Synopsis
MRI can be used to
non-invasively encode and quantify different biophysical aspects of tissue
including flow and motion. In this presentation, we will review methods that generate
motion related image contrast, through manipulation of longitudinal
magnetization prior to excitation, such as Time-of-flight MR Angiography, Arterial Spin Labeling, Myocardial Tagging; as well as through manipulation of
transverse magnetization between excitation and acquisition, such as Phase Contrast MR Angiography, Diffusion Imaging, and MR Elastography.
Syllabus
MRI can be used to
non-invasively encode and quantify different biophysical aspects of tissue
including flow and motion.
A typical pulse sequence
involves excitation, where longitudinal magnetization is converted to
observable transverse magnetization using a combination of RF and gradient
pulses; and data acquisition, where transverse magnetization is spatially
encoded using additional imaging gradients for signal detection and image
reconstruction. Gradients, used in conjunction with RF pulses to nutate
magnetization, can be used to modulate longitudinal and/or transverse
magnetization, by changing the local Larmor frequency and the phase of spins.
Methods for generating motion
related image contrast, that can be categorized as follows, will be reviewed in
this presentation:
A) Methods
that modify Mz prior to excitation: These methods use a
combination of RF and gradient pulses to modify the spatial distribution of
longitudinal magnetization and observe the effects of the transport of these
‘tagged’ spins from one location to another. Since they are based on modulation
of Mz, the time scale for these methods is limited by T1.
- Time-of-flight
MR angiography: maximizes vessel conspicuity by manipulating
imaging parameters in a conventional imaging sequence. When a fast gradient
echo sequence is used with a large flip angle, the steady state magnetization
in static spins in the imaging volume is low. However, blood flow continually
transports relaxed blood spins into the imaging volume, therefore blood flowing
in vessels tends to have larger longitudinal magnetization (and MR signal) than
surrounding static tissue. Simple image processing methods such as Maximum
Intensity Projection then serve to exclude non-flowing spins from the final
image.
- Arterial
spin labeling: differentiates inflowing arterial water from
water in the tissue of interest by labeling the arterial water magnetization using
spatially selective RF inversion (or saturation) pulses. A delay is inserted
into the sequence to allow for labeled blood to reach the tissue of interest.
Arrival of labeled blood in the tissue of interest decreases the tissue MRI
signal in proportion to blood flow and other measurable parameters. Subtraction
of labeled and control images (acquired without labeling) provides the perfusion
sensitive signal, which with appropriate modeling assumptions (and additional
data such as the T1 of tissue), can provide quantitative estimates of regional
perfusion.
- Myocardial
tagging: relies on temporary magnetic fiducial
markers, or tags, which are generated by modulating the underlying image
intensity using a specific presaturation (i.e. exciting multiple planes of
magnetization, which for saturation tagging are then dephased) such that they make
no significant contribution to subsequently acquired images (leaving null
regions at tagged locations). After a certain time (i.e. a different phase of
the cardiac cycle) another image is acquired where the tags are still visible.
Deformation of the tags is analyzed to obtain a model of the underlying tissue
motion. DENSE and HARP encode a uniform pattern of phase modulation into the
tissue at a chosen time, and detect the deformation of that pattern at a later
time to estimate local displacements.
B) Methods
that manipulate Mxy (after excitation before data acquisition): These methods use
only gradient pulses, i.e. motion sensitizing gradients. Contrast is based on
phase shifts from motion along the gradient. A basic waveform is a bipolar
gradient pulse, the first half of which generates a linear phase modulation in
the direction of the gradient vector, and the second half reverses this
modulation. If motion occurs during the pulse, the phase modulation is incompletely
rewound, leaving a residual phase shift, proportional to the moving spins velocity,
the strength of the applied gradient, and the square of the length of time it
moves within that gradient. Since they use Mxy, the time scale for these
methods is limited by T2 decay.
- Phase
contrast MR angiography: can be used to quantify velocity of moving
fluids such as blood and CSF. Bipolar gradients are typically added to a fast
gradient recalled imaging sequence. A reference dataset corrects for other
phase contributions that do not originate from motion. After subtraction of the
reference phase from the phase image obtained with bipolar gradients, the phase
of the image is directly proportional to the component of velocity that is
along the direction of the applied gradient. Duration and magnitude of the
bipolar gradient pair determines the velocity encoding (VENC). VENC is
inversely related to the size of these gradients. For an optimal velocity to
noise ratio, VENC should be set such that the highest velocity likely to be
encountered within the vessel of interest corresponds to a 180° phase shift.
Velocity aliasing occurs when VENC is set too low. Flow can be calculated by
integrating the area-velocity product over an area of interest, e.g. a vessel.
- Diffusion
imaging: provides information on tissue microstructure,
similarly, using bipolar gradients to produce motion related phase shifts in
individual spins. Since motion of spins due to diffusion is not coherent, there
is a distribution of velocities (directions and/or magnitude), thus a distribution
of phases (there is no net phase shift). Phase dispersion across spins means a less
coherent signal with an attenuated magnitude. Diffusion weighting gradients
typically need to be a few orders of magnitude larger than flow weighting
gradients in order to generate measurable signal attenuation. The attenuation is
related to the typical distances travelled by the diffusing spins (thus the
diffusion coefficient D), the time between the diffusion weighting gradients,
their amplitude and duration. Relevant experimental factors (i.e. gradient
amplitudes and timings) are grouped together into the b-value (larger b-values induce
more dephasing, hence more signal loss). Quantitative estimates of the apparent
diffusivity along the direction of the applied motion-encoding gradient can be
obtained from the ratio of the diffusion weighted intensities measured at
different b-values.
- MR elastography: involves
estimation of tissue stiffness through introduction of a known oscillation into
the tissue of interest (i.e. acoustic shear waves applied with a mechanical
driver) to generate standing waves in the tissue. The standing waves are imaged
using the principles of phase contrast MRA, but because the displacement is
periodic, periodic motion encoding gradients tuned to the vibration frequency
can selectively sensitize the sequence to the applied vibrations while
filtering out other sources of motion. The train of bipolar gradients is
externally triggered to synchronized them with the external vibration, such
that the shear waves generated in the tissue are encoded into the phase of the
signal. Wave propagation, the local wavelength, and tissue deformation are then
analyzed to estimate viscoelastic properties such as shear modulus.
Acknowledgements
No acknowledgement found.References
Bernstein, King, Zhou.
Handbook of MRI Pulse Sequences
Liang, Lauterbur.
Principles of Magnetic Resonance Imaging
Nishimura. Principles
of Magnetic Resonance Imaging
Haacke, Brown,
Thompson, Venkatesan. Magnetic Resonance Imaging Physical Principles and
Sequence Design