, Schmid T, Brunner DO, Pruessmann KP. Real-time feedback for spatiotemporal field stabilization in MR systems. Magn Reson Med. 2015;73(2):884-893. PMID: 24634192.
30. Vaidya M V., Collins CM, Sodickson DK, Brown R, Wiggins GC, Lattanzi R. Dependence of B1- and B1+ field patterns of surface coils on the electrical properties of the sample and the MR operating frequency. Concepts Magn Reson Part B Magn Reson Eng. 2016.
31. Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63(5):1315-1322. PMID: 20432302.
32. Khalighi MM, Rutt BK, Kerr AB. RF pulse optimization for Bloch-Siegert B+1 mapping. Magn Reson Med. 2012;68(3):857-862. PMID: 22144397.
33. Wrede KH, Johst S, Dammann P, Umutlu L, Schlamann MU, Sandalcioglu IE, Sure U, Ladd ME, Maderwald S. Caudal Image Contrast Inversion in MPRAGE at 7 Tesla. Problem and Solution. Acad Radiol. 2012;19(2):172-178. PMID: 22104286.
34. Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of the effects of high-permittivity pads in 7 Tesla MRI of the brain. Magn Reson Med. 2012. PMID: 21826732.
35. Katscher U, Boernert P. Parallel RF transmission in MRI. NMR Biomed. 2006;19(3):393-400. PMID: 16705630.
36. Cloos MA, Knoll F, Zhao T, Block KT, Bruno M, Wiggins GC, Sodickson DK. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016.
37. Gras V, Vignaud A, Amadon A, Le Bihan D, Boulant N. Universal pulses: A new concept for calibration-free parallel transmission. Magn Reson Med. 2017;77(2):635-643. PMID: 26888654.