Azma Mareyam1, Jitendra Sharma1,2,3, Mitul Desai4, Christin Y Sander1,2, Eric Frederick5, Benjamin B Bartelle4, Atsushi Takahashi1,6, and Lawrence L Wald1,2
1Department of Radiology, A.A Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Department of Brain and Cognitive Sciences, MIT, Cambridge, MA, United States, 4Department of Bioengineering, MIT, Cambridge, MA, United States, 5Bruker Biospin, Billerica, MA, United States, 6McGovern Institute for Brain Research, MIT, Cambridge, MA, United States
Synopsis
High-field functional MRI of non-human primates
provides elevated BOLD contrast in a well-established neuroscience model but
requires an optimized coil design to boost the SNR, facilitate image
acceleration while maintaining mechanical and electrical stability during the
awake primate experiment. In this work we have constructed and tested 4-channel
9.4T and a 5-channel 3T receive-only array optimized for high-resolution
whole-brain imaging of the anesthetized and awake marmosets.
Introduction
Functional MRI of awake non-human primates,
especially in small primates like marmosets, present unique challenges due to
increased demands on image encoding and SNR required for high resolution
imaging, along with the need to mitigate body-motion induced susceptibility
distortions. High field strengths and receive-coil arrays can both boost the
sensitivity required for smaller voxel sizes, but require dedicated receive
coil geometries for accelerated primate imaging studies and careful attention
to the stresses of the primate and the high-field environment. Previous studies
have illustrated the benefits of using custom built monkey coils for brain
imaging [1]. In our work, we extend this
philosophy of imaging marmosets also beyond 7T [2]. Here we present a 4-channel
9.4T and a 5-channel 3T coil, compatible with a head restraint system. The coil
diameter was chosen as a trade-off between element count, depth penetration and
whole-brain coverage. The coil array is embedded on a custom head restraint
system for use across different bore sizes, providing mechanical stability
while minimizing the distance between the array elements and the marmoset
brain. The coverage and sensitivity of the array is demonstrated using in vivo images in marmosets.
Methods
Fig. 1a shows the 4-channel
9.4T array with elements laid out on an ABS plastic former fabricated with a 3D
printer, and positioned to surround the brain fully, while not obstructing the
head-post system or the visual stimulation. Occipital and frontal port holes
allow access to various brain areas to combine imaging with optogenetic
manipulations and electrophysiological recording, as well as head-post
implants, to address various experimental needs. Ports aligning ear canals
accommodate custom designed ear bars. Fig. 1b
shows the completed 5-channel 3T array. The array, transmit coil and preamps
are mounted to the monkey cradle. The 4-channel 9.4T array (Fig. 1c) consists of four
elements made of 16 AWG wire loops [3], 5 cm in diameter, with four equally-spaced
capacitors. Preamp decoupling is achieved with a cable length of 25 cm, with
the preamps placed some distance away to allow access and room for
immobilization. This necessitated the use of lattice baluns at the drivepoint
to each element to eliminate cable common-mode current issues. The lattice
balun with a PIN diode provides matching and detunes each element during
transmit. In addition to this, cable traps are placed behind the preamps. A 9 cm
in diameter detunable single-loop coil provides excitation.
Data
was acquired on an animal 9.4T scanner (Bruker Biospin, Billerica, MA) and the
3T scanner (Siemens Healthcare, Erlangen, Germany). Array noise covariance was
estimated from thermal noise data acquired without RF excitation, and SNR maps
were computed following the method of Kellman & McVeigh [4]. While
comparing SNR maps between arrays at the 9.4T, the same transmit coil was used. We performed
structural imaging in an adult marmoset, initially under anesthesia, using the
5-channel 3T array.
Results
Each element (Fig. 1c) shows a Q unloaded-to-loaded
ratio of ~260/100 with a spherical phantom 3.5 cm in diameter. Loaded S21
between neighboring elements ranges from ‑26 dB to ‑15.5 dB. S11
reflections showed the elements tuned and matched to 50 Ohms. Fig. 2
compares the SNR maps between the 4-channel 9.4T marmoset array and a 4-channel
Bruker rat array applied to the 3.5cm phantom, showing about a significant SNR
increase. Fig .3 compares the SNR maps between the 5-channel 3T marmoset array
with a volume coil (diameter 63mm and length 51mm) showing the potential for
SNR increases in many brain regions. The measured EPI time-course stability of
the 4-channel 9.4T array coil was 1.3% for peak-to-peak and 0.2% for the
5-channel 3T array. It is to be noted the numbers are relative at each field strength but cannot be used to
compare the coils at two different field strengths.
Fig. 4 shows structural images of an adult marmoset brain at
350 um isotropic resolution with excellent temporal stability, no motion
artifacts and high resolution throughout the brain. The entire assembly – head
coil and cradle was easy to use for placing the animal, provided excellent
stability across scans and uniform
coverage of entire brain.
Conclusions
The 9.4T 4-channel
monkey coil and the 3T 5-channel monkey coil provides an increase in overall SNR
compared to arrays not designed for marmoset imaging. The EPI stability was adequate,
indicating the coil’s utility for functional brain imaging. This design meets
the additional demands placed on the coil by the high field strength, the
mechanical constraints of the monkey cradle and stimulus, and measurement apparatuses.Acknowledgements
Support
from Simons Center for Social Brain and Poitras Foundation. Thanks to Jose
Estrada, Gao Yang.References
[1] Mareyam et
al. (2011) Proc ISMRM. [2] Papoti
et al. (2017) MRM 78(1):387-98. [3] Wiggins et
al. (2006) MRM 56:216-23. [4] Kellman et
al. (2005) MRM, 54(6):1439-47.