Elena Kaye1, Daniel V Litwiller2, Maggie Fung2, Stephen Solomon1, and Majid Maybody1
1Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2GE Healthcare, New York, NY, United States
Synopsis
MRI guided biopsy
plays an important role in the diagnosis of bone lesions. Although T1w FSE offers
exceptional signal-to-noise ratio and resolution, its acquisition time is
relatively long. In this study, we evaluate a single-shot FSE with centric partial Fourier encoding, variable refocusing flip
angle, and an inversion recovery preparation pulse (vrfSSFSE-IR) for MRI-guided bone biopsy application with the goal to reduce the overall procedure time.
Introduction
Magnetic resonance imaging (MRI) guided biopsy
plays an important role in the diagnosis of bone lesions1-3. Despite excellent MRI
visualization of bone tumors and lack of ionizing radiation, the majority of current
procedures are carried out under the guidance of computerized tomography4 owing to the superior quality of the biopsy
instrument and speed of this imaging modality. The large size of the currently available MR-compatible titanium biopsy needles
(3-4 mm diameter) results in substantial image artifacts thus rendering gradient
echo-based sequences unsuitable for this procedure. T1-weighted fast spin-echo (FSE)
sequence, on the other hand, helps minimize the susceptibility artifact around
the needle while still visualizing the bone lesions (Figure 1). Although T1w
FSE offers exceptional signal-to-noise ratio and resolution, its acquisition
time is relatively long. Hence, when repeated multiple times to guide the
placement of the biopsy needle, it substantially impacts the duration of the
procedure. On the other hand, single-shot FSE with centric partial Fourier
encoding, variable refocusing flip angle, and an inversion recovery preparation
pulse (vrfSSFSE-IR)5 can produce T1-like
contrast while keeping acquisition time very short. The goal of this work was
to evaluate the vrfSSFSE-IR pulse sequence for MRI-guided bone biopsy
application.Methods
We evaluated the
vrfSSFSE-IR sequence on a 1.5T MRI scanner (Discovery MR450w, GE Healthcare,
Waukesha, WI) using an 8-channel head coil. An ex vivo bovine bone-muscle
tissue sample with a 4-mm titanium bone biopsy needle (Invivo, FL, USA) was first
scanned with a conventional FSE sequence from our clinical bone biopsy protocol
and then with the vrfSSFSE-IR sequence. The vrfSSFSE-IR sequence was repeated for
inversion time (TI) ranging from 300 to 1500 ms. Regions of interest were
placed in muscle and bone marrow tissue on both conventional FSE and
vrfSSFSE-IR images to calculate the average signal intensity (SI) (Figure 2A).
The ratio of muscle to bone marrow average SI obtained from all images was
analyzed, and the TI value producing the SI muscle to bone marrow ratio equivalent
to that measured on the FSE image was determined. The scan time as a function
of number of slices was also evaluated for a fixed set of parameters (Table 1).
The appearance of the biopsy needle artifact was also evaluated.Results
Figure 2A shows the images
obtained with the vrfSSFSE-IR sequence for a TI of 300 to
1400 ms. An exponential function of the form SImuscle/SIbone
= a*(1 – exp(-b*TI) was found to describe the relationship between the muscle
to bone marrow SI ratio and the TI with 0.99 correlation coefficient of the fit
(Figure 2B). The value varied from 0.29 to 0.89. The SI ratio measured in the
FSE image was 0.35. Based on the exponential fit obtained above it was found
that a TI of 318 ms would produce a FSE-equivalent muscle to bone marrow SI
ratio using the vrfSSFSE. Figure 3 shows the comparable muscle to bone marrow
contrast in conventional FSE and vrfSSFSE-IR with TI of 300 ms. The
susceptibility artifact caused by the biopsy needle is comparable between the
two sequences, with the needle appearing narrower in the SSFSE image due to the
slightly higher bandwidth (Table 1). The total scan time, as a function of the
number of slices, is shown in Figure 4.Discussion
The results of this
study demonstrate that T1-like contrast in muscle to bone marrow tissue can be
achieved using the vrfSSFSE-IR sequence at a shorter acquisition time while
keeping the biopsy needle artifact comparable. Depending on the conspicuity and
size of the bone lesion being targeted, the lower signal-to-noise ratio and the
blurring that is inherent to the SSFSE sequences may be a limitation. However,
even in that scenario, vrfSSFSE-IR may be useful to guide placement of the
needle after the conventional FSE image is acquired for target visualization.
The linear relationship between the scan time and number of slices is
especially valuable in an interventional setting where only a very limited
region needs to be imaged repeatedly in order to depict the progress of the
needle placement. Using vrfSSFSE-IR to image 6 slices, for example, there will
be ~4 fold reduction in scan time compared to imaging with FSE.Conclusion
This preliminary
evaluation shows that the vrfSSFSE-IR sequence may play a valuable role in
reducing the overall procedure time of the MRI-guided bone biopsy. As part of future
work, this sequence will be accessed in the bone biopsy procedures to establish
its true clinical value. Acknowledgements
No acknowledgement found.References
1. König
CW, TRüBENBACH J, Böhm P, Fritz J, Duda SH, Pereira PL. Magnetic
resonance-guided transcortical biopsy of bone marrow lesions using a magnetic resonance
imaging-compatible piezoelectric power drill: preliminary experience.
Investigative radiology. 2003;38(3):159-63.
2. Neuerburg JM, Adam G, Buecker A, et
al. MRI‐guided biopsy of bone in a hybrid system. Journal of Magnetic Resonance
Imaging. 1998;8(1):85-90.
3. Parkkola R, Mattila K, Heikkilä J,
et al. Dynamic contrast-enhanced MR imaging and MR-guided bone biopsy on a 0.23
T open imager. Skeletal radiology. 2001;30(11):620-4.
4. Leffler SG, Chew F. CT-guided
percutaneous biopsy of sclerotic bone lesions: diagnostic yield and accuracy.
AJR American journal of roentgenology. 1999;172(5):1389-92.
5. Litwiller DV, Taviani, V, Estkowski
L, Bayram E. Fast, motion-robust T1-weighted body imaging with single-shot fast
spin echo via centric partial Fourier encoding and variable refocusing flip
angle. ISMRM; 2016; Singapore.
6. Busse
RF, Hariharan H, Vu A, Brittain JH. Fast spin echo sequences with very long
echo trains: design of variable refocusing flip angle schedules and generation
of clinical T2 contrast. Magnetic resonance in medicine. 2006;55(5):1030-7.