Lucia I Navarro de Lara1, Laleh Golestani Rad1, Sergey Makarov1,2, Jason P Stockmann1, Lawrence L Wald1, Thomas Witzel1, and Aapo Nummenmaa1
1Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Worcester Polytechnic Institute, Worcester, MA, United States
Synopsis
The multichannel TMS approach
enables steering the stimulation “hot spot” electronically making the execution
of concurrent TMS and fMRI experiments potentially more flexible. An integrated
multichannel TMS/MRI system is currently under development and here we
investigate how TMS coils influences the B1+ field of a
birdcage coil. We compare detailed electromagnetic simulations with
empirical data using a single three-axis TMS coil element as well as generate
simulated data for 4x4 TMS coil array. Both simulations and empirical results
indicate that an array of TMS coils should not distort the B1+
significantly at distances bigger than 2cm from the coil.
Purpose
To develop a new type of multichannel TMS/MRI stimulation/imaging array
for 3T, we evaluated the effects of the TMS coil elements on the B1+
of a birdcage coil. Introduction
The combination of non-invasive neuromodulation and
simultaneous imaging techniques constitutes a powerful approach for
neuroscience and potentially treatments. For example, Transcranial Magnetic
Stimulation (TMS) during fMRI could help to evaluate
causal relationships between the activations of different cortical regions. This combination was presented for the first time
by Bohning et al. 20 years ago (1). Dedicated coil arrays have been developed
to increase the sensitivity of the MR imaging (2) for these experiments. However,
to enable even broader range of potential applications, a multichannel TMS/MRI
head coil array for 3 T utilizing a novel three-axis TMS coil design is
currently under development (see Fig.1A).
When bringing multiple TMS coils inside the MRI
scanner it is important to assess the influence of these conducting elements on
the RF excitation. Here, we studied the effects of such system on the transmit
field based on EM simulations. In addition, a prototype TMS coil element has
been fabricated (TristanTechnologies, San Diego, US) and it was used to compare
the simulations with experimental data (see Fig.1B). Methods
A Skyra body
coil (Siemens, Erlangen, Germany) was modeled in HFFS (Ansys, USA) and tuned
and matched with a gel phantom as load. The dimensions of the phantom
were 60cmx41cmx6cm. Three models were built: (i) the birdcage coil loaded only
with the phantom, (ii) the coil loaded with the phantom and a three-axis TMS
coil element placed over the center of the phantom, and (iii) the coil loaded
with the phantom and 16 three-axis TMS elements distributed in a cartesian grid
over the center of the phantom (see Fig.2).
To compare the
simulations with experimental data, B1+ measurements using
the Skyra body coil were done. As load, an in-house built phantom
(0.82% NaCl, 17% Agar-Agar) with the same dimensions was used. Two setups were employed:
(i) only the phantom and (ii) the phantom and one TMS prototype over the center
and placed in the isocenter. Gradient Echo (GRE) images with 6 different flip
angles (FA) (TR=3000ms, TE=10ms, FA=10,20,30,50,70 and 90, 50 slices,
3mmx3mmx3mm, MA=96x54) were acquired for each setup to calculate B1+
maps. The measured data was fitted pixel-wise to B1*sin(FA) to
calculate the field. The reference voltage was set to be 300 V. Results
Fig3A and Fig3B show the B1+ simulated over the same
field of view as the measurement. Fig3C
and Fig3D show the corresponding results of the calculated B1+
field based on the measured data.
Profiles of the B1+
at 3 different points on the phantom, for both the simulations and the
measurements with and without the coil are shown in Fig4. The table on the figure summarizes the changes of the
B1+ when the coil is over the phantom at 3 different
depths, for both simulated and measured fields.
The simulated B1+ obtained for the multichannel
TMS array is shown in Fig5A. For the sake of a clear comparison, Fig5B shows
the simulated B1+ for the no coil case. The table on the
figure summarizes the changes of simulated B1+ when the
array is over the phantom at 3 different depths, for 7 distinct positions.Discussion
The simulated and measured B1+ fields show in general similar
characteristics. Both simulations and measurements show enhancement and decrease
of the field close to the TMS coil.
The profiles show that B1+
enhancement at 4.2cm based on the simulation (between 4-12%) is slightly higher
than the measured data (1-5%). This difference is larger for closer regions to
the coil. The reason for this is the potential susceptibility effects due to
presence of the housing.
For the multichannel TMS
case, the results show an enhancement of the B1+ at the
depth of 4.2cm between 7-19%, and for closer regions, between 15-25%. On the
other hand, the field homogeneity at distances greater than 2cm is not significantly
affected. No drastic artifacts are expected based on either the simulated or
the experimental data.
Our conclusion from this
study is that an additional transmit coil will not be required even for a
large-scale multichannel TMS array positioned inside the MRI environment.
However, if using a standard body transmit coil the enhancement of the B1+
must be considered. Further safety assessment must be done and also the
effects of the TMS cables should be included in future analysis. The framework
presented here may also be useful for other MRI compatible TMS coil design
considerations.Acknowledgements
This work was funded by NIH
R00EB015445, R01MH111829 and NIH K99EB021349.
We want to thank Dr. Douglas Paulson, Mr. Kevin Pratt, Mr. Paul Mascarenas, and
Mr. Paul Miller from Tristan Technologies for their contribution in fabricating
the coil. We also thank Dr. Yoshio Okada from Moment Technologies and Dr. Bastien
Guerin from MGH Martinos Center for valuable discussions on the project. References
(1) Bohning et al.,Invest Radiol,33(6):336-340,1998
(2) Navarro de Lara et al., Magn Reson Med, 74(5):1492-1501,2015