Iain P Bruce1, Christopher Petty1, and Allen W Song1
1Brain Imaging and Analysis Center, Duke University, Durham, NC, United States
Synopsis
Diffusion-weighted images acquired with multi-shot EPI at
high-resolution are susceptible to inter-shot motion artifacts and geometric
distortions resulting from magnetic-field inhomogeneities. This study presents
an efficient means of inherently accounting for both motion and distortion
artifacts by alternating the phase encoding gradient polarities in odd/even
shots of multi-shot EPI. When acquired in this fashion, the proposed RPG-MUSE
model simultaneously accounts for both shot-to-shot motion induced artifacts
and geometric distortions during the reconstruction of multi-shot diffusion
weighted images. This technique requires no additional scan time, and accounts
for B0 and eddy current induced distortions specific to each diffusion-weighted
volume.
Purpose
Recent advances in multi-shot echo-planar-imaging (ms-EPI)
methodologies have seen increased utility in achieving high-resolution
diffusion MRI (dMRI). Due to the prolonged readout time associated with sizeable
acquisition matrices, however, the spatial distortions resulting from B0 and eddy
current induced magnetic field (B-field) inhomogeneities remain a challenge in
dMRI acquisitions. To address these challenges, this study presents an
efficient and effective method of inherently accounting for both B-field
distortions and shot-to-shot motion induced artifacts common in multi-shot dMRI.
This is achieved by incorporating reversed polarity gradients (RPG)1,2,3
along the phase encoding direction (PE) into a multiplexed sensitivity encoding
(MUSE)4 acquisition. Specifically, this technique, termed RPG-MUSE, applies
opposite PE gradient polarities to odd/even shots of the ms-EPI acquisition scheme
employed in a conventional MUSE scan, thereby allowing B0 and eddy currents induced
distortions to be estimated for each diffusion volume individually, without
increasing the acquisition time.Methods
Whole brain in-vivo multi-shot DWI
data was acquired in a commercial 3T GE MR750 scanner using a 32-channel head
coil. Volumes with b=0 and b=800s/mm2 were acquired with 1.2x1.2x1.2mm3 isotropic voxels
in 80 axial slices with a 30.7cm FOV and a 256x256
acquisition matrix. Four interleaved shots were acquired with PE=Anterior/Posterior
in all odd shots and PE=Posterior/Anterior in all even shots (66%
partial-Fourier along PE), as illustrated in Fig. 1. The interleaved shots were
individually reconstructed with SENSE5 to produce four full-FOV shot
images (two AP/PA pairs) for each slice in each dMRI volume. These shot images
were then used to estimate 1) low frequency motion-induced phase variations
between shots, and 2) B0/eddy current induced distortion maps for each AP/PA
shot pair using FSL’s topup tool2,3. The shot phase images and field
distortion maps were then fed into RPG-MUSE to reconstruct the raw interleaved
ms-EPI data into full FOV slice images while simultaneously accounting for
shot-to-shot motion artifacts and B-field distortions. Finally, a bias-field
correction was applied to volumes generated from shot 1 (PE=AP), shot 2
(PE=PA), and RPG-MUSE using FSL6.Results and Discussion
The SENSE reconstructed axial and
sagittal slice images for shot 1 (PE=AP) in Fig. 2a exhibits stretching in the entire
anterior region of the brain (white ovals) for both b=0 and b=800s/mm2
volumes, while the corresponding slice image for shot 2 (PE=PA) in Fig. 2b conversely
exhibits compression in the same region (yellow ovals). With B-field distortions
estimated from the b=0 and b=800/mm2 volumes individually in Figs.
2a and 2b, RPG-MUSE simultaneously reconstructs all shot images into a single
image while correcting the distortions observed in this region (blue ovals). Derived
from a combination of all four shots, the RPG-MUSE reconstructed images in Fig.
2c have an increased SNR when compared to the SENSE reconstructed shot images
in Figs. 2a-b, and the improved fidelity achieved through inherent shot-to-shot
motion and distortion corrections provides enhanced anatomical detail and
accuracy.Conclusion
In this study, a conventional
ms-EPI sequence used in acquiring dMRI data with MUSE was modified to apply
alternating phase encoding gradient polarities to odd/even shots. When acquired
in this manner, the RPG-MUSE algorithm presented here simultaneously accounts
for shot-to-shot motion artifacts as well as geometric distortions that arise
from B0 and eddy current induced B-field inhomogeneities. With such an
acquisition, the overall scan time is equivalent to that of a conventional MUSE
acquisition, with no need to acquire redundant data in the form of either additional
b=0 volumes with RPG at the beginning of a dMRI scan1,2,3, or an
entire dMRI acquisition repeated with RPG. Additionally, as the effects of eddy
currents on B-field inhomogeneities are dependent on the direction of the
diffusion gradients applied, this technique is able to more accurately estimate
and account for the distortion characteristics associated with each diffusion
volume than through distortion characteristics estimated from b=0
volumes only.Acknowledgements
This work was supported in part by NIH grants R01-NS-075017, R01-
NS-074045 and R24-106048, and a grant from GE Healthcare.References
1. Bowtell R, et al. Correction of geometric distortion in echo planar images. Proc. Intl Soc. Magn. Reson. Med. 1994, p 411.
2. Andersson
JLR, Skare S, Ashburner J. How to correct susceptibility distortions in
spin-echo echo-planar images: application to diffusion tensor imaging.
NeuroImage, 2003; 20(2):870-888.
3. Smith
SM, et al. Advances in functional and structural MR image analysis and implementation
as FSL. NeuroImage, 2004; 23(S1):208-219.
4. Chen NK, Guidon A, Chang HC, Song AW. A robust multi-shot
scan strategy fot high-resolution diffusion weighted MRI enabled by multiplexed
sensitivity encoding (MUSE). Neuroimage, 2013; 72:41-47.
5. Pruessmann KP, et al. SENSE: Sensitivity Encoding for
Fast MRI. Magn. Reson. Med. 1999; 42:952-962.
6. Jenkinson M, et al. FSL. Neuroimage, 2012;
63:782-790.