Leonardo A Rivera-Rivera1, Tilman Schubert2, Patrick A Turski1,3, Oliver Wieben1,3, and Kevin M Johnson1,3
1Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 2Department of Radiology, Basel University Hospital, Basel, Switzerland, 3Department of Radiology, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
Physiological parameters
derived from quantitative flow MRI can potentially improve characterization of
a large spectrum of vascular diseases if routinely used in a clinical setting.
However, current barriers limit the use of quantitative flow MRI in a clinical
setting, partially due to a lack of calibration tests, and concerns regarding
accuracy and reproducibility. In this study we investigate the potential
induced bias of flow measurements in a cranial 4D flow MRI acquisition due to
signal magnitude heterogeneity, and the implications for comparing protocols
with differing flip angle or contrast agent usage.
INTRODUCTION:
Recent advances have
facilitated the use of 4D flow MRI in clinically acceptable acquisition times1, however accuracy and reproducibility are still major concerns limiting
widespread clinical implementation. Velocity maps in PC-MRI are known to be susceptible
to partial volume effects, arising from both low resolution and the unequal
signal magnitude in background and within the vessel of interest.2 These
effects are pronounced in smaller vessels such as those found in the intracranial
vasculature. Such artifacts are often not readily observed from the raw data, therefore
it is of importance to optimize 4D flow MRI protocols and/or correct for these
effects using additional geometric modeling.3 In this study, we investigate
the potential flow bias introduced by magnitude changes when using protocols
with differing flip angles or contrast media (with vs without) in a cohort of
healthy volunteers.METHODS:
Subjects: 13 healthy volunteers
(age range 24-51y, mean=32y, 5F) participated in this study. With IRB approval
and HIPAA compliance, written informed consent was obtained from all study
subjects. For 9 volunteers scanning was performed before and
after the administration of ferumoxytol (AMAG Pharmaceuticals, Inc., Cambridge,
MA). Ferumoxytol was diluted to 60 ml and injected as a slow infusion. Images
were acquired at 2 time points: pre-contrast, and after a dose of 5 mg/kg of ferumoxytol.
MRI: Volumetric, time-resolved PC MRI data with 3-directional velocity encoding
were acquired on a 3.0T clinical MRI system (MR750, GE Healthcare) with an 32 channel
head coil (Nova Medical, Wilmington, MA, USA), with a 3D radially undersampled
sequence, PC VIPR4 with the following imaging parameters: flip angle = 16˚,
Venc = 80 cm/s, imaging volume = 22x22x16 cm3, 0.7 mm acquired
isotropic resolution, TR/TE=5.78/1.74ms,
scan time ~ 5 min, retrospective cardiac gated into 20 cardiac phases
with temporal interpolation5. Four volunteers did not receive contrast
agent, and were scanned four times with the same imaging parameters but varying
flip angles [4˚, 8˚, 12˚, 16˚]. Flow analysis: Automatic vessel segmentation
and flow quantification was performed in a customized Matlab tool6 (Mathworks, Natick, MA) from the reconstructed 4D flow MRI data. A centerline
guided flow tracking algorithm was used to visualize and select the arterial
segments for further analysis. Flow rates and signal of the magnitude images were
calculated for each segment by averaging data obtained in five local cross-sectional
cut-planes automatically placed in every centerline point perpendicular to the
axial direction of the vessel. For this purpose 4 segments were selected
(figure 2 a.): distal cervical internal carotid artery (ICA) (left and right), distal
petrous ICA (left and right), and MCA M1 segment (left and right). RESULTS:
Figure
1 shows signal magnitude data from a volunteer shown as colormap on the maximum
intensity projection (MIP) of the PC-MRA with a 4º flip angle. As the flip
angle is increased the signal of the magnitude data increases near the region
closer to the slab placement (distal cervical ICA), while it decreases along
the ICA length and into smaller vessels (white arrow). Figure 3 shows how the
flow and magnitude changes along the ICA to MCA as a function of the flip
angle. A flip angle of 8˚ shows the least variation along the vessel length,
while 16˚ shows the largest variation. Figure 2 shows MIP images from a different volunteer before and after
the administration of 5 mg FE/kg. These images were acquired with the same
parameters and a flip angle of 16˚. Post contrast images demonstrated a
significantly higher SNR especially in smaller vessels when compared to the pre
contrast scan. Figures 4 and 5 show box plots of the flow and magnitude signal
in the pre- and post-contrast images. The flow and magnitude decreased
considerably along the ICA and MCA in the pre contrast scan.DISCUSSION AND CONCLUSION:
This work demonstrates that caution should be
placed when setting the flip angle for intracranial flow quantification.
Theoretically, a flip angle creating flat, proton density weighted image should
allow for the most accurate, although potentially noisier, flow quantification.
Using a higher flip angle, can result in either an increase or decrease in the
flow value as demonstrated with the increase in flow in non-contrast studies
(4˚ vs 8˚) and decrease from (12˚ to 16˚). Using contrast media unifies the
signal along the vasculature but is not practical in all studies and may
potentially overestimate flow. Thus we recommend flip angle be set as low as
reasonably allowable for maximal study reproducibility.Acknowledgements
We gratefully acknowledge R01NS066982, and GE
Healthcare for assistance and support.References
- Markl
M, et al. 4D flow MRI. J Magn Reson Imaging 2012; 36: 1015–1036.
- Peng
SL, et al. Optimization of phase-contrast MRI for the quantification of
whole-brain cerebral blood flow. J Magn Reson Imaging 2015; 42:1126–1133.
- Bouillot
P, et al.3D phase contrast MRI: Partial volume correction for robust blood flow
quantification in small intracranial vessels. Magn Reson Med. 2017 Feb 28. doi:
10.1002/mrm.26637
- TL
Gu et al. PC
VIPR: a high-speed 3D phase-contrast method for flow quantification and
high-resolution angiography. AJNR 26(4):743-9 (2005).
- J
Liu et al. Generation and visualization of four dimensional MR angiography data
using an undersampled 3-D projection trajectory. IEEE TMI 25(2):148-57 (2006).
- Schrauben
E, Wåhlin A, Ambarki K, Spaak E, Malm J, Wieben O, et al. “Fast 4D flow MRI
intracranial segmentation and quantification in tortuous arteries.” J Magn
Reson Imaging. 2015 Nov;42(5):1458-64.