Susceptibility, Flow, Chemical Shift, Motion Artifacts & Imaging
Frederik Laun1

1University Hospital Erlangen, Germany

Synopsis

Following this lecture, the audience will understand the effect of magnetic susceptibility, flow, chemical shift, and motion on MRI. They will be able to use this knowledge to identify and minimize related artifacts or, alternatively, to measure the underlying effects.

Target Audience

Researchers who are interested in measuring the effects related to susceptibility, flow, chemical shift, motion or who want to minimize artifacts caused by these effects.

Outcome / Objectives

Following this lecture, the audience will understand the effect of magnetic susceptibility, flow, chemical shift, and motion on MRI. They will be able to use this knowledge to identify and minimize related artifacts or, alternatively, to measure the underlying effects.

Methods

The transversal magnetization of water protons in the main magnetic field $$$B_0=|(0,0,B_0)^T| $$$ rotates with the angular frequency $$$\omega_0=-\gamma B_0 $$$ around the z-axis. Here, $$$ \gamma $$$ is the gyromagnetic ratio. In the following, the reference frame rotating with $$$ \omega_0$$$ is used.

In this lecture, the following effects are considered:

Chemical shift

Protons residing in a different chemical environment than water have a slightly different rotation frequency, i.e. $$$\omega \rightarrow \gamma B_0 (1+\sigma) $$$ with the chemical shift $$$ \sigma$$$. E.g. for the dominant chemical group in fat, the aliphatic CH2 group, a chemical shift of 3.5 ppm arises with respect to water (1).

Magnetic susceptibility

Human tissue becomes slightly magnetized in a magnetic field according to the equation $$$\mathbf{M}=\chi \mathbf{H}$$$ with the magnetic susceptibility $$$\chi$$$ (2). $$$\chi$$$ is a rather small unitless quantity (roughly 1 ppm) and is larger or smaller than zero for paramagnetic or diamagnetic tissue, respectively. In general, the magnetic susceptibility and the magnetization dependent on position, i.e. they may be noted as $$$\mathbf{M}(\mathbf{r})$$$ and $$$\chi(\mathbf{r})$$$. A point source, i.e. an infinitesimal volume $$$dV$$$ at position $$$\mathbf{r}_1$$$ with magnetization $$$\mathbf{M}(\mathbf{r}_1)$$$, generates the additional magnetic field $$$\Delta B_{\textrm{point}}(\mathbf{r}-\mathbf{r}_1)=B_0(4\pi)^{-1}(3 \cos^2(\theta)-1)|\mathbf{r}-\mathbf{r}_1|^{-3}dV$$$. Here, $$$\theta$$$ is the angle between z-axis and $$$\mathbf{r}-\mathbf{r}_1$$$. These additional fields add up linearly if more than one point source is present. Consequently the complete field perturbation is $$$\Delta B(\mathbf{r})=\Delta B_{\textrm{point}}(\mathbf{r})*\chi(\mathbf{r})$$$, where “$$$*$$$” denotes a convolution. If all tissues had the same magnetic susceptibility, then $$$\Delta B(\mathbf{r})$$$ would have almost no spatial dependency. But if tissue interfaces – in particular tissue-air interfaces – are present, a stronger spatial dependency of $$$\Delta B(\mathbf{r})$$$ arises resulting in $$$\omega(\mathbf{r})=\gamma \Delta B(\mathbf{r}) $$$.

Flow and motion

Some spins in the human body move on a macroscopic scale and their position becomes dependent on time $$$t$$$, i.e. $$$\mathbf{r}\rightarrow \mathbf{r}(t)$$$ . For example, blood often flows approximately coherently with a velocity $$$\mathbf{v}$$$ so that $$$\mathbf{r}(t)=\mathbf{r}(0)+\mathbf{v}t$$$. If a magnetic field gradient $$$\mathbf{G}(t)$$$ is applied, then the angular frequency becomes $$$\omega(\mathbf{r})=\gamma \mathbf{G}(t) \cdot \mathbf{r}(t)=\gamma \mathbf{G}(t) \cdot (\mathbf{r}(0)+\mathbf{v}t)$$$.

Results and Discussion

Chemical shift

(a) Single line sequences. Applying a readout gradient $$$G_{\textrm{read}}$$$ along x-direction, the angular frequency of water protons is $$$\omega_{water}=\gamma G_{\textrm{read}}x$$$. The angular frequency of chemically shifted protons is $$$\omega_{CS}=\gamma G_{\textrm{read}}x+\gamma B_0 \sigma$$$. Detecting the signal, one does not know its origin, one just sees a signal component at angular frequency $$$\omega_{\textrm{measured}}$$$. In image reconstruction, the following positions are attributed to this angular frequency. For water: $$$x_{\textrm{water}}=\gamma^{-1}G_{\textrm{read}}^{-1}\omega_{\textrm{measured}}=x$$$ thus finding the true position. For the chemically shifted group: $$$ x_{\textrm{CS}}=\gamma^{-1}G_\textrm{{read}}^{-1}\omega_{\textrm{measured}}-G_{\textrm{read}}^{-1}B_0\sigma=x+\Delta x_{\textrm{CS}}$$$ with $$$\Delta x_{\textrm{CS}}=-G_{\textrm{read}}^{-1}B_0\sigma$$$ thus finding an incorrect position shifted by $$$\Delta x_{\textrm{CS}}$$$ in read direction.

(b) Echo planar imaging (EPI). In EPI, the shift $$$\Delta x_{\textrm{CS}}$$$ occurs dominantly along phase direction, since $$$G_{\textrm{phase}}$$$ is effectively much smaller than $$$G_{\textrm{read}}$$$. The shift is usually much larger than for single line sequences, which makes necessary a good fat suppression (see e.g. chapter 12 of (3)).

(c) Spectroscopy. For example, using free induction decay signal acquisition, the frequency composition of spins contained in the measured volume can be retrieved, which reveals the presence of molecules other than water (4).

Magnetic susceptibility

(a) Phase images. Due to $$$\Delta B(\mathbf{r})$$$, an additional phase $$$\varphi(\mathbf{r},t)=\gamma \Delta B(\mathbf{r})t$$$ arises, which can be depicted in “phase images”.

(b) T2’ signal decay. If $$$\varphi(\mathbf{r},t)$$$ varies within a voxel, then this phase dispersion results in a reduced signal in this voxel. This effect becomes more pronounced at larger $$$t$$$. Spin echoes can be used to mitigate this effect if particle motion is negligible (static dephasing regime).

(c) Susceptibility-weighted imaging (SWI) (5,6). Images are generated by multiplying magnitude images with a function $$$f_\varphi(\varphi(\mathbf{r},t))$$$, which is, e.g., defined as: 1 for $$$\varphi>0$$$, and $$$\varphi\pi^{-1}$$$ for $$$\varphi<0$$$ using a high-pass filtered map of $$$\varphi(\mathbf{r},t)$$$. Consequently, the image is darkened for negative $$$\varphi(\mathbf{r},t)$$$, i.e. in veins, iron containing tissue, and hemorrhage.

(d) Quantitative susceptibility mapping (QSM). In QSM (7), phase images, i.e. $$$\varphi(\mathbf{r},t)$$$, are measured allowing to compute $$$\Delta B(\mathbf{r})=\gamma^{-1}t^{-1}\varphi(\mathbf{r},t)$$$. Then the equation $$$\Delta B(\mathbf{r})=\Delta B_{point}(\mathbf{r})*\chi(r)$$$ is used to compute $$$\chi(\mathbf{r})$$$. This involves the use of advanced post processing techniques taking care of issues such as phase unwrapping, background field removal, and the actual field-to-susceptibility inversion. QSM can be used, for example, to detect demyelination (8), calcifications (9) and to differentiate them from hemorrhage (10). Since $$$\chi(\mathbf{r})$$$ is a tensor quantity in myelinated tissue, it may be used to perform fiber susceptibility tensor imaging (STI) and fiber tracking (11).

(e) Image distortions. In the same manner as for chemical shifts, $$$\Delta B(\mathbf{r})$$$ can cause image distortions (replace $$$\Delta \chi_{CS}$$$ by $$$\Delta x_{\Delta B}(\mathbf{r})=-G_{read}^{-1}\Delta B(r)$$$).

Flow and motion

(a) Flow compensation. Since $$$\varphi(\mathbf{r},t)=\gamma\int \mathbf{G}(t) \cdot (\mathbf{r}(0)+\mathbf{v}t)dt=\mathbf{k} \cdot \mathbf{r}(0) + \mathbf{k}_1 \cdot \mathbf{v}$$$ with $$$\mathbf{k}=\gamma\int \mathbf{G}(t) dt$$$ and $$$\mathbf{k}_1=\gamma \int \mathbf{G}(t) t dt$$$, adapting the temporal evolution of the applied gradients such that $$$\mathbf{k}_1=0$$$ suppresses the effect of flow on images (see e.g. chapter 5 of (12)). This also minimizes ghosts arising in phase direction originating from blood in major vessels.

(b) Measuring flow velocity. Performing two exams with different $$$\mathbf{k}_1$$$, e.g. by using additional bipolar gradient pulses, allows one to extract $$$\mathbf{v}$$$ from phase maps (see e.g. chapter 11 of (12)). This can be used to measure blood flow in major vessels.

(c) Triggering/gating. Repetitive body motions can be monitored externally (e.g. by ECG, pulse oximeter, breathing belts) or by means of MRI data (self-gating, see e.g. (13)). To suppress the effect of this motion, the signal is only acquired in the desired time intervals (triggering). Alternatively the signal is acquired without interruption, but only data acquired in the desired time intervals is used for image reconstruction (gating).

(d) Breathhold examinations. The signal is acquired in one or multiple breathholds thus minimizing the effect of breathing motion.

(e) Imaging Registration. Several MRI images that are acquired at different time points and that shall be combined can be registered. This is useful, for example, in high-resolution imaging (14) or in diffusion-weighted imaging.

(f) Prospective motion correction. Using navigator echoes, the read-phase-slice system can be adjusted during the examination to correct for rigid motion (see e.g.(15)).

Acknowledgements

No acknowledgement found.

References

1. Brix G, Heiland S, Bellemann ME, Koch T, Lorenz WJ. Mr-Imaging of Fat-Containing Tissues - Valuation of 2 Quantitative Imaging Techniques in Comparison with Localized Proton Spectroscopy. Magn Reson Imaging 1993;11(7):977-991.

2. Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Medical physics 1996;23(6):815-850.

3. Jones DK. Diffusion MRI: Theory, Methods, and Applications: Oxford University Press, New York, USA; 2011.

4. Bottomly PA, Griffiths JR. Handbook of Magnetic Resonance Spectroscopy In Vivo: MRS Theory, Practice and Applications: Wiley; 2016.

5. Haacke EM, Mittal S, Wu Z, Neelavalli J, Cheng YCN. Susceptibility-Weighted Imaging: Technical Aspects and Clinical Applications, Part 1. Am J Neuroradiol 2009;30(1):19-30.

6. Reichenbach KR, Haacke EM. High-resolution BOLD venographic imaging: a window into brain function. Nmr Biomed 2001;14(7-8):453-467.

7. Schweser F, Deistung A, Reichenbach JR. Foundations of MRI phase imaging and processing for Quantitative Susceptibility Mapping (QSM). Z Med Phys 2016;26(1):6-34.

8. Liu CL, Li W, Johnson GA, Wu B. High-field (9.4 T) MRI of brain dysmyelination by quantitative mapping of magnetic susceptibility. Neuroimage 2011;56(3):930-938.

9. Straub S, Laun FB, Emmerich J, Jobke B, Hauswald H, Katayama S, Herfarth K, Schlemmer HP, Ladd ME, Ziener CH, Bonekamp D, Rothke MC. Potential of Quantitative Susceptibility Mapping for Detection of Prostatic Calcifications. J Magn Reson Imaging 2017;45(3):889-898.

10. Ciraci S, Gumus K, Doganay S, Dundar MS, Ozcora GDK, Gorkem SB, Per H, Coskun A. Diagnosis of intracranial calcification and hemorrhage in pediatric patients: Comparison of quantitative susceptibility mapping and phase images of susceptibility-weighted imaging. Diagn Interv Imag 2017;98(10):707-714.

11. Liu CL. Susceptibility Tensor Imaging. Magn Reson Med 2010;63(6):1471-1477.

12. Callaghan PT. Translational Dynamics & Magnetic Resonance: Oxford University Press, New York, USA; 2011.

13. Jiang WW, Ong F, Johnson KM, Nagle SK, Hope TA, Lustig M, Larson PEZ. Motion robust high resolution 3D free-breathing pulmonary MRI using dynamic 3D image self-navigator. Magn Reson Med 2018;79(6):2954-2967.

14. Lusebrink F, Sciarra A, Mattern H, Yakupov R, Speck O. T1-weighted in vivo human whole brain MRI dataset with an ultrahigh isotropic resolution of 250 mu m. Sci Data 2017;4.

15. Mattern H, Sciarra A, Godenschweger F, Stucht D, Lusebrink F, Rose G, Speck O. Prospective motion correction enables highest resolution time-of-flight angiography at 7T. Magn Reson Med 2018;80(1):248-258.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)