Omer F. Oran1, L. Martyn Klassen1, Hacene Serrai1, and Ravi S. Menon1,2
1Centre for Functional and Metabolic Mapping, The University of Western Ontario, London, ON, Canada, 2Department of Medical Biophysics, The University of Western Ontario, London, ON, Canada
Synopsis
Mapping B1+
magnitude fields for any arbitrary RF-configuration of interest is desirable
and often required in parallel multi-transmit systems. Conventional mapping
methods work by taking either the magnitude ratio or the phase difference of
MRI images, and thus they fail in regions of low MRI-signal which can easily
occur for an arbitrary RF-configuration. We propose a new BSS based method
which always uses the CP-mode for excitation while the RF-configuration of
interest is used only for the off-resonance RF-pulse. Through this approach, the
proposed method works even if the RF-configuration of interest has very-low
magnitudes within the imaging volume.
Introduction
Parallel-transmit
(pTx) systems consisting of RF-coil arrays bring in the possibility of
tailoring $$$B_1^+$$$ field by configuring the magnitude and phase of each coil element
(RF-configuration). These systems can be used for mitigating $$$B_1^+$$$ inhomogeneity and for
accelerating multidimensional excitation RF pulses.
For any arbitrary pTx RF-configuration, $$$B_1^+$$$ field
distribution is required for $$$B_1^+$$$ shimming, design and development of pTx coils,
and SAR estimation. Several $$$B_1^+$$$ magnitude mapping methods have been previously
developed1,2,3 However, since these methods work by taking either the magnitude
ratio or the phase difference of MRI images, they suffer in regions of low
MRI-signal which can easily occur for an arbitrary RF-configuration. In this
study, we propose a Bloch-Siegert Shift (BSS) based method which is capable of
measuring $$$B_1^+$$$ field even in the mentioned null regions.Theory
In
traditional BSS method, an off-resonance RF pulse (OffRFP) ($$$\Delta f>2$$$kHz) is applied right after the excitation RF pulse
(ExcRFP) and then the signal is read out using a GRE sequence. The OffRFP does
not excite spins but, considering the effective-B1 field theory, it acts as an additional
B-field applied in the z-direction which causes phase shift in the transverse
spins given by3 $$\phi _{BSS} =\int_{0}^{T}\frac{(\gamma B_1(t))^2}{2\Delta f}dt$$ where $$$B_1(t)$$$, $$$\Delta f$$$, and $$$T$$$ are
the envelope, frequency shift and the duration of the OffRFP. To calculate $$$\phi _{BSS}$$$, the phase difference is taken between two
scans acquired with positive and negative frequency shifts.
As mentioned
above, the BSS method fails in the null regions of arbitrary RF-configurations due
to the inaccuracy of the phase in these regions. To overcome this problem,
we propose the use of two different RF-configurations for the ExcRFP and
OffRFP: For the excitation RF pulse, we use only the CP-mode of the coil, obtained by a one-time optimization of the $$$B_1^+$$$ homogeneity for an average head, which gives high-signal in all regions. As
for the OffRFP, we use the RF-configuration of interest for which we want to map $$$B_1^+$$$ field.
Methods
All
scans were conducted on a Siemens 7T MRI scanner (Erlangen, Germany) equipped
with a pTx system of 8 transmit and 32 receive channels. The manufacturer’s GRE
sequence is modified to fulfill the BSS pulse sequence. Axial 2D brain images
and corresponding $$$B_1^+$$$ maps are acquired with the following parameters for two
volunteers: TR 600 ms, TE 12 ms, FOV 240x240(128x128) mm, flip angle 60 degree, OffRFP with a Fermi shape ($$$\Delta f=4$$$kHz, $$$T=8$$$ms, nominal peak 5.8uT). Discrete
Fourier transform matrix coefficients were used to obtain 8 different
RF-configurations4. For Subject 1, these configurations were applied
both to the ExcRFP and OffRFP (traditional BSS method). For Subject 2, the
configurations were applied only to the OffRFP and, for ExcRFP, the CP-mode was
always used (proposed BSS method).Results and Discussion
Figure 1 shows
the MRI magnitude and phase difference images for Subject 1 for which the traditional
BSS method was used. The null regions are apparent in magnitude images except
in the first RF-configuration which corresponds to CP mode excitation. In null
regions, the phase difference is distorted and thus the traditional BSS method
fails. Figure 2 shows the MRI magnitude image for Subject 2. Since the
excitation RF-configuration is always the CP-mode, the same MRI magnitude
images are obtained for all 8 RF-configurations. Figure 3 shows additional
magnitude images which are acquired with the FLASH sequence to show null regions for the different RF-configurations.
Figure 4 shows the phase difference and $$$B_1^+$$$ magnitude images for the different
configurations. In contrast to the case where the RF-configurations were the
same for the ExcRFP and OffRFP, the phase difference images show no sign of
distortion in null regions. Therefore, we were able to apply the proposed BSS
method and get $$$B_1^+$$$ magnitude maps for each configuration.Conclusion
Mapping $$$B_1^+$$$ magnitude fields for any arbitrary
RF-configuration is desirable and often required in pTx systems. For
this purpose, we propose a new BSS based method which always uses the CP-mode for
ExcRFP while the RF-configuration of interest is used only for the OffRFP. Through
this approach, the excitation and mapping processes are separated, allowing the
proposed BSS method to work even if the RF-configuration of interest has
very-low magnitudes within the imaging volume. Another important feature of the
method is that the acquired $$$B_1^+$$$ magnitude maps have the same noise
characteristics in all regions since these maps are always obtained by using
MRI signal resulting from CP-mode excitation. We believe this method
can be a useful tool for the design, development and verification of pTx
systems.Acknowledgements
This
work was supported by The Natural Sciences and Engineering Research Council of
Canada (NSERC) Discovery Grant to Ravi Menon.References
1. Cunningham CH, Pauly JM, Nayak KS.
Saturated double-angle method for rapid B1+ mapping. Magn. Reson. Med.
2006;55(6):1326–1333
2. Yarnykh VL. Actual flip-angle imaging in
the pulsed steady state: A method for rapid three-dimensional mapping of the
transmitted radiofrequency field. Magn. Reson. Med. 2007;57(1):192–200.
3. Sacolick LI, Wiesinger F, Hancu I, Vogel
MW. B1 mapping by Bloch-Siegert shift. Magn. Reson. Med. 2010;63(5):1315–1322.
4. Nehrke K, Börnert P. Eigenmode analysis of
transmit coil array for tailored B1 mapping Magn. Reson. Med. 2010;63(3):754–764