High field Magnetic Resonance Angiography on mouse eye can be used to investigate the vasculature near the mouse eye. One of the difficulties with using on mice is the induced moving artifact. To overcome this obstacle, a custom mouse restrainer containing both a tooth bar and ear bars was designed and constructed for mouse eye imaging. The restrainer, in combination with a rigid surface coil, enabled three-dimensional mouse eye images to be captured with 30 um isotropic resolution. A three-dimensional mouse angiography was created by segmenting ocular blood vessels from these images.
An 8 mm surface coil was fabricated as indicated by the schematic in figure 1 a. Printed circuit board was used to mechanically and electrically stabilize the coil and the coupling network. The restrainer for the mouse head shown in figure 1 b was designed using SolidWorks (Dassault Systemes SOLIDWORKS Corp. MA, USA). The main function of the restrainer was to restrain the movement of the mouse head due to possible ‘hiccup’ breathing and other motion. The restrainer was printed using PolyLactic Acid (PLA) with a Lulzbot TAZ5 3D printer (Aleph Objects Inc., CO, USA). The surface coil and the restrainer were assembled in a home built probe as shown in figure 1 c.
The mouse preparation for this particular set up was performed similarly as in previous work [1]. The only addition to this set up were the ear bars. The ear bars were inserted and anchored to the zygomatic arch hole to inhibit translation movement of the head. Proper set up of nose cone, biting bar, and ear bars allowed to restrain the mouse head from any movement (figure 2).The fabricated setup was tested at 14.1 T. Two-dimensional gradient echo imaging was performed and compared with the previous setup [1].
MR angiography of the mouse eye was performed three-dimensionally using a gradient echo sequence with 6mm x 7.2mm x 7.2mm Field Of View (FOV) and 200 x 240 x 240 pixel sizes (30 um isotropic pixel resolution). 40 ms repetition time (TR) and 5 ms echo time (TE) were used to create the contrast between blood vessels and surrounding tissues. The total experiment time for the scan was 38 mins 24 secs. The blood vessels were segmented and the shape of the eye was volume rendered using Avizo (Thermo Fisher, OR, USA).
Figure 3 shows the two dimensional retinal images using the old set up (a) and the new set up with restrainer (b). The inlays in figure 3 show an enlarged part of the retina. The contrast of the layers seen with the restrainer is much clearer than the one acquired with the old set up.
The middle slice of the three-dimensional MRI of the mouse eye is shown in figure 4 a. The micro-vessels that are represented by only a few pixels in this slide are well-resolved. The segmented vessels are shown in figure 4 b. The retinal blood vessels, color coded in red, are exhibited in respect to the position of the volume rendered eye.The choroidal blood flow and other blood vessels are also shown.
Both the previous [1] and the current work succeeded in determining the isolated retinal vessels. The outline of the choroidal vasculature was only shown in the current work. Choroidal vasculature edges were smeared in the previous work.