YunKyoung Ko1, Chang-Hoon Choi1, and N.Jon Shah1
1Institute of Neuroscience and Medicine - 4, Juelich, Germany
Synopsis
A crossbar switch matrix can be flexibly connected to
any input and output paths via RF switches. However, this type of matrix
creates open-stubs in the RF lines. Since the RF wavelength becomes shorter as
the magnetic field increases, signal loss due to impedance variations in RF
pathways becomes severe, thus degrading image quality. In this study, we
propose an advanced compensation method and verify its performance in
ultra-high field MRI with single and multi-channel array coils.
Purpose
RF receive channels are the path between the RF coil
and the receivers on the MRI spectrometer. Ordinarily, the number of receive coil connectors and receive
paths are different; however, recently the channels of receive coils
have been increased to correspond to the number of receive paths, including
analogue-to-digital converters (ADCs). Therefore, modern MRI scanners need, for
example, a switch matrix to automatically connect receive coils with their appropriate
receive paths. The most commonly used matrix structure is a crossbar type
switch matrix,1 where a crossbar switch matrix can be flexibly
connected to any input and output paths via RF switches. However, this type of matrix
creates open-stubs in the RF lines. Since the RF wavelength becomes shorter as
the magnetic field increases, signal loss due to impedance variations in RF
pathways becomes severe, thus degrading image quality.2,3 In this study,
we propose an advanced compensation method and verify its performance in
ultra-high field MRI with single and multi-channel array coils.Methods
Fig.
1(a) shows a cross-bar-type switch matrix, including the compensation circuits
on the edge of the RF signal path. To evaluate the effect of the open-stub and
compensation circuit, a 4x4 switch matrix was constructed with a double-layer
FR4 PCB. The dimensions of the matrix board were 300 mm × 210 mm as shown in
Fig. 1(b). When the length of the open-stub is close to λ/4, and the signal
loss is large, this compensation circuit switches to a specified capacitor
value. In a
previous study, the compensation circuitry was only present on the opposite
side of the input stage.2,3 However, in ultra high-field MRI, the compensation
circuitry is also required on the opposite side of the output, thus, as shown
in Fig. 1(b), the compensation circuitry is located on both the opposite side
of the input port and output port. To measure signal loss according to the stub length,
the RF path loss was measured using a network analyser. MR images were acquired
using an in-house assembled 9.4 T animal MRI scanner4 with a gradient
echo sequence (TR = 40 ms, TE = 3.85 ms, averages = 2, acquisition time = 20
seconds, resolution = 100 µm x 100 µm, and slice thickness = 1 mm). To
calculate the SNRs, noise images were separately acquired using the identical protocol
with a zero transmit power. A home-built quadrature birdcage coil and a commercial
4ch Rx only coil (RAPID, Germany) were used to evaluate the switch matrixResults
Fig. 2 shows the measured signal path loss of each
channel of the switch matrix at 9.4 T, with and without the compensation
capacitor (91 pF). Channel 1 was used as a reference. The open stub length of
channels 2 and 3 were close to λ/4, and the signal losses were about 14 dB and
11 dB, respectively. By the application of compensation, the signal losses were significantly
reduced. Phantom images acquired using a single-channel coil are shown in Fig.
3 and indicate a significant signal loss in channel 2 and channel 3. However,
by using the compensation capacitors, the losses were minimised. As shown in
Fig. 4, multi-channel images were acquired using a 4-channel Rx only coil. SNR
was degraded by around 34% without compensation, compared to that of the
reference. It was also shown that SNR was recovered using the compensation
circuit.Discussion
We have demonstrated the performance of the proposed
switch matrix using either a single-channel coil or multi-channel coils with an
ultra-high field MRI scanner. Due to the shorter wavelength at ultra-high field,
it was necessary to compensate the signal loss both in vertical and in horizontal
paths, and we have confirmed that this compensation method is effective.Acknowledgements
No acknowledgement found.References
[1] Kröckel
H. Switching matrix with two control inputs at each switching element. Google
Patents; 2010.
[2] Lee H. Reconfigurable multi-channel RF crossbar
switch matrix for MRI receiver front-end. Electron Lett. 2017;53(6):380-1.
[3] Ko Y-K, Lee HL, Choi C-H, Shah NJ. A
technique to compensate signal loss in an RF switch matrix system in MRI. In
Proceeding of the 25th Annual Meeting of ISMRM, Honolulu,USA; 2017.
[4] Choi C-H, Ha Y, Veeraiah P, Felder J, Möllenhoff
K, Shah NJ. Design and implementation of a simple multinuclear MRI system for
ultra high-field imaging of animals. J
Magn Reson. 2016;273:28-32.