Christoph Juchem1, Michael Mullen2, Chathura Kumaragamage3, Lance DelaBarre2, Gregor Adriany2, Peter B Brown3, Scott McIntyre3, Terence W Nixon3, Michael Garwood2, and Robin A de Graaf3
1Departments of Biomedical Engineering and Radiology, Columbia University, New York, NY, United States, 2Department of Radiology, Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, MN, United States, 3Department of Radiology and Biomedical Imaging, Magnetic Resonance Resesarch Center (MRRC), Yale University, New Haven, CT, United States
Synopsis
To
date, spatial encoding for human MRI is based on linear X, Y and Z field
gradients generated by dedicated X, Y and Z gradient coils. We recently
introduced the Dynamic Multi-Coil Technique (DYNAMITE) for B0 magnetic field
modeling and demonstrated DYNAMITE MRI in a miniaturized setup. In this study
we report the first realization of multi-slice DYNAMITE MRI of the in vivo human
brain in which all gradient fields are purely DYNAMITE-based. The obtained
image fidelity is comparable to MRI with conventional gradient coils, paving
the way for full-fledged human DYNAMITE MRI systems.
INTRODUCTION
Spatial
encoding for magnetic resonance imaging (MRI) is traditionally based on
magnetic field gradients generated by dedicated X, Y and Z wire patterns. The
DYNAmic Multi-coIl TEchnique (DYNAMITE)1,2 provides
a new and generalized method for the synthesis of advanced magnetic field
distributions with generic (i.e. non-orthogonal) basis fields from a set of
localized coils. DYNAMITE MRI in which all encoding fields are purely
multi-coil-generated has been shown in miniaturized setups for radial MRI3, Cartesian
MRI and EPI4, and non-linear
encoding5 as
well as for the concurrent generation of both MRI and B0 shim fields.4 Here
we report the first realization of DYNAMITE MRI of the in vivo human brainMETHODS
All experiments were performed at Yale
University on a 4T Magnex 94 cm magnet interfaced to a Bruker spectrometer.
DYNAMITE magnetic fields were generated with a 28 coil matrix distributed as 4
rows of 7 coils on a cylindrical former (Ø 344 mm, spanning 359 mm along the z
direction). Each DC coil was composed of two nested rectangular elements of 100
windings each with center dimensions 144 x 80 mm and 118 x 54 mm run in series
and interfaced to a ±5 A constant current amplifier (Resonance Research
Inc, Billerica, MA). DYNAMITE field modeling and synthesis of the event files
for rotating gradient shapes (10 events, figure 1), gradient-echo MRI (18432
events, figure 2) and spin-echo MRI (22528 events) was achieved with customized
B0DETOX software.6 The amplitudes and timings of the 28
channels were set by a multi-channel gradient waveform controller, developed
in-house.7 RF pulse transmission and reception was
achieved with a 16-channel stripline RF coil.8 The calibration of the multi-coil B0
field modeling system and basic functionality testing was done in a head-sized
oil phantom (rounded in-plane: 160 x 200 mm, height 20 cm). DYNAMITE MR images
were acquired in 5 healthy volunteers with gradient-echo (TR/TE = 1000/14 ms)
and spin-echo (TR = 2500/50 ms) sequences as 128 x 128 matrices over 256 x 256
mm (16 slices over 3 mm) and compared to conventional MRI at identical
parameters employing the scanner's built-in XYZ gradient coils.RESULTS
Quality assurance checks of the established multi-coil hardware (figure
1A) included the synthesis of well-defined, dynamically cycled field patterns
such as a 100 Hz/cm linear field gradient that was rotated by 45-degree
increments between slices (figure 1B). High accuracy of both shape and amplitude
of the slice-specific gradient fields confirmed the general design and functionality
of the derived multi-coil hardware and computational infrastructure (figure 1C).
Our goal in this proof-of-principle research was to provide a physical replica
of standard MRI sequences with DYNAMITE and to demonstrate full MRI capability
in the human head at image quality comparable to conventional MR systems. To
this end, all gradient patterns for slice-selection and in-plane phase and
frequency encoding of conventional gradient-echo (figure 2A) and a spin-echo
(not shown) sequences were replaced piece-by-piece with DYNAMITE-generated
fields. The DYNAMITE gradient shapes were then played out by a series of
well-timed TTL trigger signals to resemble the full MRI sequence (figure 2B). The
resultant multi-slice spin-echo DYNAMITE images acquired from the human brain resembled
those acquired with conventional gradient hardware throughout the FOV (figure 3).
Similarly, high-level DYNAMITE MRI was achieved with both spin-echo (figure 4)
and gradient-echo (figure 5) encoding from multiple subjects. Minor differences
between conventional and DYNAMITE MRI can be explained by imperfectly
calibrated MC magnetic field maps and will be addressed in the future by
performing a higher-precision calibration over an extended phantom. Notably,
the established sequence implementation allowed the replacement of individual
sequence components (such as slice selection or in-plane encoding) and equivalent
MRI results were obtained for all these scenarios (data not shown). The correspondence
of conventional and DYNAMITE-encoded MRI strongly implies that the hundreds of
gradient fields dynamically generated by DYNAMITE indeed closely resembled
those provided by conventional gradient coils.
CONCLUSIONS
The
first successful realization of DYNAMITE MRI of the in vivo human brain at image
fidelity comparable to MRI using conventional gradient coils is expected to pave
the way for full-fledged human DYNAMITE MRI systems.Acknowledgements
This
research was supported by NIH grants R24-MH105998, R01-EB014861, T32-EB008389 and U01-EB025153.References
1. Juchem, C., T.W. Nixon, S. McIntyre,
D.L. Rothman, and R.A. de Graaf, Magnetic
field modeling with a set of individual localized coils. J Magn Reson
204:281-9 (2010).
2. Juchem, C., D. Green, and R.A. de
Graaf, Multi-coil magnetic field
modeling. J Magn Reson 236:95-104 (2013).
3. Juchem, C., O.M. Nahhass, T.W.
Nixon, and R.A. de Graaf, Multi-slice MRI
with the dynamic multi-coil technique. NMR Biomed 28:1526-34 (2015).
4. Rudrapatna, S.U., F. Fluerenbrock,
W.T. Nixon, R.A. de Graaf, and C. Juchem, Combined
Imaging and Shimming with the Dynamic Multi-Coil Technique. Magn Reson Med
(2018), epub ahead of print.
5. Juchem, C., T.W. Nixon, and R.A. de
Graaf. Multi-Coil Imaging with Algebraic
Reconstruction. Proc. ISMRM
2012, p.2545.
6. Juchem, C. B0DETOX - B0 Detoxification Software for Magnetic Field Shimming.
Columbia TechVenture (CTV), License CU17326 (2017). Available from: innovation.columbia.edu/technologies/cu17326_b0detox.
7. Nixon, T.W., S. McIntyre, and R.A.
de Graaf. The Design and Implementation
of a 64 Channel Arbitrary Gradient Waveform Controller. Proc. ISMRM 2017, p.0969.
8. Adriany, G.,
P.F. Van de Moortele, F. Wiesinger, S. Moeller, J.P. Strupp, P. Andersen, C.
Snyder, X. Zhang, W. Chen, K.P. Pruessmann, P. Boesiger, T. Vaughan, and K.
Ugurbil, Transmit and receive transmission
line arrays for 7 Tesla parallel imaging. Magn Reson Med 53:434-45 (2005).