Kai Herz1, Chirayu Gandhi1, Klaus Scheffler1,2, and Moritz Zaiss1
1Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Department of Biomedical Magnetic Resonance, University of Tübingen, Tuebingen, Germany
Synopsis
Off-resonant spin-lock imaging enables a lot of
possibilities for T1ρ and chemical exchange (CE) sensitive applications. For
this purpose, a matching amplitude of the tipping and the locking pulse is
required, which can be difficult due to the high power requirements of
adiabatic pulses. In this work, we present a newly shaped adiabatic
half-passage pulse, usable at low power to match the amplitude of the pulses.
Off- and on-resonant saturated images acquired at 9.4T are shown. The new pulse
shape is able to generate robust images with comparatively low power at
ultra-high-field strengths.
Introduction
In previous work it was shown, that hyperbolic secant (HS)
adiabatic spin-lock (SL) pulses are a very useful tool to investigate chemical
exchange (CE) processes at ultra-high field (UHF), at either on-resonant 1
or off-resonant frequencies 2. Despite these promising results and the
fact that SL pulses are in principle robust against B1
inhomogeneity, they have still some disadvantages. First, HS pulses need very
high power for robust results, which can be difficult in terms of SAR and
hardware limitations, especially at UHF. In addition, efficient labeling of CE
effects requires a low amplitude for various metabolites. Second, an amplitude
difference between the tipping and locking pulses makes the images prone to B0
artifacts and prohibits off-resonant imaging. Therefore, in this work we aim to
design a pulse that shows adiabatic behavior with low power, to match the
amplitude of the tipping and locking pulses.Methods
To get more freedom in the pulse design, we introduce a
Blackman-windowed HS amplitude modulation in combination with a weighted
exponential frequency modulation (HSExp). The pulse shape is thus defined by
the bandwidth (BW), maximum amplitude (B1,max), pulse duration (tp),
length of the Blackman window (twindow) and the dimensionless
scaling factor µ. In order to find the most stable pulse, a Bloch-Simulation
was used to simulate Z-spectra at different levels of B0 and B1
inhomogeneity, as well as varying pulse shape parameters. Fixed parameters were
B1,max = 5 µT and tp = 8 ms. The optimized shape was then
used to acquire images in a glucose phantom and two healthy volunteers at 9.4T
(Siemens MAGNETOM) using a 16Tx/31Rx channel head coil 3. SL
prepared Imaging consists of a SL saturation pulse followed by a single shot 3D
GRE readout 4 (FA=5°, TE=1.96 ms, TR=3.86 ms, Grappa=3, elliptical
scanning, resolution 1.5x1.5x2 mm3). Z-spectra were acquired with the HSExp
pulse and with a Gaussian saturation pulse for comparison. For the phantom, 96 offsets
between ±4.5 ppm were acquired with a locking duration tlock = 35
ms. In-vivo, 92 offsets between ±3 ppm were acquired with tlock=50 ms.
Since the actual B1 is reduced through inhomogeneity, B1,max
was set to 6.5 µT to achieve a mean B1 of 5 µT in the FOV.
Results
The best pulse found by simulation is defined by the
following parameters: BW: 3000 Hz, µ: 45, and twindow: 2.5 ms. The
amplitude of the rectangular locking pulse is set to the B1,max of
the tipping pulse. The Z-spectra calculated from the HSExp prepared images show
a smooth transition from off- to on-resonance without apparent oscillation
artifacts (Figure 1). The solution with the higher glucose concentration shows
a slightly higher asymmetry with both pulses. In addition, MTRasym
images show a stronger contrast (Figure 2) with HSExp than Gauss pulses.
However, B1 inhomogeneity can still cause oscillation artifacts at
frequency offsets close to the Larmor frequency.
The in-vivo Z-spectra show a similar behavior as in the
phantom. Again, the transition from off- to on-resonant sample points is smooth
with the HSExp pulse. However, they are broader in general since the saturation
is sensitive to T2 and MT effects. (Figure 3). Both pulses show
clear correlation between the contrast and the B1 amplitude (Figure
4). Images acquired with HSExp show only slight B0 artifacts in
regions with very high inhomogeneity.Discussion
An amplitude matched HS pulse for T1ρ
measurements has been used in the past 5. Although off-resonant SL
imaging was shown in these studies, the applied B1 amplitude was
very high. The proposed HSExp prepared images show a robust contrast, for both
on- and off-resonant frequencies with a lower B1 amplitude. Thus, in
the presence of a very strong B1 inhomogeneity, slight oscillation
artifacts appear due to the extreme amplitude decrease. However, B1,max
of the pulse could be increased if the ROI lies in such a region, to overcome
this problem. Dynamic glucose enhanced imaging (DGE) is another field of
application for SL pulses 6. The presented HSExp pulse adds the
possibility of off-resonant imaging to DGE applications. A potential clinical
benefit of this method must be investigated in the future.Conclusion
By introducing a new pulse shape, we were able to generate a
SL prepared contrast with low power and a matching tipping and locking
amplitude. Since these pulses can be used for off-and on-resonant SL imaging,
there are other possible applications besides the presented Z-spectrum
acquisition, e.g. off-resonant T1ρ-mapping or dynamic glucose
enhanced imaging. It could be shown that efficient glucose labeling is possible
in a phantom. This is currently under investigation in-vivo.Acknowledgements
The financial support of the Max Planck Society, German
Research Foundation (DFG, grant ZA 814/2-1, support to M.Z.), and European
Union’s Horizon 2020 research and innovation programme (Grant Agreement No.
667510, support to M.Z.) is gratefully acknowledged.References
[1] SCHUENKE, Patrick et al. Adiabatically prepared spin‐lock approach for
T1ρ‐based dynamic glucose enhanced MRI at ultrahigh fields. Mag Reson Med 2017;
78 (1): 215-225.
[2] JIN, Tao et al. Advantages of chemical exchange‐sensitive spin‐lock
(CESL) over chemical exchange saturation transfer (CEST) for hydroxyl–and
amine–water proton exchange studies. NMR in Biomed 2014; 27(11): 1313-1324.
[3] GUNAMONY, Shajan et al. A 16‐channel dual‐row transmit array in
combination with a 31‐element receive array for human brain imaging at 9.4 T. Mag
Reson Med 2014; 71(2): 870-879.
[4] ZAISS, Moritz et. al. snapCEST – A single-shot 3D CEST sequence for
motion corrected CEST MRI. In Proceedings of 25th Annual Meeting of ISMRM 2017,
3768.
[5] GRÖHN, Olli HJ et al. On‐and
off‐resonance T1ρ MRI in acute cerebral ischemia of the rat. Mag Reson Med 2003;
49(1): 172-176.
[6] SCHUENKE, Patrick et al. Fast and Quantitative T1ρ-weighted Dynamic
Glucose Enhanced MRI. Scientific Reports 2017; 7:42093.