Ana E RodrÃguez-Soto1, Michael C Langham1, Nadav E Schwartz2, and Felix W Wehrli1
1Department of Radiology, University of Pennsylvania, Philadelphia, PA, United States, 2Department of Obstetrics and Gynecology, Division of Maternal-Fetal Medicine, University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Quantitative
MRI allows the estimation of fetal oxygen transport in vivo, for which knowledge of the oxygen saturation (HbO2)
of blood in the umbilical vein (UV) is required. The method of choice to
estimate HbO2 in fetal applications is T2-based oximetry,
which requires a sequence-specific calibration equation to convert blood T2
to HbO2. Therefore, in the present work we examined the feasibility of
using susceptometry-based oximetry (SBO) to measure HbO2 at the UV as
it is calibration-free and implementable across field strengths. Results show,
in a limited number of participants, no difference in HbO2 measured
with both MRI-based oximetric techniques.
INTRODUCTION
Quantitative
MRI allows the estimation of fetal oxygen transport in vivo,1 for which knowledge of the oxygen saturation
(HbO2) of blood in the umbilical vein (UV) is required. Two MRI-based
oximetric methods exist: T2– and susceptometry–based oximetry (T2Ox
and SBO, resp.).2,3 Both techniques have unique strengths and
limitations that make them suitable for different applications. For example, T2–based
oximetry requires a sequence-specific calibration curve to convert whole blood
T2 to HbO2 levels. Such calibration curves are laborious
to produce, and once established, identical scanning parameters must be
maintained. In fact, the study of fetal oxygen metabolism has primarily relied
on T2Ox.1 On the other hand, SBO is a phase-based method
limited to relatively straight vessels w.r.t. the main field. In addition, SBO
requires vessel-adjacent tissue as a phase reference, a condition which is not
always met. However, SBO is calibration-free, resulting in a sequence implementable
across field strengths and system platforms. Therefore, in the present work we
explore the feasibility of utilizing SBO to quantify HbO2 in the UV.METHODS
Fourteen pregnant women (gestational age, GA=26±7wks,
range=17-37wks) were scanned at 1.5T
(Siemens Avanto, Erlangen, Germany) with two flexible body coils placed on the
abdomen. UV HbO2 was
estimated using SBO as $$$HbO_2=[1-\frac{2|\Delta\phi|}{\gamma\chi_{do}\Delta TE \cdot{B_0}(\cos^2\theta-1/3)Hct}]$$$, where Δφ is the
average phase difference between intravascular blood and the surrounding
tissue, χdo is the
susceptibility difference between fully deoxygenated and fully oxygenated erythrocytes, ΔTE is the
inter-echo spacing, θ is the
angle of the UV w.r.t. B0 and Hct
is the hematocrit.3 Once a relatively straight section of the UV was identified SBO data were acquired
using an axial single-slice three-echo RF-spoiled gradient echo (GRE)
acquisition (Figure 1A) interleaved with a dual-slice acquisition using simultaneous echo
refocusing (SER, Figure 1B). These
two slices were located 10mm superior and inferior to the SBO imaging slice,
yielding a total of three images (Figure 2A). Imaging parameters: TRGRE=25ms, TEGRE=10ms,
TRSER=20ms, TESER,1=8.8ms, TESER,2=3.5ms, FOV=307×307mm2,
voxel size=1×1×5mm3, FA=15°, and scan duration of 20sec. Phase accrual
difference between the UV and surrounding tissue, Δφ , was
estimated between the first and third echoes of the RF-spoiled GRE multi-echo
GRE (Figure 2B). UV and reference tissue regions of interest (ROIs) were manually traced on
the magnitude image using ImageJ4 and applied to a bias field corrected
phase map.5 The angle θ was estimated using the two images acquired
with SER (Figures 2C&D). Hematocrit (Hct) is needed in both oximetric methods, in the present work
population averages were used.6 The average Hct for GA<22wks was
not available, in such cases Hct(GA=22) was used. T2Ox HbO2 data were
acquired using a background suppressed (BS)7 T2-prepared
balanced steady-state free precession (BS-T2bSSFP)8
sequence, not necessarily in the axial plane. Fetal blood T2 values were converted
to HbO2 values using an established calibration equation.9
Imaging parameters: T2-preparation TEs=0,48,96,144,192ms, bSSFP TR/TE=4.1/2.05ms,
duration of BS module 3000ms, FOV=300×300mm2, voxel size=0.8×0.8×5mm3,
FA=60°, half-Fourier with 14 reference lines, inter-pulse interval (τ180)
of 12ms and scan duration of 1:06 minutes. Intra-class correlation
coefficient (ICC), Bland-Altman and paired t-tests were used to evaluate agreement
between both oximetry methods.RESULTS
Example datasets are shown in Figures 3 and 4. T2Ox was successfully acquired
in 8 pregnant women of GA≥26wks, while SBO was possible in 10 cases at GA≥18wks
(Table 1). HbO2 was
estimated with both T2Ox (average
HbO2=83±7%) and SBO (average HbO2=83±6%) in seven cases
all at GA≥26wks (Table 1). Results
from the two methods were well correlated (HbO2,SBO=0.7·HbO2,T2+0.2, R2=0.84,
intra-class correlation coefficient 0.71, Figure
4C). However, no significant difference was found between methods (paired
t-test, p>0.05), confirmed by the Bland-Altman analysis showing a mean
difference (HbO2,T2–HbO2, SBO) of -0.3%, well within the
95% confidence interval (±2.9%, Figure 4D).DISCUSSION AND CONCLUSIONS
Good agreement was found between BS-T2Ox
and SBO in the UV. The earliest GA at which it was possible to extract UV HbO2
via T2Ox was 26
weeks for the present study and 23 weeks in general in the authors’ experience.
Here, we explored the feasibility of utilizing SBO for this purpose and found
that such measurements were achievable as early as 18 weeks of GA. Although HbO2
measurements by both oximetry methods were not found to be different from each
other, these data should be interpreted with some caution given the small number
of participants. Further, it must be kept in mind that SBO HbO2 may
be susceptible to partial volume effects, which lead to overestimated HbO2
values and, of course, is subject to the usual limitations in terms of vessel
angle with respect to static magnetic field. It is of particular
clinical relevance to determine how early during pregnancy either method is
applicable. Acknowledgements
Human Placental Project U01 HD087180.References
1.
Sun L, Macgowan CK, Sled JG, Yoo SJ, Manlhiot C, Porayette P, Grosse-Wortmann L, Jaeggi E, McCrindle BW, Kingdom J and others. Reduced fetal cerebral oxygen consumption is associated with smaller brain size in fetuses with congenital heart disease. Circulation 2015;131(15):1313-23.
2.
Wright G, Hu B, Macovski A. Estimating oxygen saturation of blood in vivo with MR imaging at 1.5T. Magn Reson Med. 1991;1:275-83.
3. Fernandez-Seara MA, Techawiboonwong A, Detre
JA, Wehrli FW. MR susceptometry for measuring global brain oxygen extraction.
Magn Reson Med. 2006;55:967-73.
4. Schneider, C.A., Rasband, W.S., Eliceiri, K.W.
"NIH Image to ImageJ: 25 years of image analysis". Nature Methods 9,
671-675, 2012.
5.
Langham MC, Magland JF, Floyd TF, Wehrli FW. Retrospective correction for induced magnetic field inhomogeneity in measurements of large-vessel hemoglobin oxygen saturation by MR susceptometry. Magn Reson Med 2009;61(3):626-33.
6.
Jopling J, Henry E, Wiedmeier SE, Christensen RD. Reference ranges for hematocrit and blood hemoglobin concentration during the neonatal period: data from a multihospital health care system. Pediatrics 2009;123(2):e333-7.
7.
Maleki N, Dai W, Alsop DC. Optimization of background
suppression for arterial spin labeling perfusion imaging. MAGMA
2012;25(2):127-33.
8.
Rodríguez-Soto AE, Abdulmalik O, Langham MC, Schwartz N, Lee
H, Wehrli FW. T2 -prepared balanced steady-state free precession (bSSFP) for
quantifying whole-blood oxygen saturation at 1.5T. Magn Reson Med 2017.
9.
Rodríguez-Soto AE, Langham MC, Abdulmalik O, Englund E, Schwartz
N, Wehrli FW. MRI
Quantification of Human Fetal O2 Delivery Rate in the Second and
Third Trimesters of Pregnancy. Magn Reson Med
2017. In Press.