Suryanarayanan Sivaram Kaushik1, Graeme McKinnon1, Matthew Koff2, Hollis Potter2, and Kevin Koch3
1MR Applications and Workflow, GE Healthcare, Waukesha, WI, United States, 2Hospital for Special Surgery, New York City, NY, United States, 3Medical College of Wisconsin, Milwaukee, WI, United States
Synopsis
3D
Multi-spectral imaging sequences like MAVRIC SL have been used overcome susceptibility
artifacts caused by metallic hardware. Several studies have attempted to reduce long
scan times by using parallel imaging and compressed sensing, but PD images are
commonly displayed with evident blurring. MAVRIC STIR images, due to their
sparseness, may be more amenable to compressed sensing acceleration. The
current work focuses on accelerating MAVRIC SL STIR images using compressed
sensing and a total generalized variation reconstruction (TGV).
Introduction
3D Multi-Spectral Imaging sequences, like MAVRIC
SL and SEMAC, significantly reduce the susceptibility artifacts caused by metal
instrumentation [1,2]. This artifact reduction is achieved by sampling the
broadened frequency distribution and acquiring images (spectral bins) at
distinct frequency offsets from the Larmor frequency, which are combined to minimize the susceptibility artifacts. The longer scan times of
these series have been accelerated using parallel imaging, partial Fourier
acquisition, and cutting the corners of k-Space. In addition, compressed
sensing has been used in combination with parallel imaging to accelerate the 3D
MSI acquisition [3,4,5]. Koch et al. showed that wavelet based regularization
in the ARC+CS algorithm leads to prohibitive blurring in PD images [3]. More
recently, Levine et al. used a Low Rank + sparse algorithm – which used
in-plane acceleration and further accelerated the bin dimension using a variable
density sampling pattern – to significantly reduce scan time [5]. While this resolved
blurring away from the metal, regions closer to the implant were still affected.
In this project, we focus the compressed sensing reconstruction on the sparser
STIR images, which should more readily lend itself to compressed sensing
acceleration. Here, we apply a total generalized variation (TGV) reconstruction
[6] to reconstruct individual spectral-bins before they are combined to yield
an artifact minimized image. Methods
Images were acquired on a clinical 1.5T MR450W
scanner (GE Healthcare, Waukesha WI), under an IRB approved study with written
informed consent from the patients. A spectral calibration scan was run to
determine the number of bins needed for each implant [7]. A conventional MAVRIC
SL STIR was acquired with 24 spectral bins and the following parameters: matrix:
256x192, FOV: 20 – 40cm, slice thickness: 5.5mm, ETL: 20, TE/TI/TR: 7/150/4237ms,
NEX: 0.5, ARC: 2x2. The STIR acquisition was repeated with a B1-optimized
inversion pulse [8], fewer spectral bins (as determined by the calibration
scan), and the following parameters: matrix: 256x192, FOV: 20 – 43cm, slice
thickness: 4mm-5.5mm, ETL: 48 with variable flip angles, TE/TI/TR:
7/150/4237ms, NEX: 1, ARC: 2x2, CS factor: 1.4. To test the impact of the
sampling pattern on image quality, in one subject, higher resolution images
were acquired using both the ARC+CS sampling pattern and a variable density
Poisson disc sampling (VDPDS). The images were reconstructed first using the
ARC+CS algorithm which first performs wavelet regularized reconstruction using
a conjugate gradient descent optimizer to fill the randomly undersampled
locations, and then performs ARC reconstruction to fill the uniformly under
sampled locations. The data were also reconstructed using the TGV algorithm
which iteratively fills in both the regularly and randomly undersampled regions
of k-space. The coil sensitivity profiles were determined by low pass filtering
the k-space data, the TGV regularization constant was empirically chosen to be
0.5, and the algorithm used 50 iterations for each spectral bin. Reconstruction
was done in C++ using the Orchestra SDK (GE Healthcare, Waukesha WI). Results
Figure 1 shows data from a subject with an
implanted suture anchor to secure the supraspinatus tendon to the proximal
humerus. This data set was acquired with 8 spectral bins, in 3:30 min. The
conventional ARC+CS reconstruction retains a significant degree of noise, which
has been minimized with the TGV reconstruction. An additional example of the
improvement to the STIR image quality can be seen in Figure 2, where the lower
resolution of the product STIR could be traded in for higher resolution, a
full-Fourier acquisition, thinner slices, and with the addition of compressed
sensing, can be achieved in a scan time that is on par with the lower
resolution clinical scan (4:23 min vs. 5:11 min). Different sampling patterns
are compared in figure 3, with the VDPDS sampling resulting in a more uniform
image. Discussion and Conclusions
For
the sparse MAVRIC SL STIR images, the TGV regularizer is advantageous as it does
not severely penalize edges and also denoises the image. This features permits
reduction of the voxel size with a concomitant full-Fourier acquisition to negate
the SNR losses induced by a homodyne reconstruction. Most importantly, these
features can be implemented while maintaining a scan time similar to the lower
resolution clinical STIR acquisition. The data set acquired with a variable
density Poisson disc sampling suggests that a choice of sampling pattern may potentially
improve the performance of the TGV reconstruction.Acknowledgements
No acknowledgement found.References
1. Koch, KM,
Brau, AC, Chen, W et al. “Imaging near metal with a MAVRIC-SEMAC hybrid”, Magn
Reson Med 2011;65:71-82.
2. Lu, Wenmiao, et al. "SEMAC: slice encoding
for metal artifact correction in MRI." Magnetic resonance in
medicine 62.1 (2009): 66-76.
3. Koch, KM, King, KF “Combined parallel
imaging and Compressed Sensing on 3D Multi Spectral Imaging near metal
implants”, #3172, Proceedings of the ISMRM 2011.
4. Worters,
Pauline W., et al. "Compressed‐Sensing
multispectral imaging of the postoperative spine." Journal of
Magnetic Resonance Imaging 37.1 (2013): 243-248.
5. Levine,
Evan, et al. "Accelerated three‐dimensional
multispectral MRI with robust principal component analysis for separation of on‐and
off‐resonance
signals." Magnetic Resonance in Medicine (2017).
6. Knoll,
Florian, et al. "Second order total generalized variation (TGV) for
MRI." Magnetic resonance in medicine 65.2 (2011):
480-491.
7. Kaushik,
S.S. et al., "External calibration of the spectral coverage for three‐dimensional
multispectral MRI." Magnetic resonance in medicine 76.5
(2016): 1494-1503.
8. Kaushik, S.S. et al.,
“Lipid suppression around metal implants using a B1-optimized adiabatic
inversion pulse”, #5028, Proceedings of the ISMRM 2016.