Marcel Gratz1,2, Jasmin Jäger3, Mathias Nittka4, Stephan Kannengiesser4, Josef Pfeuffer4, Gregor Koerzdoerfer4,5, Rainer Kirsch4, Florian Meise4, Harald H. Quick1,2, Vikas Gulani6, Mark Griswold6, Michael Forsting3, and Lale Umutlu3
1Erwin L. Hahn Institute for MRI, University of Duisburg-Essen, Essen, Germany, 2High Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany, 3Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany, 6Case Western Reserve University, Cleveland, OH, United States
Synopsis
This
study puts the focus on estimating the stability of MRF data in
phantoms, volunteers and patients and investigates the options to
integrate MRF into routine clinical workflows.
The
phantom and in-vivo scans yielded consistent and reproducible results
throughout different scan days. Thus, MRF enables a fast and reliable
generation of quantitative data in vivo and, along with
appropriate reference values, has the potential to improve
diagnostics in neuroimaging.
Introduction
Magnetic
resonance fingerprinting (MRF) enables the direct derivation of
tissue-characteristic quantities such as spin density and relaxation
constants T1 and T2. The additional information on tissue
characteristics derived from MRF bears the potential to leverage the
understanding of tumor biology to advanced levels and potentially
improve current clinical MR imaging procedures. However, the
reliability of these novel mapping techniques needs to be thoroughly
evaluated before introducing it into a productive clinical
environment. This study puts the focus on estimating the stability of
MRF data in phantoms, volunteers and patients and investigates the
options to integrate MRF into routine clinical workflows.Methods
A
non-commercial MRF sequence prototype1 was implemented on a 1.5T
and a 3T MR system (MAGNETOM Aera and Skyra, Siemens Healthcare,
Erlangen, Germany). For repeated stability scans, a standardized NIST
system phantom2 was used, which contains multiple compartments
resulting in a broad range of T1 and T2. The obtained quantities
were compared to their corresponding reference values provided by
NIST and analysed with respect to their variability throughout
different scan days (Fig. 1 and 2).
Moreover,
MRF was applied in n = 10 healthy volunteers (at both field
strengths), acquiring maps from three reference slices covering the
upper lobar, mid-ventricular and cerebellar neurocranium. Based on
these scans, 16 anatomical regions were selected in each of the
datasets (Fig. 3) and compared in an intra-individual fashion. In
addition, a total of n = 80 patients with various neurological
pathologies that were scheduled for a routine clinical protocol at 3T
received an additional MRF protocol prior to the application of
contrast agent. The anatomical regions that were not infiltrated by
the pathology were extracted from the patient imaging data and
compared across subjects and to their corresponding values in healthy
volunteers, where applicable.Results and Discussion
The
NIST phantom scans yielded consistent and reproducible results
throughout different scan days with a standard deviation of less than
5% for T1 and less than 8% for T2 (Fig. 2). All measured T1 and T2
values agreed well with the reference values as given by NIST with a
consistent tendency to underestimate the long relaxation times and
overestimate the short relaxation times which confirms an earlier
finding of Jiang et al.3.
All
subjects tolerated the MRF protocol well. Given the rather short
acquisition and reconstruction times on the order of less than one
minute, an integration into the routine scanning workflow seems
feasible. However, particular care needs to be taken when positioning
the field-of-view to adequately cover the anatomic regions under
study.
The
obtained T1 values were reproducible and showed a standard deviation
of less than 10% across volunteers and patients. The same held true
for T2 values, yet with a higher spread of up to 20% (Fig. 4 and 5).
In general, quantities obtained from patient data showed higher
deviations from their mean value (T1: 7.2%, T2: 14.8%) than those in
healthy subjects (T1: 4.9%, T2: 8.6%). This may be attributed to the
larger spread in age and gender of the patient cohort as shown in
previous publications. It is thus evident, that the latter parameters
have to be considered when trying to establish a clinical reference
database of normal and pathological T1 and T2 in various anatomical
regions.Conclusion
MRF
is feasible in phantoms, healthy volunteers, and patients. It enables
a fast and reliable generation of quantitative data in vivo and thus,
along with appropriate reference values, has the potential to improve
diagnostics in neuroimaging.Acknowledgements
No acknowledgement found.References
1.
Jiang,
Y., et al., MR
fingerprinting using fast imaging with steady state precession (FISP)
with spiral readout.
MRM, (2015)
2. Russek et al., Characterization
of NIST/ISMRM MRI system phantom.
Proc. of the 20th Annual Meeting of ISMRM (2012), #2456
3.
Jiang et al., Repeatability
of magnetic resonance fingerprinting T1 and T2 estimates assessed
using the ISMRM/NIST MRI system phantom,
MRM (2016)