William Devine1, Francesco Giganti2, Edward Johnston1, Eleftheria Panagiotaki3, Shonit Punwani1, Daniel C. Alexander3, and David Atkinson1
1Centre for Medical Imaging, University College London, London, United Kingdom, 2Department of Radiology, University College London Hospital NHS Foundation Trust, London, United Kingdom, 3Centre for Medical Image Computing, University College London, London, United Kingdom
Synopsis
This work aims both to show that using a constrained fitting method for Luminal Water Imaging with fewer echoes produces accurate parameter estimates and that this fitting method is feasible for detecting and grading Prostate Cancer. Simulated signals were produced and the mean LWF across multiple iterations was calculated for two models. Then 19 patients were imaged and the images were contoured in both cancerous and benign regions. The simulation showed that the proposed constrained method is accurate and the in-vivo imaging reinforced the
idea that multi-echo T2 modelling shows promise in detecting and
grading PCa.
Introduction
Similar to methods developed for imaging myelin in the brain1,
Luminal Water Imaging (LWI), a fitting technique for multi-echo T2 data recently
proposed by Sabouri et al.2, calculates the percentage volume of
lumen within a voxel. LWI has been shown to be closely related to the luminal space in prostatic tissue2, with a recent study
demonstrating the feasibility of luminal water imaging in the detection and
grading of PCa3. However, the current method requires a 64-echo
acquisition which is time consuming and may not be available without dedicated
scanner research software. This work
seeks to show that the combination of an appropriately refined fitting method and
fewer acquired echoes provide values of the Luminal Water Fraction (LWF) that
are in close agreement with those of the original LWI. Correlations to Gleason
score were then performed along with the AUC, sensitivity and specificity of the
detection of PCa.
Methods
Model The original fitting proposed by Sabouri et al1 used a smoothed non-negative least squares (NNLS) fitting over many variables to produce a distribution $$$p$$$ of T2 values4. The method proposed here constrains this distribution to consist of two Gaussians. This constraint was imposed because the vast majority of voxels in the Sabouri studies2,3 showed two main signal components, which may provide a more robust fitting. The model relating the Signal $$$S$$$ to the distribution $$$p$$$ of T2 values is: $$S = M_0 \int_{0}^{\infty}p(T2)exp(-\frac{TE}{T2})dT2$$ where $$$M_0$$$ is the overall magnitude of the signal and $$$TE$$$ is the echo time. For each voxel six parameters were estimated: overall scaling ($$$M_0$$$), relative fraction between the two compartments (α), mean and standard deviation of the short T2 peak (μ1 and σ1) and mean and standard deviation of the long T2 peak (μ2 and σ2). The LWF was calculated as the area under the long T2 component divided by the area under both components.
Simulation
Simulated signal decays were produced for two separate
underlying T2 distributions (two delta functions at T2short and T2long and two Gaussian distributions with means T2short and T2long, fixed at T2short = 75ms and T2long = 600ms) with varying values of LWF, Signal-to-Noise
Ratio (SNR) and Number of Echoes (NE). For each combination of parameters 50
iterations of the signal were created. For each iteration the original and
new fitting methods were tested using 64-echo
and 32-echo simulation signals respectively. The mean LWF across the
50 iterations was then calculated for both models in order to show how the two
compare over a range of LWF and SNR values.
Data
In
order to investigate the relationship between this new fitting method and
Gleason Score, 19 subjects were imaged for this study. The scan parameters are
included in Figure 1. A board-certified radiologist contoured 15 lesions in 19
patients, which were confirmed to be cancers following targeted biopsy. Regions
of interest (ROIs) were also placed in 16 areas of normal tissue in each of the
subjects, all of which were histologically confirmed as benign.
Statistics
A
Spearman’s rank correlation test was used to find the correlation between
Gleason Score and LWF and the Area Under the Curve (AUC), sensitivity and
specificity values were calculated using a Receiver Operating Characteristic
(ROC) analysis with 5-fold cross validation to reduce problems arising from
multiple ROIs coming from the same image.
Results
Figure 1
shows the scan parameters of the 32-echo acquisition. Figure 2
shows that for nearly all values of LWF and SNR, using either the delta
function or Gaussian assumption, the LWF value produced by the new method is at
least as accurate as the original.
Using a Spearman’s rank correlation test
a correlation of -0.667 was found between Gleason Score and LWF in 31 ROIs. Figure 3 shows the AUC, sensitivity and specificity
achieved by the new fitting method in 31 ROIs.
Figure 4 is an example LWF map of the prostate. Note the
higher LWF in the peripheral zone (PZ), consistent with histological findings
of large regular acini and loosely woven stroma in the PZ5.
Discussion
The LWF values calculated in simulation using both fitting
methods show that the new constrained method is just as accurate as the
original. The correlation between Gleason Score and LWF then reinforces the
idea that multi-echo T2 modelling shows promise as a method for detecting and
grading PCa. The AUC, sensitivity and specificity of the detection of PCa are
all lower in this study when compared to previous results6 but this
may be down to a much lower number of ROIs.
Conclusion
This new method shows promise in being able to significantly
reduce the number of measurements needed to carry out LWI, making it a more
clinically viable method.
Acknowledgements
Cancer Research UK through
the Comprehensive Cancer Imaging Centre, Prostate Cancer UK and the NIHR funded biomedical research centre at UCLH.References
- Whittall K., et al. "In vivo measurement of T2
distributions and water contents in normal human brain." Magnetic
Resonance in Medicine 37, no. 1 (1997): 34-43.
- Sabouri S., et al. “MR Measurement of Luminal
Water in Prostate Gland: Quantitative Correlation Between MRI and Histology.” Journal of Magnetic Resonance Imaging 46,
no. 3 (2017): 861-869
- Sabouri S., et al. “Luminal Water Imaging: A New
MR Imaging T2 Mapping Technique for Prostate Cancer Diagnosis” Radiology 284, no. 2 (2017): 451-459
-
Bjarnason T., et al. “AnalyzeNNLS: Magnetic
resonance multiexponential decay image analysis” Journal of Magnetic Resonance 206, (2010): 200–204
- McNeal J. “Normal Histology of the Prostate” The American Journal of Surgical Pathology 12,
no. 8 (1988):619-633
- Sabouri S., et al. “Luminal water imaging: a
novel MRI method for prostate cancer diagnosis” Proc. Intl. Soc. Mag. Reson. Med. 24 (2016) 2489