David Otto Brunner1,2, Betram Jakob Wilm1,2, Simon Gross1, Benjamin Emmanuel Dietrich1, Christoph Barmet1,2, and Klaas Paul Pruessmann1
1Institute for Biomedical Engineering, ETH and University of Zurich, Zurich, Switzerland, 2Skope Magnetic Resonance Technologies Inc., Zurich, Switzerland
Synopsis
Devices
operating in the magnet bore, such as coil arrays, receivers or shim inserts
become ever more sophisticated as well as the employed image acquisition
schemes. However, quantitative testing for their influence on the gradient
switching performance is often difficult and time consuming. In this work we
present a methodology for measuring, evaluating and predicting eddy current
effects fast and quantitatively, employing a dynamic magnetic field camera and
a framework of linear time invariant system characterization. For
demonstration, standard components and several frequently applied RF coils are
assessed.
Introduction
Ever more
sophisticated devices are being used in MR imaging, such as high-density receive
coils, $$$B_0$$$-shim inserts [1] or combined RF and shimming arrays [2]. All devices require careful testing for
electromagnetic compatibility with the MRI system and its protocols. Typical procedures employ the
imaging capabilities of the scanner to detect potentially occurring problems
such as $$$B_1^+$$$ degradation, static $$$B_0$$$ distortions or added noise.
However, it is difficult to assess if a device impacts the accuracy of the
gradients by carrying eddy currents. Although the typically performed evaluation
of artefacts of demanding read-outs (e.g. EPI) allows determining overall
adequacy for the particular sequence, it is difficult to ascribe each artefact
to unwanted interactions of the device and the gradient system and to localize
and quantify the cause of the problem. This is because gradient distortions interfere
with the encoding and corrections required for the acquisition of the
validation data itself resulting in a complicated entanglement of various
effects. Additionally, an evaluation [3] is most valuable during development
but at the same time particularly difficult to perform if not all resources are
fully operational.
The assessment
of the suitability of devices and components remains therefore largely based on
the experience of developers.
As a
solution, we present a framework to measure and to quantify the effect of a device
on the spatio-temporal gradient accuracy working independently from the scanner
by employing a dynamic field camera (Fig. 1) and signal models based on
linear time-invariant systems (LTI) theory. Based on this analysis the source
of the distortions can be localized and the impact on image acquisitions can be
quantitatively predicted.Methods
The fields
($$$B_{DUT}(\omega)= F(b_{DUT}(t))$$$)
produced by eddy
currents on the DUT’s (Device Under Test) conductors, being small compared to
the gradient fields, are treated as a perturbation of the system without DUT ($$$B_{ref}$$$).
Such an LTI is characterized by its impulse response ($$$h_{DUT}(t)$$$) or its Fourier transform, the transfer function
$$$H_{DUT} (\omega)=F(h_{DUT} (t))$$$ [4]:
$$b_{DUT}(\vec r, t)=b_{ref}(\vec r,t) \otimes
h_{DUT} (\vec r, t)=F(B_{ref}(\vec r, \omega) \cdot H_{DUT} (\vec r,
\omega)).$$
Spherical
harmonic functions ($$$Y_{l,m}(\vec r)$$$) are employed for capturing the
spatial distribution of the field:
$$ b_{DUT} (\vec r, t)=\sum_{l’,m’}b_{DUT} (t)\cdot
Y_{l’,m’}(\vec r)=\sum_{l’,m’} Y_{l’,m’}(\vec r) \sum_{l,m} ( h_{DUT}^{l,m->l’,m’}(t)
\otimes b_{ref}^{l,m}(t)).$$
Acquiring the
field evolution with DUT ($$$b_{tot}=b_{DUT}+b_{ref}$$$) and without while driving
frequency sweeps subsequently on all axes yields the transfer functions:
$$ H_{DUT}^{l,m->l’,m’}(\omega)=F(h_{DUT}^{l,m->l’,m’}(t))=\frac{B_{tot}^{l’,m’}(\omega)-
B_{ref}^{l’,m’}(\omega)}{ B_{ref}^{l,m}(\omega)}.$$
A field
camera [6] (DFC-16, Skope AG, Zurich,
Switzerland) is employed in a 3T scanner (Achieva, Philips Healthcare, Best,
Netherlands). As DUT components such as a brass mesh, a PCB and two commercial
coils, a TR head-birdcage and an 8-channel receive-only head-array, have been
employed.
For
illustration, image acquisitions have been simulated using the measured reference
impulse responses and reconstructed using that of the scanner with the coil.Results
See Figures 3-5.Conclusion
The proposed measurement of the transfer
function of devices gives detailed insight into the spatio-temporal field
distortions induced during scanning. Forward calculation of the field evolution
employing the response of the scanner and the device allows evaluating the
imaging performance time and cost effectively in-silico for all sequences and
in principle for various scanners. Similarly, the measurements of the properties
of isolated components can predict the behaviour of the entire assembly. These abilities
grow ever-more useful, as in-bore deployed setups become more sophisticated and
the gradient systems more performant while the requirements on image quality
and quantitation are steadily growing.
In addition, the captured spatial distribution
of the distorting fields allows localizing its source. The frequency response relates
to the lifetime of the eddy currents and the mechanical response of the device
being determined by the geometry and the involved materials’ properties. Such
information is key for determining the source of the problem and improving the
design, but difficult to predict otherwise since all, material parameters,
their macroscopic assembly and the interaction with the various scanner induced
fields have to be accounted for in detail. Doing so at an early design stage is valuable in minimizing development
time.
The distortions induced by standard coils
varied strongly and can generate visible effects in advanced protocols. While
EPI phase correction [7] and other delay compensations [8] can largely correct for direct
terms, cross terms require more advanced eddy current compensation schemes even
involving dynamic shimming for higher orders. Hence, it is imperative to
minimize perturbations by hardware design in the first place, especially for use
in high-performing gradient systems.Acknowledgements
No acknowledgement found.References
1. C.
Juchem, et al., Dynamic Multi-Coil
Technique (DYNAMITE) Shimming of the Rat Brain at 11.7 Tesla. NMR Biomed,
2014. 27(8).
2. J.P. Stockmann, et al., A 32-channel combined RF and B0 shim array
for 3T brain imaging. Magn Reson Med, 2015
3. M. Alecci, et al., Characterization and reduction of
gradient-induced eddy currents in the RF shield of a TEM resonator. Magn
Reson Med, 2002. 48(2).
4. S.J. Vannesjo, et al., Gradient and shim pre-emphasis by inversion
of a linear time-invariant system model. Magn Reson Med, 2016
5. J. Busch, et al., Analysis of temperature dependence of
background phase errors in phase-contrast cardiovascular magnetic resonance.
Journal of Cardiovascular Magnetic Resonance, 2014. 16(1).
6. B.E. Dietrich, et al. An Autonomous System for Continuous Field
Monitoring with Interleaved Probe Sets. in Proc Int Soc Magn Reson Med. 2011 p. 1842.
7. P. Jezzard, et al., Characterization of and correction for eddy
current artifacts in echo planar diffusion imaging. Magn Reson Med, 1998. 39(5).
8. N.-k. Chen, et al., Removal of EPI Nyquist ghost artifacts with
two-dimensional phase correction. Magn Reson Med, 2004. 51(6).