Seong Dae Yun1 and N. Jon Shah1,2
1Institute of Neuroscience and Medicine, Medical Imaging Physics (INM-4), Forschungszentrum Juelich, Juelich, Germany, 2Faculty of Medicine, Department of Neurology, JARA, RWTH Aachen University, Aachen, Germany
Synopsis
EPI
has been widely used for fMRI due to its relatively fast imaging speed.
However, as one of its drawbacks, ghost artefacts need to be corrected. In the
community, an approach that utilises three non-phase encoded navigator echoes
is commonly used for the correction. Although this scheme is quite effective,
as the matrix size increases for high-resolution imaging the navigator echoes constitute
a significant contribution in increasing the minimum required TE. To overcome
this issue, this work employs an alternative navigator echo scheme. Here, its
performance was analysed and whole-brain submillimetre-resolution fMRI (0.75 ×
0.75 mm2) was demonstrated at 3T.
Purpose
EPI
has been widely used for fMRI due to its relatively fast imaging speed. However,
as one of its drawbacks, ghost artefacts need to be corrected. In the
community, an approach that utilises three non-phase encoded navigator echoes
is commonly used for the correction. This
scheme has been shown to be robust and effective in removing the artefacts. However,
as the matrix size increases for high-resolution imaging, the navigator echoes play
a significant part in increasing the minimum required TE. To overcome this
issue, this work proposes to employ an alternative scheme which allocates the
navigator echoes in a preceding, separated TR loop.1 This scheme has
not yet been analysed in terms of the steady-state and optimal flip-angle. The
purpose of this study is to perform analysis for the aforementioned issues and
demonstrate the method for submillimetre-resolution fMRI at 3T.Materials and methods
Figure
1 shows two different navigator echo schemes for EPI: one with the ‘navigator
echoes’ in the same TR loop as the ‘main EPI readout’ (standard; see Fig. 1a)
and the other with the ‘navigator echoes’ in a separated TR loop (alternative;
see Fig. 1b). As shown here, the alternative scheme requires additional
excitation (flip angle of α1) for the navigator echoes, which leads
to a slight increase in scanning time. However, in contrast to the standard
scheme, the navigator echoes do not have any impact on the increase of minimum
required TE. This implies that a larger number of EPI echoes can be allocated until
the given optimal TE for fMRI. Here, in order to check if the alternative
scheme reaches a steady-state, the equation for the longitudinal magnetisation was
derived under the ‘steady-state incoherent’ condition (i.e. elimination of
remaining transverse magnetisation by spoiling).2 The derived equation
is given as follows:
$$ M_{z(n)}^{alt}(0^{-})=M_{eq}\cdot\left[(E_{TR1}\cdot \cos \alpha_{1})^{n-1}\cdot(E_{TR2}\cdot \cos \alpha_{2})^{n-2}+(1-E_{TR1})\cdot\frac{1-(E_{TR1}\cdot \cos \alpha_{1}\cdot E_{TR2}\cdot \cos \alpha_{2})^{n-1}}{1-(E_{TR1}\cdot \cos \alpha_{1}\cdot E_{TR2}\cdot \cos \alpha_{2})}\\+(1-E_{TR2})\cdot(E_{TR1}\cdot \cos \alpha_{1})\cdot\frac{1-(E_{TR1}\cdot \cos \alpha_{1}\cdot E_{TR2}\cdot \cos \alpha_{2})^{n-2}}{1-(E_{TR1}\cdot \cos \alpha_{1}\cdot E_{TR2}\cdot \cos \alpha_{2})}\right] $$
where Mz(n)alt(0-)
and Meq denote the
longitudinal magnetisation just before the second TR loop excitation (α2)
and the magnetisation at thermal equilibrium, respectively. Moreover, ETR1
and ETR2 indicate exp(-TR1/T1)
and exp(-TR2/T1),
respectively.
For the standard scheme, as it is
well-known from previous works,3 the equation is given as follows:
$$ M_{z(n)}^{std}(0^{-})=M_{eq}\cdot\left[(E_{TR}\cdot \cos \alpha)^{n-1}+(1-E_{TR})\cdot\frac{1-(E_{TR}\cdot \cos \alpha)^{n-1}}{1-(E_{TR}\cdot \cos \alpha)}\right] $$
where ETR denotes exp(
-TR/T1).
Results
The
two equations above were simulated under the following conditions: α/α1/α2
= 90/10/90°, TR/TR1/TR2 = 3000/10/2990 ms and T1
= 1.2 s. The results are plotted in Fig. 2a. It reveals that the signals are
saturated to a certain level (i.e. steady-state) in both the standard and the alternative
schemes. This was further verified with real in vivo data obtained with the same conditions. As shown in Fig.
2b, the signal behaviour is quite similar to the simulation data; although the
real signals present transverse magnetisation in contrast to the above
equations, the evolution of the transverse magnetisation is congruent with that
of the longitudinal magnetisation. Here, in the alternative scheme, an optimal flip
angle for navigator echoes (α1) was determined based on the
steady-state level and the performance of ghost artefact removal. Figure 3
depicts in vivo data results where it
is observed that as α1 increases, the stead-state level decreases, but
the performance of ghost removal enhances (i.e. remaining ghost energy
decreases). Nine degrees is a good compromised value for both issues. For this
angle, the steady-state level of the alternative scheme is about 98% of that of
the standard scheme and the performance of ghost removal is also quite similar
each other. For the determined optimal angle, a high-resolution visual fMRI
study was performed with the following settings: TR/TE = 3000/35 ms, FOV = 216 × 216 mm2,
matrix = 288 × 288 (0.75 × 0.75 mm2), partial Fourier = 5/8, 2-fold
in-plane/3-fold inter-plane (multi-band) acceleration and 48 slices with 2.3 mm
thickness. Here, the minimum possible TE of
the alternative scheme was 5.86 ms shorter than that of the standard scheme (see Fig. 4). Figure 5 shows a
representative reconstructed image and three slices from a structural scan with
activated regions superimposed on them. It is observed that the activated
voxels are precisely localised along with the cortical regions, which successfully
demonstrates high-resolution mapping.Conclusions
In this work, an alternative navigator
echo scheme was analysed in terms of evolution of longitudinal magnetisation,
steady-state level and the performance of ghost artefact removal. The method
was shown to be effective in decreasing the minimum required TE and was
demonstrated for whole-brain submillimetre-resolution fMRI.Acknowledgements
No acknowledgement found.References
-
Wielopolski
PA. Echo-planar imaging pulse sequences in Echo-planar imaging theory,
technique and application, ed. Schmitt F, Stehling MK, Turner R. 1998. p. 96.
-
Chen
YN, Haacke EM. Current Protocols in Magnetic Resonance Imaging. John Wiley
& Sons, Inc. 2001. B5.1.1-B5.1.16.
-
Gescheidtova
E, Bartusek K. Longitudinal Relaxation Time Measurement in MR with
Transient-state Magnetization. 2008;4(1):61-64.