Djaudat Idiyatullin1, Wei Zhu1, Yi Zhang1, Xiao-Hong Zhu1, Wei Chen1, and Kâmil Uğurbil1
1Radiology, Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States
Synopsis
T2 weighted fMRI has
been shown to provide higher spatial specificity than commonly used T2* contrast. However, 2D-spin-echo-EPI
method, which is the most commonly employed approach, suffers from a low functional
mapping contrast. Versions of single-shot 3D-GRASE sequence provide
improvements in fMRI contrast, but suffer limitations particularly at ultrahigh
magnetic fields due to shorter T2.
To solve this problem, a new multi-slab version of the 3D-GRASE sequence with
and without the capability of inner-volume selection, abbreviated as ivmsGRASE
and msGRASE, respectively, are proposed. The theoretical and practical
considerations of these methods in comparison with spin-echo-EPI for fMRI
application are presented.
Introduction
Comparatively,
T2
weighted
spin-echo contrast fMRI has been proved to yield higher spatial specificity
than T2* contrast, due to
its selectivity for smaller blood vessels involved in the neuro-vascular
coupling, which provides the functional mapping signals in fMRI1. However,
conventional 2D-spin-echo-EPI (seEPI) suffers from a low BOLD functional contrast-to-noise
ratio (fCNR)2. To overcome this
obstacle, single-shot 3D-GRASE combining fast spin echo acquisition scheme with
EPI readout has been implemented3. The multiple
refocusing pulses employed after a single excitation increase signal-to-noise
ratio (SNR) and fCNR due to the Fourier averaging. By using slab-selective
excitation and refocusing pulses in orthogonal axes an inner-volume can be
imaged4. The inner-volume
selection allows shorter echo train, which alleviates blurring effect on the
phase-encoding direction. This blurring remains a major limitation, especially at
ultrahigh magnetic field due to the significantly shorter T2. To solve this problem, we propose a single-shot multi-slab
3D-GRASE sequences with and without the capability of inner-volume selection,
respectively abbreviated as ivmsGRASE and msGRASE. Below we present the
theoretical and practical considerations of these methods in comparison with seEPI
for fMRI application.Methods
Fig.1
schematically presents the discussed pulse sequences. All three sequences use
the similar seEPI readout scheme. The difference lies only in the way the third
dimension is encoded. EPI uses stacked slices while GRASE uses slabs,
consisting of spin echoes (encoding in plane defined by the readout and phase
encoding) with a second phase encodings in the third dimension. These three
sequences were compared under condition of the same spatial resolution, matrix
size (unless specified for the inner-volume case), and repetition time between
single-shot fMRI image collections. The noise was assumed to be the same for
all sequences and was not considered. Similarity of these sequences allows
simplified SNR comparison based on differences of the three components, such
as:
SNR~“Excitation_efficiency”x“T2_relaxation”x“Acquisition_efficiency”.
These
components are presented in detail in Table1. For discussion below we use a normalized
SNRNmsGRASE and SNRNivmsGRASE calculated
as ratio of SNR of GRASE sequences to SNR of seEPI.
Experiment: The single-shot msGRASE sequences where implemented and tested on an in vivo rat brain on a Varian 16.4T animal
scanner with single loop 2.5cm surface coil. Resting-state fMRI was further
acquired with conventional seEPI and proposed GRASE sequences.
Results
Higher
values of npe2 benefits SNR, however to avoid blurring we must keep npe2<πT2/2τ
and reach a target resolution by increasing the number of slabs. The
inner-volume case uses smaller npe and allows increased npe2. The “Excitation_efficiency”
depends on effective
TR and
is
lowest for ivmsGRASE. Accelerating recovery of longitudinal magnetization due
to blood in-flow effect as well as longer effective
in Carr-Purcell echo train with
decreased echo-time in brain tissue5,6 highly improve SNR for ivmsGRASE. The
results of theoretical predictions for the normalized SNRs are presented in Table2.
The experimental data presented at Fig.2 in
general agreed with theoretical predictions. We found highly shortened magnetization
recovery time in the brain tissue (~0.5s), which is about 4 times lower than T1 of rat brain tissue at 16.4T (~2.3s)7. This expected
in-flow (perfusion) effect arising from the use of a small surface coil allowed
us to increase flip angles considerably, which, as a result, improved SNR,
especially for ivmsGRASE images.Discussion
The
theoretical and practical considerations based on fMRI experiments show that single-shot
multi-slab 3D-GRASE is a good alternative to seEPI for functional studies. The
multi-slab inner-volume capability is even more practical, especially at
ultrahigh field with shortened T2
relaxation. Imaging with surface coils is very beneficial in this case due to
blood in-flow (perfusion) signal enhancement. However, influence of this effect
on BOLD contrast is yet to be determined. One considerable complication in
multi-slab acquisition is the striped artifacts along the slab selection due to
non-ideal excitation profile. However it was shown that uneven profiles could
be compensated almost lossless during image post-processing8. Even without
compensation, these artifacts have low impact on fMRI results. Additionally,
the performance of these sequences could be enhanced by utilizing partial
Fourier9 and multi-band
accelerations10.Conclusion
A novel
multi-slab single-shot 3D-GRASE sequences have been presented and compared with
conventional spin-echo EPI at high field for high resolution fMRI study. In in-vivo application of the presented
technique shows high flexibility and considerable better SNR and T2 contrast compared to
conventional spin-echo EPI.Acknowledgements
This
study was supported by NIH grants R01 MH111447 and MH111413; P41 EB015894; P30 NS07640.References
1. Yacoub E, Shmuel A, Logothetis N, Ugurbil K. Robust
detection of ocular dominance columns in humans using Hahn Spin Echo BOLD
functional MRI at 7 Tesla. NeuroImage 2007;37(4):1161-77.
2. Ugurbil K. What is feasible with imaging human brain
function and connectivity using functional magnetic resonance imaging.
Philosophical Transactions of the Royal Society B: Biological Sciences
2016;371(1705):20150361.
3. Oshio K, Feinberg DA. Single-Shot GRASE Imaging without
Fast Gradients. Magn Reson Med 1992;26(2):355-360.
4. Feinberg DA, Harel N, Ramanna S, Ugurbil K, Yacoub E.
Sub-millimeter single-shot 3D GRASE with inner volume selection for T2 weighted
fMRI applications at 7 Tesla. In: 16th Annual Meeting International Society for
Magnetic Resonance in Medicine. Toronto, ON; 2008. p. 2373.
5. Jensen JH, Chandra R, Yu H. Quantitative model for the
interecho time dependence of the CPMG relaxation rate in iron-rich gray matter.
Magn Reson Med 2001;46(1):159-65.
6. Michaeli S, Garwood M, Zhu X-H, DelaBarre L, Andersen P,
et al. Proton T2 relaxation study of water, N-acetylaspartate, and creatine in
human brain using Hahn and Carr-Purcell spin echoes at 4T and 7T. Magn Reson
Med 2002;47(4):629-633.
7. Pohmann R, Shajan G, Balla DZ. Contrast at high field:
Relaxation times, magnetization transfer and phase in the rat brain at 16.4 T.
Magn Reson Med 2011;66(6):1572-1581.
8. Wu W, Koopmans PJ, Frost R, Miller
KL. Reducing slab boundary
artifacts in three-dimensional multislab diffusion MRI using nonlinear
inversion for slab profile encoding (NPEN). Magn Reson Med
2016;76(4):1183-1195.
9. Feinberg DA, Hale JD, Watts JC, Kaufman L, Mark A.
Halving MR imaging time by conjugation: demonstration at 3.5 kG. Radiology
1986;161(2):527-531.
10. Feinberg DA, Setsompop K. Ultra-Fast MRI of the Human Brain
with Simultaneous Multi-Slice Imaging. Journal of magnetic resonance (San
Diego, Calif : 1997) 2013;229:90-100.
11. Ernst RR, Anderson WA. Application of Fourier transform
spectroscopy to magnetic resonance. Rev Sci Instrum 1966;37:93-102.
12. Ernst RR, Bodenhausen G,
Wokaun A. Principles of Nuclear Magnetic Resonance in One and Two Dimensions.
5-th ed. Oxford: Clarendon Press; 1997. 518 p.