Seung-Kyun Lee1,2 and Sukhoon Oh3
1Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Republic of Korea, 2IBS Center for Neuroscience Imaging Research, Suwon, Republic of Korea, 3Bioimaging Research Team, Korea Basic Science Institute, Cheongju, Republic of Korea
Synopsis
We present theory
and experimental validation of reconstruction of all three components
of RF vector magnetic field (B1 vector) in MRI. Using eight-orientation B1+
magnitude mapping, the magnitudes and relative phases of all Cartesian
components of B1 vector can be unambiguously determined. The method is
demonstrated on a phantom in a birdcage transmit coil with a standard B1+
mapping sequence at 3T. Our method can be used to validate assumptions in MR-based
electrical properties tomography (MR-EPT) and specific absorption rate (SAR) modeling.
Introduction
Increased tissue-RF
field interactions at ultra-high-field MRI presents challenges in two ways:
image shading and tissue heating. While the former involves only one (proton
co-rotating) component of the RF field ($$$B_1^+$$$), heating is contributed by all
three components of the RF vector magnetic field (B1 vector) through induction
of an electric field. Experimental measurement of the B1 vector field is therefore
relevant to modeling specific absorption rate (SAR) in high-field MRI. Also,
MR-based electrical properties tomography (MR-EPT)1,2 could benefit
from additional B1 field information. While conventional B1 mapping only
measures $$$B_1^+$$$, Florian et al.3 noted in 2006 that other B1 components
could be accessed by multi-orientation measurements. Here, we present a theory of
B1 vector field mapping from multi-orientation scans, and provide experimental
validation through phantom measurements and RF field simulations.Theory
We will assume that
the RF transmitter is small enough compared to the scanner’s RF shield that the
B1 field of interest is not significantly affected by the shield. Accordingly,
we define an “RF system” as the transmitter and the object being scanned with
their relative positions fixed. We aim at finding the B1 components $$$(B_x, B_y,
B_z)$$$ in the RF system before rotation.
Conventional $$$B_1^+$$$ mapping only measures $$$B_{1,orig}^+
= |B_x – iB_y|/2$$$. If we now rotate the RF system by ϑ
around an axis $$$\hat{n} = (n_x,n_y,n_z)$$$, the B1 vector in the magnet’s reference frame becomes $$$\vec{B}_{1new} = R(\hat{n},\theta)
\vec{B}_{1orig}$$$, where $$$R$$$ is a rotation matrix. The corresponding new,
measureable $$$B_1^+$$$ magnitude is given by
$$ |B_{1,new}^+|=|(\cos\theta-in_z\sin\theta)(B_x-iB_y) +B_z(n_y+in_x)\sin\theta + \vec{B}\cdot\hat{n}(n_x-in_y)(1-\cos\theta)|.$$ Multiple-orientation measurements of the above quantity can allow
disentangling of magnitudes and relative phases of $$$(B_x, B_y, B_z)$$$. Many multi-orientation
strategies can be devised, balancing mathematical complexity and experimental practicality.
Here we present an 8-orientation method which involves large angles but
allows mathematically straightforward solution. The relevant equations are
summarized in Tables 1-2.Methods
Experiment
$$$B_1^+$$$ maps on a
cylindrical phantom were acquired in a 3T whole-body scanner (Magnetom Prisma,
Siemens Healthineers, Germany) with a vendor-provided extremity (knee) coil,
consisting of 1 quadrature (birdcage) transmit coil and 15 receive elements. A
vendor-provided fast $$$B_1^+$$$ mapping sequence4 was applied to each of
the eight orientations to produce a set of 2D $$$B_1^+$$$ maps on the phantom’s axial
plane. During prescription, scan planes were rotated to follow the phantom
orientation. The phantom was a cylinder with length (flat part) = 42 cm and
diameter = 14.4 cm, consisting of aqueous solution of NaCl and NiSO4,
as provided by the vendor. The acquired $$$B_1^+$$$ maps were processed according to
Tables 1-2 to calculate the magnitudes and relative phases of the RF-frame
components of B1.
Simulation
The RF
fields were simulated in XFdtd v7.6 (Remcom, USA) on the following model:
high-pass birdcage coil with 12 rungs, diameter/length = 20/20 cm, frequency =
123.25 MHz; an axially located cylindrical phantom with σ=1 S/m, εr=78,
length/diameter = 42/14.4 cm. The effect of an RF
shield was tested and found to affect B1 field by less than 4% at tested
orientations.
Results
Figure 2 shows the magnitudes
of $$$(B_x,B_y,B_z)$$$
reconstructed from experimental data in
comparison with simulation. The successive slices are 18 mm apart. The agreement
is generally good. The results on the middle slice also agreed well with
analytical solutions for an infinitely long cylinder5. Small
differences between the experimental and simulated data can be partly attributed to asymmetric
positioning of the phantom inside the coil, which was visible in localizer
scans (not shown). Our data show the spatial inhomogeneity of the linear
components of the transverse B1 field at 3T, and how Bz grows towards the end
of the birdcage coil. In the relative phase maps (Fig. 3), the phase of By is
about π/2
ahead of Bx, consistent with left circular polarization. Bz phase is sensitive
to experimental (positioning) errors due to small Bz magnitudes. Measured
relative phases involving Bz resemble simulation better on the end slices where
Bz is stronger.Discussion
To our knowledge
this is the first experimental demonstration of RF vector field mapping (up to
a common phase) in MRI. Our method could inform MR-EPT methodology and SAR
modeling. Extension of the method to in-vivo studies obviously requires
reduction of the rotation angles as well as streamlining of multi-orientation scan
workflow. Given existing literature on multi-orientation human scans for susceptibility6
and $$$B_1^-$$$ mapping7, we do not
think this is an insurmountable problem. Future research directions include development
of human study-compatible scan protocols and absolute phase determination through
additional phase mapping or theoretical constraints.Acknowledgements
This work
was supported by IBS-R015-D1.References
(1) Zhang
X. et al., IEEE Rev Biomed Engineering 7:87-96 (2014)
(2)
Katscher U. et al., NMR in Biomed (2017) DOI: 10.1002/nbm.3729
(3) Wiesinger F. et al., Prospects
of absolute B1 calibration, in Unsolved Problems and Unmet Needs in MR, 14th
Annual Meeting of ISMRM (2006).
(4) Fautz H.P. et al., B1 mapping of coil arrays for parallel
transmission, 16th Annual
Meeting of ISMRM (2008), Abstract 1247.
(5)
Glover G.H. et al., J Magn Res 64:255-270 (1985)
(6)
Liu T. et al., Magn Res Med 61:196-204 (2009)
(7) Lee S-K. and Dixon, W.T., 19th
Annual Meeting of ISMRM (2011), Abstract 4429.