Shanshan shan1, Mingyan Li1, Yaohui Wang2, Fangfang Tang1, Deming Wang1, Haiwei Chen1, Ewald Weber1, Rafael Franco1, Craig Freakley 1, Feng Liu1, and Stuart Crozier1
1The University of Queensland, Brisbane, Australia, 2South China University of Technology, Guangzhou, China
Synopsis
MRI-guided radiotherapy requires precise
image geometric information to target a tumour without unnecessary radiation on
healthy surrounding tissue. Due to the imperfections in the gradient system and
engineering limitations, however, gradient non-linearity (GNL) inevitably
occurs and causes image distortions if not properly accounted for. Here we
propose a novel method to estimate the gradient field using stream function
methods with a grid phantom. The estimated gradient field was then used for GNL
distortion correction and image reconstruction. Initial simulations demonstrated
that the image geometric distortion in a combined MRI and linear accelerator (MRI-Linac)
system was effectively improved by the proposed method.
Purpose
MRI-Linacs provide image-guided, real-time treatment for
cancer patients 1. When a split magnet is used, it is very challenging
to generate perfectly linear gradient fields with a split gradient coil system 2. Phantom-based image domain correction methods 3 employ a fitting procedure to extract gradient
non-linearity (GNL) information, but with a limited number of marker points,
the estimation accuracy may be not guaranteed near the imaging boundary area. Electromagnetic
(EM) modelling can be used to estimate the gradient field, but the numerically
calculated gradient fields do not always match the measured values. In this
work, to improve the accuracy of estimating nonlinear gradient fields, a hybrid
method that uses a stream function 4,5 and grid phantom-based measurement 6 is developed. The image distortion caused by GNL in
MRI-Linac system can then be effectively corrected.Methods
The proposed method corrects GNL in three
steps. Step 1: The gradient field profile in the imaging region is acquired via
using the grid phantom; Step 2: by means of a stream function method, the
current density distribution on a 40 cm diameter sphere is calculated to
produce gradient field profile that matches phantom images acquired in
step 1. Step 3: with the determined current density distribution, the
non-linear gradient fields at any position in the imaging volume can be
calculated and used for GNL correction. The kernel of the proposed method is to
solve the following optimisation problem:$$ {\mathop{\arg\min}_{\ I_{n\_k}} (\| B(I_{n\_k},r)-B_{k\_nominal}\|_{2}+\alpha W_{magn} \tag{1}}$$where the magnetic field $$${B(I_{n\_k},r)}$$$
at position r was calculated as$$${B(I_{n\_k},r)\approx \frac{\mu_{0}}{4\pi}\sum_{n=1}^N I_{n\_k}\int_{s'} \triangledown \times\frac{f_{n} (r')}{\mid r-r'\mid}ds'}$$$.$$${f_{n} (r')}$$$ denotes the circulating current on the surface
of the spherical volume, which is discretised with ‘’elements’’ and N
‘’nodes’’ shown in Figure 1.$$$ I_{n\_k}$$$is the coefficient of
the stream function for each node n.$$${\alpha}$$$ is the weight of the magnetic energy, $$${W_{magn}}$$$. According
to the phantom mapping from the nominal position$$${r\epsilon C}$$$to the actual position h(r), the magnetic field in measured marker position is
calculated as$$${B_{k\_nominal}(r)=G_{k}*h(r)}$$$,
where$$$G_{k}$$$ is the k-axis gradient. Once the surface
currents on three axes are calculated (shown in Fig. 2) the gradient field at
any location inside the 30 cm DSV (Diameter of Spherical Volume) can be
calculated for GNL correction (see abstract 5644 of ISMRM 2018 ). Then the
undistorted image m can be obtained by
solving the following optimization problem:$${\parallel E_{GNL}\cdot Fm-b\parallel_{2} \tag{2}}$$where$$$E_{GNL}$$$
denotes
the GNL field encoding, F is the
non-distorted Fourier encoding matrix and b
is the acquired k-space signal. In this paper, the gradient field and k-space
data was simulated according to Liu’s design 2. The image size is Nx = Ny = Nz =
256, voxel size is 1mm isotropic.Results
Phantom images before and after GNL
correction are shown in Fig. 2 (a-c) and (d-f), respectively. In the top row,
geometric distortion is clearly visible before correction, especially at the
outer regions denoted with red arrows. However, after applying the proposed
method, the geometric distortion caused by GNL is successfully corrected as
demonstrated by images (d) to (e) in Figure 2. In Table 1, the root-mean-square-error (RMSE)
and the maximum error of the marker coordinates are compared before and after
GNL correction. Before correction, the RMSE of marker locations in x, y and z
directions was 1.7mm, 1.8mm, and 1.3mm, respectively; with 2.9mm in the
positional displacement (r). The RMSE of marker locations in GNL-corrected
images decreased significantly to less than 0.15mm. The maximum errors of
marker coordinates in x, y and z directions before correction were 5.5mm, 6.1mm,
3.4mm, respectively; with a 8.4mm positional displacement. In comparison, the
maximum errors of marker coordinates after GNL correction decreased to 0.5 mm
or less, meeting the geometric accuracy requirement of radiotherapy 7.Discussion
Here, the gradient fields and related
phantom images were numerically calculated according to our previously designed
split gradient coils for MRI-Linac system.
The actual gradient fields may have minor deviations from the simulation
results. In addition, the gradient field profiles obtained from numerical
modelling are usually smoother than those taken from measurement, thus a
filtering effect is realized during the implementation of the proposed hybrid
method. The practicality of the proposed method will be further validated with
experimental data. Conclusion
In this proof-of-concept work, a
hybrid method combining both phantom-based mapping and stream function was
proposed for gradient field estimation and GNL-correction in MRI-Linac system. The
preliminary simulation results demonstrated that the proposed method can correct
GNL-induced image distortions efficiently. In the future, we will verify the
proposed method by experiment.Acknowledgements
No acknowledgement found.References
1. Metcalfe
P, Liney G, Holloway L, et al. The potential for an enhanced role for MRI in
radiation-therapy treatment planning. Technology in cancer research &
treatment 2013;12:429-446.
2. Liu L, Sanchez-Lopez H, Liu F,
Crozier S. Flanged-edge transverse gradient coil design for a hybrid LINAC-MRI
system. J Magn Reson 2013;226:70-78.
3. Tao S, Trzasko JD, Gunter JL, et al.
Gradient nonlinearity calibration and correction for a compact, asymmetric
magnetic resonance imaging gradient system. Phys Med Biol 2017;62:N18-N31.
4. Lemdiasov RA, Ludwig R. A stream
function method for gradient coil design. Concepts in Magnetic Resonance Part
B: Magnetic Resonance Engineering 2005;26B:67-80.
5. Poole M, Bowtell R. Novel gradient
coils designed using a boundary element method. Concepts in Magnetic Resonance
Part B: Magnetic Resonance Engineering 2007;31B:162-175.
6. Wang D, Doddrell DM, Cowin G. A
novel phantom and method for comprehensive 3-dimensional measurement and
correction of geometric distortion in magnetic resonance imaging. Magn Reson
Imaging 2004;22:529-542.
7. Thwaites D. Accuracy required and
achievable in radiotherapy dosimetry: have modern technology and techniques
changed our views?, In Journal of Physics: Conference Series, IOP Publishing,
2013.