Michael Carl1, Yajun Ma2, and Jiang Du2
1GE Healthcare, San Diego, CA, United States, 2UCSD, San Diego, CA, United States
Synopsis
We compared different
center out 3D radial trajectories and assess their advantages and disadvantages
for short T2 MSK imaging. We found that while ZTE and Cones may
provide some unique capabilities, UTE represents a good compromise between the
two.
Introduction
Direct MR imaging of tissues such as tendons, or ligaments,
which have short transverse relaxation times (T2s) has become
possible using ultrashort echo time (UTE) sequences [1-4]. There are several
excitation RF and k-space trajectories available for UTE imaging. In this work, we compare different center out 3D radial
trajectories and assess their advantages and disadvantages for short T2
MSK imaging.Theory
The pulse sequences are shown in Fig.1. Both UTE and Cones employ
their RF pulses while the read gradients are ramped down, while in ZTE the read
gradients are already ramped-up during excitation (see Fig.1A). Some consequences
are:
1) The flip angle in ZTE is limited due to the need to keep the RF
excitation pulse very short.
2) UTE and Cones can apply a slice-selection gradient, which allows
application of slab selection.
3) UTE and Cones acquire data during partially time-varying read
gradients, while ZTE uses an already fully ramped read gradient which means
that ZTE does not suffer from gradient delay errors.
4) Both UTE and Cones allow to acquire multiple echoes at later
TEs.
5) Cones has a flexible read gradient design that allows a
spiral-like k-space trajectory, which can cover more k-space data per RF
excitation.
Methods
To experimentally study
the blurring characteristics, we scanned a plastic doll made of soft rubber (short
T2* of about 400𝜇s). Both UTE and ZTE scans were performed with BW of 31.25kHz,
62.5kHz, and 125kHz. For the in-vivo scans
we imaged the axial tibia using inversion recovery (IR) prepared UTE at two BWs
(31.25kHz and 125kHz). To study the RF
excitation, we scanned a resolution-phantom in the axial plane, focusing only
on a small axial section that contained structure. The UTE scan were acquired
using slab selective pulses (at Ernst angle) to show the best possible
image quality (artifacts and SNR). Similar scans were performed in the sagittal
spine.
Finally, multi-echo UTE and Cones scans were performed in the sagittal
knee. Two separate dual echoes acquisitions were obtained to highlight the
flexibility of the Cones trajectory. Relevant scan parameters were FOV=25cm, matrix=256, slice thickness=3mm,
20slices, BW=125kHz. For the first scan, the second echo was chosen near the
minimum available echo spacing (TE2=2.2ms). For the second acquisition, the
second echo was chosen larger at TE2=8.8ms, meaning that the Cones sequence was
able to acquire more k-space data per TR period.Results
The images of the short T2
doll-phantom are shown in Fig.2. For either sequence there is significant short
T2 blurring for the 31.25kHz BW scans, while the best result is obtained with
UTE using BW=125kHz.
The axial tibia images are shown in
Fig.3, which show similar blurring for the lower BW acquisitions.
The axial scans in the resolution
phantom to study the RF excitation differences are shown in Fig.4. Fig.4A shows
the center slice of the ZTE acquisition. Since no slab selection was applied,
there exist a lot of excited signals from the phantom outside the encoded 3D
volume, which manifests as artifacts in the final image. Fig.4B shows
the UTE acquisition using a slab-selective RF pulse. The corresponding in-vivo
images are shown in Fig.5 in the sagittal spine. Since no slab-selection was
applied in ZTE, there exists a lot of excited signals from either the left or
right side of the spine which manifests as artifacts in the final image. In the corresponding UTE acquisition using a slab-selective RF pulse,
these artifacts have been eliminated.
Finally,
Fig.6 shows several dual-echo scans using UTE (top) and Cones (bottom). The left
column corresponds to the acquisitions with the shorter second echo
(TE2=2.2ms). Here both UTE and Cones show similar image quality and SNR. The right
column shows the corresponding results for the longer second echo (TE2=8.8ms).
In this case, compared to UTE, the Cones acquisition shows both less artifacts
and more SNR, due to the more efficient k-space sampling duty cycle of Cones.Conclusion
We have examined three
ultra-short-TE pulse sequences for their performance in SNR, artifact
characteristics and multi-echo capability. While ZTE and Cones may provide some
unique capabilities, UTE represents a good compromise between the two. Although
our current version of ZTE was not able to acquire multi-echo images,
future work may include a version of that psd that has such capability (such as
looping-star).Acknowledgements
No acknowledgement found.References
[1] Rahmer et al, Magn
Reson Med, 2006. 55(5): p.1075-82.
[2] Du et al, Magn Reson
Imag 2011:29:470–482
[3] Weiger et al, NMR Biomed. 2015 28(2):247-54
[4] Li et al, Magn Reson Med 2012 68(3):680