Vijayaraghavan Panda1, Siva Sai Krishna Puranam1, Lance Delabarre2, Gregor Adriany1,2, and Anand Gopinath1
1Electrical and Computer Engineering, University of Minnesota, Minneapolis, MN, United States, 2Radiology, Center for Magnetic Resonance Research, Minneapolis, MN, United States
Synopsis
Slots are introduced to the standard microstrip transmission line RF coil
element, to enhance the current density and obtain better improved transmit
efficiency. A prototype of such a
slotted microstrip element is compared with the standard microstrip element at
297.25 MHz for 7T MR scanner to show the feasibility of the design at higher
magnetic fields. A 60% improvement in
the transmit efficiency and increased penetration of field into the phantom are
observed experimentally. The B1+ field distribution in
both transverse and sagittal plane are shown for the comparison.
Purpose
The purpose of this work is to design an efficient RF transmission line
coil with increased B1+ efficiency for high SNR of the
received MR signal and better image contrast with 7T and higher magnetic field
MR scanners.Method
The microstrip transmission line coil element [1] is a half wavelength
resonator which is foreshortened with the help of termination capacitors as
shown in Fig 1. A single series capacitor or a simple pi-matching network [2]
is used to match the element to input impedance. The field distribution in the
microstrip element can be approximated to say that it is homogeneous. The
microstrip transmission line element carries maximum current along their edges
due to skin effect, and these currents are the main cause for the time varying
magnetic fields (B1+ field). The transmit efficiency
which is defined as the ratio of magnitude of B1+ field
to the applied power (µT/√W), of these transmission line elements can be
improved by increasing the current density. Like the techniques used in the
past [3-4], slots are created in the microstrip line to increase the current
density along the edges of the conductor. As the B1+ field strength directly depends on the current
through the conductor, an improved B1+ transmit
efficiency can be achieved with these slots. The feasibility of the design at
high magnetic field strengths, is verified using a single channel comparison
between a slotted and a standard microstrip element on a 12.7 mm thick Teflon (εr = 2.2) substrate at
297.25 MHz. A length of 160 mm, line width of 10 mm and copper cladding of 35 mm are kept same for
both the elements. For the prototype of a slotted microstrip element, six set
of slots of size 6 mm x 15 mm are used in two columns as shown in Fig 1. Each
coil element was elevated 60 mm off the patient table and placed 12 mm away
from the acrylic wall of a tall sucrose phantom having a diameter of 180 mm, εr of 58.1 and
conductivity of 0.54 S/m. The tuning and matching of the elements at 297.25 MHz
are verified with the network analyzer. The strong coupling of the elements
with the phantom is observed by the field pickup probe measurement and with the
calculation of Q-factor. To increase the SNR, a metamaterial transmission line
element [5] as a receive only element, is placed on the opposite side. The
microstrip elements are assumed to have zero coupling across the phantom with
this metamaterial element. The
transmit efficiency (in µT/√W) of each coil element is measured from the
excitation flip angle induced in the phantom by an RF pulse waveform of known
power. The flip angle is measured using a method called Actual Flip-angle
Imaging AFI [6], a 3D dual repetition-time sequence. The AFI sequence
parameters were a non-slice selective excitation with a nominal flip angle = 60
degrees, TR1=20 ms, TR2 = 120 ms, TE = 3.06 ms and a
resolution of 2x2x4 mm.Results
Tuning and matching conditions, coupling of the elements with the
phantom, and comparison of field strength are verified on the bench. The
transmit efficiency of both the standard and slotted microstrip elements are
compared for the 7T MRI scanner. Fig. 2 shows a transverse and sagittal slice
of the phantom for the comparison of the transmit efficiency between the
standard and slotted microstrip element and Fig 3. shows the normalized
transmit efficiencies of the two elements with respect to the transmit
efficiency of the standard microstrip. An increase of 60% in the transmit
efficiency is observed with this single channel experiment. In addition, due to
the slots, narrower and more focused B1+ field
distribution is observed.Conclusion
Slots are introduced in the main conductor of the standard microstrip
transmission line element to increase its current density. A slotted microstrip
element is compared with the standard microstrip element, to validate its
performance for the high magnetic field MR scanners. The deeper penetration of
the B1+ field into the tissue or phantom from the RF coil
is observed which improves the SNR of the received MR signal and provides
better image quality in the MR imaging. A 60% improvement in the transmit
efficiency is obtained without any additional cost or loss in performance and
this can be easily implemented in any other elements that uses microstrip
technology like loop coils or dipoles.Acknowledgements
This work is
supported in part by the National Institute of Health (NIH) NIH-EB0006835, in
part by S10 RR026783 “Multichannel Transmit Frontend for 7 Tesla” WM KECK
Foundation and in part by University of Minnesota.References
[1] Zhang,
X., Ugurbil, K. and Chen, W., 2003. Journal of Magnetic Resonance, 161(2),
pp.242-251
[2] Sohn, S.M., DelaBarre, L., Vaughan, J.T. and
Gopinath, A., 2012, June. In Microwave Symposium Digest (MTT), 2012
IEEE MTT-S International (pp. 1-3). IEEE
[3]Sohn, S.M., DelaBarre, L., Gopinath, A. and
Vaughan, J.T., 2014. IEEE transactions on microwave theory and techniques, 62(8),
pp.1784-1789
[4] Son, H.W., Cho, Y.K. and Yoo, H., 2014. A
novel RF resonator for human-body MRI at 3 T. Journal of the Korean
Physical Society, 64(6), pp.813-816
[5] Panda, V et al., IEEE J-ERM 2017 (under
review)
[6] Yarnykh, V.L et al., 2007. Magnetic
resonance in Medicine