Mariska Damen1, Maarten S. van Leeuwen2, Peter R. Luijten2, Andrew Webb1, Dennis WJ Klomp2, and Catalina S. Arteaga de Castro2
1Radiology, Leiden University Medical Center, Leiden, Netherlands, 2Imaging, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
Age dependency of T1
and T2 times of the healthy pancreas at 7 Tesla with a multi-transmit system was
investigated for MRI protocol optimization of the pancreas. Three age groups were
measured (21 -25 yo, 26 -39 yo and 40 -72 yo). Measurements resulted in average
T2 times of 52 ms, 97 ms and 127 ms for the 3 different age groups respectively,
revealing an age dependency of the T2 relaxation times. Average T1 times were 856
ms, 899 ms and 858 ms for each age group respectively. No age dependency was
observed for the T1 relaxation times.
Introduction
Pancreatic cancer has the highest mortality rate
of all cancers: 10 out of 11 patients die within 5 years after diagnosis.1
Characterization of different tumor types is difficult, hampering optimal
treatment selection. 7 Tesla (7T) MR imaging potentially offers higher signal-
and contrast-to-noise and spatial resolution
than lower clinical fields which can be translated into improved diagnoses. However,
at 7T image artifacts such as signal voids can be substantial, particularly in the
abdomen, which may reduce image quality in the region of interest (ROI). By
using a multi-transmit system, the phase of each channel can be calculated to produce
constructive interference in the ROI without reducing the transmit field (B1+)
amplitude. In practice, MR protocols need to be optimized to enhance the
contrast between normal and diseased tissue, which requires knowledge of the
tissue relaxation parameters. The T2 value of pancreas at 7T has previously
been investigated in relatively young
volunteers (26±5 years) whereas the mean age of patients with pancreas cancer
is 70 years.1 It is known that the physiology of the pancreas
changes significantly with age, with factors such as pancreatic atrophy,
lobulation and fatty degeneration occurring which could substantially affect
relaxation times.3 The purpose of this study is to determine the
mean T1 and T2 values of pancreas at 7T in older healthy volunteers and therefore
the age dependency of the relaxation parameters.Methods
18 healthy volunteers, divided into three groups
with age ranges of 21-25 (group 1, averaged age 24.6±1.1 years), 26-39 (group 2,
averaged age 30.6±4.0 years) and 40-72 (group 3, averaged age 59.2±12.3 years)
years old, were scanned with a 7T MR system (Philips, Best, The Netherlands)
using eight parallel transmit channels, each connected to a transmit-receive
fractionated dipole antenna4 (MR Coils BV, Drunen, The Netherlands) and
16 additional receive loops integrated with the antennas (2 per antenna) that
were positioned symmetrically around the abdomen. The RF phase of each antenna was
optimized by a multi-transmit system. RF phase shimming was performed to
maximize the B1+ field in the pancreas region. After B1
shimming, gradient-echo images were obtained for anatomy localization. A Look-Locker
sequence was used to obtain image sets for T1 determination (Turbo field echo
(TFE), FA=3, TR/TE=9000/1.68ms, 20 IR times, 8:33 min total acquisition time,
4x4x10mm voxels, FOV= 222x340mm). Images acquired at a series of different echo
times in one slice containing the pancreas were acquired to determine T2 (2D
single shot turbo spin echo (TSE), TE/TR = 50;80;100;120ms/ 17s, SENSE factor =
2, 1.3x1.6x5mm voxels, FOV=250x350mm). To decrease motion artifacts, the study
was carried out with an approximately 3 second breath-hold for the TE series
and cardiac triggering for the Look-Locker sequence. Matlab scripts (R2014b,
The Mathworks, Inc. ©) were used to fit T1 and T2 times using mono-exponential decays with a non-linear least
squares method using the magnitude signal 5. An ROI was drawn to
include the entire pancreas. Values were obtained both as an average over the
ROI, and also on a voxel-by-voxel basis. Results and Discussion
T1 and T2 fits were
obtained for all volunteers. A substantial age dependency was observed for the
T2 relaxation times, being more than two times higher at higher ages. Figure 1
shows a single example from each group of a T2 weighted image with the
corresponding T2 map of the pancreas. Notice that with increasing age, higher
T2 values are observed in the T2 maps, which can also be seen in the increased
signal from the pancreas in the older groups. In the third age group, the
anatomy is clearly changed, showing lobulation and fat infiltration. Voxel-by-voxel
T2 fitting is more challenging in older pancreas, as observed in the T2 maps.
As the size of the pancreas decreases with age, the pancreatic ducts that
conduct fluids get closer together, causing motion artifacts and incorrect fits.
Figure 2, shows the average T2 relaxation time in the ROI per age group. The
average T2 values per group were 52±10 ms, 97±32 ms and 127±39 ms, respectively.
Figure 3 shows no significant age dependency observed for the T1 relaxation
times. The average T1 values per group were 856±134 ms, 899±108 ms and 858±209
ms, respectively.Conclusions
We have successfully measured the T1 and T2
values for pancreas at different ages at 7T. This resulted in an averaged T1 of
856±134 ms, 899±108
ms and 858±209 ms and a T2 value of 52±10 ms, 97±32 ms
and 127±39 ms for the 3 age groups, respectively. The T2 relaxation
times showed an age dependency, being higher for the older group.Acknowledgements
No acknowledgement found.References
1. National cancer
institute (2015) Cancer Stat Facts: Pancreas Cancer. Retrieved from https://seer.cancer.gov/statfacts/html/pancreas.html.
Accessed October 30, 2017.
2. Damen M., van
Houtum Q, van Leeuwen MS, Luijten Peter R., Webb A., Klomp Dennis W.J., Arteaga
de Castro CS. Quantitative T1 and
T2 measurements of pancreas at 7 Tesla using a multi-transmit system. Proc Intl
Soc Mag Reson Med; 2017 22-27 April; Honolulu, HI, USA.
3. Tomohiro Sato, et al., Age-related changes
in normal adult pancreas: MR imaging evaluation. European Journal of Radiology,
2011.
4. Raaijmakers,
A.J.E., et al., The fractionated dipole antenna: A new antenna for body imaging
at 7 Tesla. Magnetic Resonance in Medicine, 2015.
5. MacFall, J.R., et
al., An analysis of noise propagation in computed T2, pseudodensity, and synthetic
spin-echo images. Medical Physics, 1986.