Signal-to-noise ratio (SNR), a limiting factor for functional MRI, can be improved by through the use of higher magnetic fields or through the use of array coils. This work proposes a novel acquisition scheme for BOLD fMRI that has the potential to improve SNR by a factor of 2 or more. This approach uses an oscillating steady state signal that is T2*-weighting and when combined across the period of oscillation produced robust fMRI results in both 2D and 3D imaging experiments.
Balanced steady-state imaging generally uses a constant phase increment to achieve spin-echo-like and, depending on the FA and TR, T2/T1-weighted contrast. 2 The specific frequency locations of the nulls and peaks in the spectral response can be made to vary via a linearly varying phase increments: $$\phi(n) - \phi(n-1) = 2 \pi \frac{n}{ n_c},$$ leading to a quadratic phase sequence with periodicity $$$n_c$$$ TRs. If the phase increment is kept small ($$$n_c > 120$$$), the spin-echo contrast is preserved. 3 However, for larger phase increments, a new oscillating steady state emerges. An example of magnetization progression for a spin with off-resonance 20 Hz, T1 = 1300ms, T2 = 75ms, FA = 10 degrees, TR= 15ms, and $$$n_c =10$$$ is shown in Fig. 1. Off-resonance shifts lead to similar temporal waveforms but shifted in time. This phase progression is equivalent to a spoiling seed of 36 degrees. With spoiling gradients, one gets the average of all isochromats yielding a small net signal, but balanced gradients lead a large, but oscillating signal. Averages of moderate distributions of spins lead to modest reductions in signal strength, equivalent to some T2’- or T2*-weighting. For the parameters in Fig. 1, simulations show that the T2’-weighting is equivalent to 22ms at the beginning of the readout and 34ms at the end of the readout interval.
Experiment data were collected on a GE MR750 scanner, and for the oscillating steady state used FA = 10 degrees, TR= 15ms, $$$n_c = 10$$$, and a fully sampled spiral-out acquisition with diameter 35 samples, though other parameters can yield similar results. Phantom studies are used to demonstrate the principles of the oscillating steady state using balance gradients as well as with spoiling gradients. Human functional MRI studies were performed using a right-handed finger tapping task (20s on/20s off x 4 cycles) analyzed by correlation with a reference waveform. Both single-slice 2D and 12-slice 3D imaging were acquired. The $$$n_c = 10$$$ oscillations are combined by taking the 2-norm of 10 consecutive image time points, yielding a temporal resolution of 150ms ($$$n_c * TR$$$) and 1.8s ($$$n_c * n_z * TR$$$) for the 2D and 3D acquisitions, respectively. RETROICOR 4 with externally acquired respiration and cardiac signals was applied to 40s blocks was used to eliminate physiological noise in the functional data. Standard 2D BOLD imaging using a spiral-in readout was acquired for comparison.
1. Zur, Y., Wood, M. L., & Neuringer, L. J. (1991). Spoiling of transverse magnetization in steady‐state sequences. Magnetic resonance in medicine, 21(2), 251-263.
2. Bernstein, M. A., King, K. F., & Zhou, X. J. (2004). Handbook of MRI pulse sequences. Elsevier.
3. Foxall, D. L. (2002). Frequency‐modulated steady‐state free precession imaging. Magnetic resonance in medicine, 48(3), 502-508.
4. Glover, G. H., Li, T. Q., & Ress, D. (2000). Image‐based method for retrospective correction of physiological motion effects in fMRI: RETROICOR. Magnetic resonance in medicine, 44(1), 162-167.