The Bloch-Siegert-effect imprints the rf-coil’s transmission field in the phase of the acquired signal. Consequently the measured amplitude information can be used to determine other MR-Parameter. A previously published sequence based on a CPMG spin echo train uses both information pathways and allows simultaneous acquisition of $$$T_2$$$- and $$$B_1^+$$$-maps. However, the formerly proposed method produces potentially erroneous $$$B_1^+$$$-map in inhomogeneous $$$B_0$$$-fields.
This work presents a method to reduce those errors by combining signal from different echoes. This allows a robust measurement of $$$B_1^+$$$-maps in inhomogeneous $$$B_0$$$-fields as present in high-field MRI with simultaneous $$$T_2$$$ quantification.
For quantitative MRI it is important to determine the $$$B_1^+$$$-field of the transmission rf-coil. That allows to correct the impact of mismatched flip angles during the measurement on the magnetization evolution. One possibility to obtain the transmission field is based on the Bloch-Siegert effect1. This effect allows to encode the $$$B_1^+$$$-information into the signal phase while at the same time the signal amplitude is practically maintained. This allows the acquisition of the $$$B_1^+$$$-field simultaneous with other MR parameters encoded in the signal magnitude. This was shown for $$$T_2$$$ quantification using a CPMG spin-echo train2. However for imperfect refocusing pulses the signal phase of off-resonant spins is not constant in the echo train. Consequently the use of phase information from different echoes may result in erroneous $$$B_1^+$$$ values. In this work a method to minimize those errors is presented.
Material and Methods
By applying off-resonant rf-pulses with $$$\omega_{off}\gg{\gamma}B_1^+$$$ and $$$\omega_{off}\gg\Delta\omega_0$$$ the phase of the transversal magnetization is changed by:
$$\Phi_{BS}[\omega_{off}]\approx\frac{({\gamma}B_1^+)^2}{2\omega_{off}}\left(1-\frac{\Delta\omega_0}{\omega_{off}}\right)$$
while the magnitude is not influenced. Whereby the off-resnonance of the rf-puls is given by $$$\omega_{off}$$$ and the frequency shift caused by the $$$B_0$$$-field inhomogeneity by $$$\Delta\omega_0$$$. By taking the phase difference $$$\Delta\Phi$$$ of two measurement with alternated sign of $$$\omega_{off}$$$ $$$B_1^+$$$ can be calculated by1
$$B_1^+\approx\sqrt{|\Delta\Phi\,\omega_{off}|}/\gamma.\hspace{5em}(1)$$
Therefore it is essential that $$$\Delta\Phi$$$ is solely determined by the Bloch-Siegert effect. A published method2 for simultaneous measurement of $$$T_2$$$ and $$$B_1^+$$$ compares phase information taken at different echo times. However it can be shown3, that the phase for the magnetization in the nth echo of a CPMG echo train is given by:
$$\Phi_{CPMG}[n]\approx -\frac{n}{4}(-1)^n(\pi-\alpha)^2\sin[\Delta\omega_0{T\!E}].$$
It can be seen, that for imperfect
refocusing pulses ($$$\alpha\neq\pi$$$) the phase amplitude for offresonant magnetization increases
linear with echo number while the sign alternates. Consequently in the phase difference $$$\Delta\Phi$$$ of two echos with alternated $$$\omega_{off}$$$ sign but different echo number the term $$$\Phi_{CPMG}$$$ is not fully cancelled so that by using equation (1) erroneous $$$B_1^+$$$-values are produced. However
based on the character of this phase variation it is possible to choose echo number combinations so that both BS-encodings have the same mean phase offset $$$\Phi_{CPMG}$$$. For example using the mean phase from the first and fifth echo (both with positive BS-encoding) minus the third echo (negative BS-encoding) produce:
$$\Delta\Phi=\frac{\Phi_{BS}[\omega_{off}]+\Phi_{BS}[\omega_{off}]}{2}+\frac{\Phi_{CPMG}[1]+\Phi_{CPMG}[5]}{2}-\Phi_{BS}[-\omega_{off}]-\Phi_{CPMG}[3]\\=\Phi_{BS}[\omega_{off}]-\Phi_{BS}[-\omega_{off}].$$
Accordingly the term $$$\Phi_{CPMG}$$$ is removed and $$$\Delta\Phi$$$ now depends solely on the Bloch-Siegert phase difference even though different echo numbers were used.
Discussion and Conclusion
1. Sacolick Li, Wiesinger F, Hancu I, Vogel MW. $$$B_1$$$ mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63:1315-1322.
2. Sturm VJF, Basse-Lüsebrink TC, Kampf T, Stoll G, Jakob PM. Improved encoding strategy for CPMG-based Bloch-Siegert $$$B_1^+$$$ mapping. Magn Reson Med. 2012;68:507-515.
3. Bain AD, Anand CK, Nie Z. Exact solution of the CPMG pulse sequence with phase variation down the echo train: Application to $$$R_2$$$ measurements. Journ Magn Reson 2010;209:183-194.