Andras Lindenmaier1,2, Brandon Zanette1,2, Elaine Stirrat1, Brian Kavanagh1,3,4, and Giles Santyr1,2
1Translational Medicine, Hospital for Sick Children, Toronto, ON, Canada, 2Medical Biophysics, University of Toronto, Toronto, ON, Canada, 3Anesthesia, University of Toronto, Toronto, ON, Canada, 4Critical Care Medicine, Hospital for Sick Children, Toronto, ON, Canada
Synopsis
Performing 129Xe morphometry at multiple
pressures allows for direct mapping of microstructural changes and
micromechanical properties of the lung (eg. pseudo-compliance) in both health
and disease. In the present study, the effects of posture and gravity on the
linear portion of lung pseudo-compliance are investigated using this approach
in healthy mechanically-ventilated rats. Spatial gradients in lung pseudo-compliance
due to gravity and posture are shown to vary, consistent with the expectations
that the lungs are denser dorsally due to the weight above. Scaled regional
pseudo-compliance is shown to be in agreement with whole lung compliance
measured using 1H MRI.
Introduction
Hyperpolarized (HP) gas apparent diffusion coefficients
(ADC) and morphometry have been demonstrated for the microstructural characterization
of various lung diseases, both preclinically and clinically.1 Several
studies2,3,4 have used ADC imaging of HP 3He and 129Xe
in rats and humans to infer microstructural changes due to different ventilation
pressure schemes. This tool could be useful for the early detection and monitoring
of treatment of diseases affecting lung micromechanics. In the present study, we
use MRI morphometry to quantify the regional response of the acinar
microstructure, specifically the surface-to-volume ratio (SVR), to changes in
peak inspiratory pressure (PIP) in ventilated healthy rats. In this way, we present
the mapping of pseudo-compliance (CXe), which we believe reflects regional
microelastic and micromechanical properties of the lung.5 We compare
pseudo-compliance maps with whole-lung compliance obtained using conventional
multislice 1H MRI performed concurrently at the same PIPs.Methods
Twelve
week old Sprague-Dawley rats were anesthetized, tracheostomized, and continuously
ventilated in the supine position with air and given four pre-breaths of HP 129Xe
followed by a 13 second breath-hold during which diffusion-weighted MRI was
performed (3T Prisma, Siemens).5 To allow for imaging of the
gravitational gradient, sagittal images were acquired in addition to coronal
images. Six b-value images ranging 0- 30.2 s/cm2 were acquired and
averaged 2-4 times (Gmax = 7.9G/cm, diffusion time of 2.65 ms, TE =
8.25 ms, TR = 16ms, FA = ~4°, FOV = 70mm, and a matrix = 64 x 64, and a partial
echo of 62.5%) to ensure adequate SNR.
This was repeated at PIP of 6, 9 and 12 cmH2O respectively,
to span the linear region of the compliance curve.3 Images were
zero-padded, Hanning, and moving-average filtered using MATLAB 2014a
(MathWorks). Acinar SVR maps were calculated as previously described6
and co-registered using a deformable registration algorithm to account for lung
expansion (imtransform.m), based on the corresponding ventilation images (b =
0) at each PIP. SVR maps were then pixel-wise fitted as a function of PIP to
obtain CXe, defined as the slope of the SVR vs. PIP fits. Only fits
with R2>0.5 were included in the CXe map.Results
Figure
1 shows typical CXe maps for a representative rat in the coronal and
sagittal projection of the whole lung, demonstrating that the spatial
distribution of the pseudo compliance values is quite different along both views.
Regions of increased and decreased SVR with pressure are visible in all maps;
however, in Fig. 1a the dorsal regions show "truly" compliant lungs (ie.
decreased SVR with pressure), while the ventral areas show predominantly
increases in SVR. Maps similar to Fig. 1 were obtained for all animals (n=8;
data not shown). Total lung compliance values scaled by average alveolar
density values (~1.2 cm3/cmH2O) compare favorably with
values measured using ex-vivo microscopy (~0.35-0.90 cm3/cmH2O)7,8
as well as with compliance from the multi-slice 1H data
obtained in this study (0.44 cm3/cmH2O). Discussion
Figure
1 shows regional variability in CXe, which agrees with the high
heterogeneity also seen in CT scans in human and animal studies of lung
regional ventilation and elasticity.9 Furthermore, Fig. 1d also
shows evidence of a gradient along the dorso-ventral axis in the direction of
gravity, as previously established.10 Since CXe is the
change in SVR as a function of PIP, CXe is affected by two major
factors: expansion of terminal airways resulting in a decrease in SVR with
increasing pressure (showing compliance), and alveolar recruitment resulting in
an increase in SVR with increasing pressure. Areas with positive CXe
have no distensibility while those with negative CXe have normal
distensibility. Regions of null CXe are present where there is no
change in the SVR with pressure (mostly seen in the airways). While the dorsal
regions are measuring "true" lung compliance, the peripheral areas
may be suffering from atelectasis (closed alveoli) and hence biasing the
compliance characterization; changing its linearity and posing a limitation to
the presented method. Atelectasis is commonly seen in the peripheral parts of
the lung and can be recruited with increased pressure, thus potentially
explaining an increased SVR and positive pseudo-compliance in these regions. This
effect may also explain whole-lung compliance differences derived from CXe
as compared to 1H. Positive end-expiratory pressure maneuvers are
often used to re-recruit alveoli3, and this process could potentially
be tracked by mapping CXe to avoid over-distention or injury to the
lungs. This new tool provides regional compliance information that can potentially
be very useful in early detection and guiding treatment for patients suffering
from lung injury, where compliance is known to change.11Acknowledgements
The
authors are grateful to N. Kanhere, Y. Friedlander and F. Morgado for 129Xe gas polarization and help with experiments, to
Dr. M. Couch for helpful discussion on pulse-programming and software
development and to the funding sources: CIHR operating grant (MOP 123431) and
NSERC Discovery grant (RGPIN 217015-2013). AL was supported by an NSERC PGSD
award.References
1. Yablonskiy, et al. Diffusion lung imaging with
hyperpolarized gas MRI. NMR Biomed 2015; doi: 10.1002/nbm.3448.
2. Chan HF, et al. Exploring Lung Inflation
Mechanisms with 3D 3He and 129Xe Whole Lung Morphometry Mapping. Proc. of The
25th International Society for Magnetic Resonance in Medicine Annual Meeting.
Honolulu, USA: poster # 3315; 2017.
3. Cereda M, et al. Positive End-expiratory Pressure
Increments during Anesthesia in Normal Lung Result in Hysteresis and Greater
Numbers of Smaller Aerated Airspaces. Anesthesiology 2013; 119(6): 1402–1409.
4. Xin Y, et al. Hyperpolarized gas diffusion MRI of
biphasic lung inflation in short- and long-term emphysema models. Am J Physiol
Lung Cell Mol Physiol 2017; 313: L305–L312.
5. Lindenmaier AA, et al. Pseudo-Compliance Mapping of
the Rat Lung Using Hyperpolarized 129Xe Morphometry at Varying Pressures. Proc.
of The 2017 International Workshop on Pulmonary Imaging. Philadelphia, USA:
University of Pennsylvania; 2017.
6. Sukstanskii AL, et al. Lung morphometry with
hyperpolarized 129Xe: theoretical background. Magnetic Resonance in Medicine
2012; 67(3): 856-866.
7. Fox MS, et al. Comparison of hyperpolarized He and Xe
MRI for the measurement of absolute ventilated lung volume in rats. Magnetic Resonance
in Medicine 2014; 71: 1130–1136.
8. Lai YL, et al. Respiratory mechanics in the
anesthetized rat. Journal of Applied Physiology 1978; 45: 255–260.
9. I. Galvin, et al. Distribution of blood flow and
ventilation in the lung: gravity is not the only factor. British Journal of
Anaesthesia 2007; 98(4): 420–8.
10. Kaneko K, et al. Regional distribution of ventilation
and perfusion as a function of body position. Journal of Applied Physiology 1966;
21: 767–777.
11. Kuethe DO, et al. Magnetic resonance imaging provides
sensitive in vivo assessment of experimental ventilator-induced lung injury. American
Journal of Physiology-Lung Cellular and Molecular Physiology 2016; 311:
L208–L218.