Andrew Hahn1, Jeff Kammerman1, Michael Evans2, Wei Zha1, David Mummy1, Robert V. Cadman1, and Sean Fain1,3,4
1Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Biostatistics and Medical Informatics, University of Wisconsin - Madison, Madison, WI, United States, 3Radiology, University of Wisconsin - Madison, Madison, WI, United States, 4Biomedical Engineering, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
To advance hyperpolarized Xenon-129
MRI as a robust tool for probing pulmonary gas-exchange in lung diseases,
establishing repeatability of quantitative ratios and regional metrics is
critical. Repeatability was assessed
using Bland-Altman analysis. Average inter-visit tissue-to-gas, red blood
cell-to-gas (RBC-to-gas) and RBC-to-tissue ratios were highly repeatable; the
primary source of variation was due to lung inflation volume variations,
suggesting consistency of breath-hold coaching and dose delivery is important. Regional distribution of measures was also
qualitatively well-matched between visits.
Demographic factors may also influence these measures; our preliminary
data indicate a significant relationship between RBC-to-tissue ratio and
subject age.
Purpose
Hyperpolarized Xenon-129 (HP-129Xe)
MRI is a safe and increasingly attractive technique for quantification of
pulmonary functional gas-exchange in patients with lung diseases including COPD
and idiopathic pulmonary fibrosis1-3. Here we investigate the long term (~4 weeks) visit-to-visit
repeatability of these imaging based metrics in healthy adult patients.Methods
A cohort of 8 healthy adult
subjects (ages 30-69 yrs, mean 54.0, standard deviation 12.0) underwent HP-129Xe
MRI on a 1.5T scanner (General Electric, Waukesha, WI) using a quadrature vest
coil tuned to the xenon resonant frequency of 17.66 MHz @ 1.5T (Clinical MR
Solutions, Brookfield, WI). Each subject
was imaged with an identical protocol at 2 separate time points, with ~1 month
between visits (range 28-40 days, mean 32.8 days, standard deviation 4.5 days). Dissolved and gaseous phase HP-129Xe images
were acquired with a single ~14 second breath hold scan following exhalation to functional reserve capacity (FRC) then inhalation of
1L of isotopically enriched (85%) 129Xe polarized to ~14-20%, with sequence
parameters as described in Kaushik et al4. Tissue and red blood cell (RBC) components were
separated from the dissolved phase using a 1-point Dixon method5. Regional measures were compared to reference
distributions constructed from healthy normal subjects (Gaussian with a mean±standard deviation) for
ventilation (0.43±0.17),
tissue-to-gas (1.05±0.28)
and RBC-to-gas (0.33±0.18),
and lung voxels were binned based on the number of standard deviations they
were offset from the reference mean6. 3D proton ultrashort echo time (UTE) MRI was
also acquired for the purposes of lung segmentation. The lungs were semi-automatically segmented and
subsequently deformably registered to the HP-129Xe ventilation images. Lung volumes at each time point were
quantified by multiplying the voxel volume by the number of lung voxels.
Bland-Altman plots, coefficient of
variation (CV), and intra-class correlation coefficient (ICC) were used to
investigate the inter-visit repeatability of whole-lung (i.e. mean of lung
voxels) tissue-to-gas, RBC-to-gas and RBC-to-tissue ratios. The distribution of binned voxels within the
lungs was also qualitatively compared visually using histogram plots. Given previous literature describing a
relationship between gas-exchange metrics and lung inflation level2,
we also investigated the correlations between inter-visit differences in lung
volumes and whole-lung tissue-to-gas, RBC-to-gas and RBC-to-tissue ratios. Finally, we explored the dependency of these
ratios on subject age. Results
Bland-Altman and correlation plots
for mean tissue-to-gas, RBC-to-gas and RBC-to-tissue ratios are shown in Figure 1. Mean RBC-to-tissue ratio is highly repeatable
(CV=6.3%, ICC=0.98), while both tissue-to-gas and RBC-to-gas are similarly
slightly less so (Tissue-to-gas: CV=17%, ICC = 0.82, RBC-to-gas: CV=17%,
ICC=0.80). Scatterplots of difference in
lung volumes vs. gas-exchange metrics, shown in Figure 2, demonstrate one potential source of this
discrepancy. Inter-visit changes in both
tissue-to-gas and RBC-to-gas are moderately-to-strongly
correlated with changes in lung volume (Tissue-to-gas: r=-0.78, P=0.022, RBC-to-gas: r=-0.76, P=0.026). No such association is apparent between
RBC-to-tissue and lung volume changes (r=0.08, P=0.854). Histograms of
voxels binned according to normal subject reference distributions at both
visits are shown for the entire cohort in Figure
3, demonstrating good agreement, in general, in the inter-scan regional
distribution of these voxel-wise metrics.
Finally, scatterplots of gas-exchange metrics vs. age are shown in Figure 4. A significant negative correlation was found
between RBC-to-tissue ratio and age (r=-0.71, P=0.048). Correlations
between age and both tissue-to-gas and RBC-to-gas were not significant
(Tissue-to-gas: r=0.12, P=0.773,
RBC-to-gas: r=-0.56, P=0.147). Discussion and Conclusion
These results indicate that common
HP-129Xe imaging based metrics of gas-exchange are highly repeatable
visit-to-visit over the long term (~4 weeks).
While tissue-to-gas and RBC-to-gas ratios are slightly less repeatable
than RBC-to-tissue, it seems likely that this is largely due to inter-visit
variations in lung volumes, which appear to influence the former two ratios,
but not the latter. The associations (or
lack thereof) between these metrics and lung volume can explain the additional
inter-scan variation and are consistent with previous findings based on total
lung spectroscopy2. Histograms of the regional distribution of
these voxel-wise metrics are qualitatively well-matched, in the majority of
cases, further strengthening the case for repeatable measures. Generally, when the distributions are poorly
matched, histogram shape is largely preserved, indicating that a bulk shift in
the distribution (potentially related to lung volume differences) has
occurred. Notably, ventilation
histograms are very repeatable, which might be expected given that they are
less affected by variation in lung volumes.
Finally, the effect of age on RBC-to-tissue highlights the importance of
age matching in subject cohorts. These
results are very encouraging; however, more precise, quantitative, regional
analysis is important to comprehensively justify the repeatability of these
techniques and is the focus of ongoing and future work. Acknowledgements
The authors would like to
acknowledge financial support from NIH/NHLBI R01
HL126771, NIH/NCRR 1 S10 OD016394 (the Pulmonary Imaging Center), and NIH/NCATS
UL1 TR000427 to UW ICTR for funding support.References
1. Qing
K, Ruppert K, Jiang Y, Mata JF, Miller GW, Shim YM, Wang C, Ruset IC, Hersman
FW, Altes TA, Mugler JP. Regional mapping of gas uptake by blood and tissue in
the human lung using hyperpolarized xenon-129 MRI. J Magn Reson Imaging. 2014; 39(2):346–359.
2. Kaushik SS, Freeman MS, Yoon SW, Liljeroth
MG, Stiles JV, Roos JE, Foster WM, Rackley CR, McAdams HP, Driehuys B.
Measuring Diffusion-Limitation with a Perfusion-Limited Gas-Hyperpolarized
129Xe Gas-Transfer Spectroscopy in Patients with Idiopathic Pulmonary Fibrosis.
J Appl Physiol. 2014; 117:577–585.
3. Ebner L, Kammerman J, Driehuys B, Schiebler ML,
Cadman RV, Fain SB. The role of
hyperpolarized 129xenon in MR imaging of pulmonary function. Eur J Radiol 2017; 86:343-352.
4. Kaushik SS, Robertson SH, Freeman MS, He M,
Kelly KT, Roos JE, Rackley CR, Foster WM, McAdams HP, Driehuys B. Single-Breath
Clinical Imaging of Hyperpolarized 129Xe in the Airspaces, Barrier
and Red Blood Cells using an Interleaved 3D Radial 1-point Dixon Acquisition. Magn Reson Med. 2016; 75(4):1434-1443.
5. Kaushik
SS, Freeman MS, Cleveland ZI, Davies J, Stiles J, Virgincar RS, Robertson SH,
He M, Kelly KT, Foster WM, McAdams HP, Driehuys B. Probing the regional distribution of
pulmonary gas exchange through single-breath gas- and dissolved-phase 129Xe MR
imaging. J. Appl. Physiol. 2013;
115:850–860
6. Wang Z, Robertson SH, Wang J, He M, Virgincar
RS, Schrank GM, Bier EA, Rajagopal S, Huang YC, O’Riordan TG, Rackley CR,
McAdams HP, Driehuys B. Quantitative analysis of hyperpolarized 129Xe
gas transfer MRI. Med Phys. 2017;
44(6):2415-2428.