Wei-Ching Lo1, Yun Jiang2, Leonardo Kayat Bittencourt3,4, Junichi Tokuda5,6, Ravi Seethamraju7, Clare Tempany-Afdhal5,6, Ananya Panda2, Katherine Wright2, Mark Griswold1,2, Nicole Seiberlich1,2, and Vikas Gulani1,2
1Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States, 2Department of Radiology, University Hospitals Cleveland Medical Center at Case Western Reserve University, Cleveland, OH, United States, 3CDPI and Multi-Imagem Clinics, Rio de Janeiro, Brazil, 4Department of Radiology, Universidade Federal Fluminense, Niterói, RJ, Brazil, 5Department of Radiology, Harvard Medical School, Harvard University, Boston, MA, United States, 6Department of Radiology, Brigham and Women’s Hospital, Boston, MA, United States, 7Siemens Healthineers, Boston, MA, United States
Synopsis
MRF enables rapid collection of multiple tissue properties simultaneously. For
clinical applications, the T1 and T2 values must be repeatable over time and on
different MRI scanners so that any observed relaxivity difference can be
assumed to be due to differences in physiology rather than scanner instability,
differences in pulse sequence or map reconstruction implementations. This study
evaluated multicenter repeatability and reproducibility of T1 and T2 estimates of
MRF in the ISMRM/NIST MRI system phantom and normal prostate regions in
patients. The intra-scanner
variation was less than 2% for MRF T1 and 4.7% for T2 within the biological
range.
Purpose
Magnetic Resonance Fingerprinting (MRF) [1] enables
simultaneous generation of quantitative maps of multiple tissue properties and potentially enables the
development of standardized imaging
biomarkers. For clinical applications, the T1 and T2 values must be repeatable
so that any observed relaxivity difference within a tissue can be assumed to be
due to differences in physiology rather than scanner instability and difference in
pulse sequence or map reconstruction implementations. Previous
research has shown that highly repeatable T1 and T2 measurements can be made over
time with MRF [2] with excellent reproducibility across several scanner types [3].
The purpose of this study was to evaluate multicenter repeatability and
reproducibility of T1 and T2 estimates of MRF using the ISMRM/NIST MRI system
phantom [4] and normal prostate regions in patients.Methods
Experiments were performed on four different 3T scanners (1 Skyra
and 3 Verio, Siemens Healthcare,
Erlangen, Germany) with different software versions (VE11C, VB19, and VB17) in
three different medical institutions: University Hospitals Cleveland Medical
Center at Case Western Reserve University (CWRU) and Brigham and Women's
Hospital (BWH) in US, and Multi-Imagem Clinic (DASA) in Brazil using the MRF-FISP
acquisition [5] with the following parameters: FOV 400x400mm2; matrix 400x400; flip
angle 5-75°; TR 11.2-15ms; slice thickness 5mm; acquisition time 39 s/slice.
During MRF acquisition, the raw data were accumulated and passed on to the
Gadgetron MRF reconstruction pipeline [6] along with a pre-calculated
dictionary [7]. The accuracy of the reconstruction was
validated using the T2 layer of ISMRM/NIST MRI system phantom with T1 values between 307.7 and 2360ms and T2 values between 19.1 and
479.2ms. The phantom was placed in the magnet for at least 20 minutes before
the acquisition to decrease the motion effects. No B0 and B1 maps were collected
in this study. Six MRF measurements were made during one session with a delay
of 5 secs between measurements. The phantom was then moved out of the magnet and
placed again in the magnet, and the phantom was again allowed to settle for at
least 20 minutes followed by another MRF acquisition for same-day test-retest reproducibility.
The repeatability of
T1 and T2 estimates was assessed using the coefficient of
variation (CV), defined as the ratio of the standard deviation to the
mean T1 and T2 values over 6 measurements. The CV of T1 and T2 estimates between
the four clinical scanners is shown to demonstrate inter-scanner variation. The
same-day test-retest normal prostate mean T1 and T2 values from right
peripheral zone (RPZ), left peripheral zone (LPZ), right transitional zone
(RTZ), and left transitional zone (LTZ) is also shown.Results
Over the wide ranges of T1 and T2 values found
in the ISMRM/NIST system phantom, intra-scanner MRF T1 estimates had less than 2%
variation and T2 estimates had less than 4.7% variation, with the exception of
T2 relaxation of 19.1ms, which showed a variation of 8.9% (Fig.1a,b). The
short T2 relaxation times were on the order of the TR used for the MRF
measurement. The test-retest reliability coefficient for both T1 and T2
estimation was above 0.99. Inter-scanner variation of mean from all 6 measurements and only
measurement #5 showed T1 variation less than 4.9% and T2 variation less than 8.1% between multicenter
scanners, and the variation increased when T2 was lower than
28.8ms (Fig.2). The same-day test-retest in in-vivo normal prostate T1 and
T2 values showed good agreement in four different regions (Fig.3). Figure 4 shows representative prostate MRF T1
and T2 maps from the four different scanners.Discussion
The
T1 and T2 values measured with MRF-FISP using the ISMRM/NIST
MRI system phantom and from normal prostate regions in patients were
highly repeatable and reproducible over multiple repetitions and different
scanners. High repeatability is also observed in the inter-scanner data, even
with different position, room temperature and inhomogeneous B1. The lower test-retest
agreement in NPZ at very high T2s (~300 ms) is likely due to dictionary coarseness; T2 step size was set to 10ms
from 160 to 300ms and 50ms above 300ms. These settings can
be optimized for future applications. These results demonstrate that T1 and T2
values are highly comparable between multicenter scanners, and that quantitative multiparametric
imaging with MRF for multicenter clinical studies may be feasible.Conclusion
MRF measurements of
T1 and T2 are highly repeatable and reproducible between MRI scanners at
different centers on different continents.Acknowledgements
Siemens Healthineers, 1R01EB016728, 1R01DK098503, 1R01CA208236, P41 EB 015898, R01EB020667References
[1] Ma et al. Nature. 2013;495:187-192. [2] Jiang
et al. MRM. 2017;78:1452-1457. [3] Körzdörfer et al. Proc. of ISMRM Workshop on
MRF. 2017. [4] Russek et al. Proc. of ISMRM, 2012. #2456. [5] Jiang et al. MRM.
2015;74:1621-1631. [6] Lo et al. Proc. of ISMRM Workshop on MRF. 2017. [7]
McGivney et al. IEEE TMI. 2014;33:2311-2322.