Xin Chen1, Xin Li2, Danielle Kara3, Mingdong Fan3, Joseph Rispoli2, Michael Steckner1, and Robert Brown3
1Toshiba Medical Research Institute USA, Mayfield Village, OH, United States, 2Purdue University, West Lafayette, IN, United States, 3Case Western Reserve University, Cleveland, OH, United States
Synopsis
B1+
inhomogeneity is a challenge for breast imaging and B1 shimming is standard for
all 3T clinical MRI applications. While B1+ homogeneity is typically better at 1.5T
than 3T due to reduced wavelength effects, breast image quality is still
challenged by B1+ inhomogeneity at 1.5T. This work shows B1 shimming can significantly
improve B1+ homogeneity for breast imaging at 1.5T as compared to conventional
quadrature drive.
Introduction
B1+ homogeneity is an important
image quality factor in breast imaging, both to ensure optimal image quality
and to facilitate comparisons. The number of breast MRI procedures is growing
rapidly and it is important to investigate mechanisms that improves image
quality. B1+ inhomogeneity in breast imaging has been evaluated on quadrature
driven 1.5T and 3T clinical MR systems, and showed similar inhomogeneity patterns.
Quantitative measurements at both field strengths suggest inhomogeneity at 1.5T
only slightly less severe than 3T.1 While B1 shimming has been
widely applied at 3T and has provided improved B1+ homogeneity for breast imaging,1,2
application to 1.5T has not received as much consideration because, in general,
B1+ homogeneity is considered sufficient at 1.5T. This work shows B1 shimming
can improve B1+ homogeneity for breast imaging at 1.5T as compared to
conventional quadrature drive (QD).Methods
Numerical FDTD EM modeling was
performed with a generic whole body Tx coil based on ISO/TS109743 (16-rung
high-pass birdcage, diameter 750mm, length 650mm. RF shield diameter 790mm, length
850mm4,5) with software package Sim4Life (v3.4, ZMT, Zurich,
Switzerland).The coil was tuned to 64MHz, and driven with two voltage sources
located on one end ring. Female human
models Hanako6 (1.60m, 53kg) and Ella7 (1.63m, 58.7kg)
were studied respectively. The human models were positioned prone, feet first
in the Tx coil (Figure 1). Anatomically accurate breast enhancements8
were added to the models’ chest to simulate a realistic prone breast imaging
scenario. The human models were elevated by a typical breast coil height so
that the breasts were at isocenter. Ella was also posed to represent the
abdominal curvature over a breast coil. Simulations were performed by varying the
phase and amplitude of one voltage source to achieve B1 shimming. Transmit B1
field (B1+) was analyzed in the central axial slice (Z=0) across the breasts.Results
The coefficient of variation (CV) of
the modeled B1+ was calculated for each B1 shim setting over each individual
breast as well as the entire breasts. Additionally, average B1+ was calculated
over left and right breast respectively, and left/right ratio was calculated to
quantitatively measure B1+ asymmetry between two breasts. The results are shown
in Tables 1 and 2 for Ella and Hanako, respectively. QD is used as reference
for comparison (B1 shim setting #1 in both tables). All power gain settings in
Tables 1 and 2 are only 0dB (which means input power are equal between two
feeding sources) because no significant improvements were otherwise
realized. Figures 2 and 3 show B1+ maps
in the central axial slice (Z = 0) with B1 shim settings corresponding to entries
in Tables 1 and 2 for Ella and Hanako respectively. Discussion
Despite of differences of body
habitus and posture between Ella and Hanako (in particular the differences of
breasts), modeling results showed that increasing the phase difference between
two feeding sources can 1) improve B1+ homogeneity in each individual breast,
2) reduce B1+ difference between two breasts, and 3) improve overall breast B1+
homogeneity. Furthermore, the phase difference between two feeding sources had
a larger impact on B1+ than amplitude difference, as for a given phase varying
the amplitude of one source did not show significant impact on B1+.Conclusion
Modeling with anatomically accurate
and detailed female human and breast models showed that B1+ field inhomogeneity
and asymmetry presently challenging 1.5T breast imaging can be significantly reduced
by B1 shimming using two feeding sources.Acknowledgements
We would like to acknowledge Dr.
Susan Hagness’s group at University of Wisconsin-Madison for the breast models.References
-
Trop I, Gilbert G, Ivancevic M, et al. Breast MR
Imaging at 3 T with Dual-Source Radiofrequency Transmission Offers Superior B1
Homogeneity: An Intraindividual Comparison with Breast MR Imaging at 1.5 T. Radiology. 2013; 267(2):602‐608.
- Winker S, Rutt B. Practical methods for improved
B1+ homogeneity in 3T breast imaging. Proc.
Intl. Soc. Mag. Reson. Med. 21. 2013: 397.
- ISO/TS 10974:2012 Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device. Geneva, Switzerland: International Organization for Standardization, 2012.
- Murbach M, Neufeld E, Kainz W, et al. Whole-body
and local RF absorption in human models as a function of anatomy and position
within 1.5T MR body coil. Magn. Reson.
Med. 2014; 71(2):839-845.
- Murbach M, Cabot E, Neufeld E, et al. Local SAR
enhancements in anatomically correct children and adult models as a function of
position within 1.5 T MR body coil. Prog
Biophys Mol Biol. 2011; 107(3):428-433.
- Nagaoka T, Watanabe S, Sakurai K, et al.
Development of Realistic High-Resolution Whole-Body Voxel Models of Japanese
Adult Male and Female of Average Height and Weight, and Application of Models
to Radio-Frequency Electromagnetic-Field Dosimetry. Physics in Medicine and Biology. 2004; Vol.49: 1-15.
- Gosselin MC, Neufeld E, Moser H, et al. Development
of a new generation of high-resolution anatomical models for medical device
evaluation: the Virtual Population 3.0. Physics
in Medicine and Biology. 2014; 59(18): 5287-5303.
- Li X, Rispoli J. Thermal and Specific Absorption
Rate Simulation of Heterogeneous Breast Models in The Prone Position Fused with
Two Human Whole-Body Models. ISMRM Workshop on Ensuring RF Safety in MRI,
McLean, VA, 2017.