Felipe Godinez1, Arian Beqiri1, Jose N Teixeira1, Joseph V Hajnal1, and Shaihan Malik1
1School of Biomedical Engineering & Imaging Sciences, King's College London, LONDON, United Kingdom
Synopsis
A multislice method for
measuring relative coupling between elements of a parallel transmit (PTx) array
and conductive structures such as guidewires at multiple internal locations. Fast
2D image slices are used to measure the relative coupling along the guidewire
inside a large phantom. This is useful during conditions where the coupling modes
can only be measured from within the object volume and where current sensors
cannot be placed. Such conditions exist in large phantoms and can exist in
humans.
Introduction
MRI guidance of
interventional procedures using guidewires is limited by potential heating of
tissue caused by radio frequency (RF) coupling with the MRI transmit coil 1–3. In a PTx system decoupling modes exist
that can reduce RF coupling risks and a procedure for finding them using current
sensors on the wire has been demonstrated 4. Previously we demonstrated that the
relative coupling could be measured with a single 3D volume acquisition with very
low flip angle (vLFA) 5, providing a single measurement that cannot
determine all coupling modes. We propose
to extend this method to achieve the coupling measurement at multiple locations
along the internal portion of the guidewire by acquiring multiple 2D images. This
can allow determination of multiple coupling modes offering potential for
comprehensive safety control.Theory
The presence of a linear conductor
leads to local enhancement of the RF magnetic field (B1+) for any
driven transmit coil element, producing a local B1+ enhancement that
is determined by currents on the wire. vLFA
gradient echo scans have the property that the MR signal is linear in B1+.
Since the B1+ field drops off quickly with radial distance, signals acquired at
vLFA are dominated by contributions very close to the conductor. This is
further reinforced by local receive sensitivity enhancement due to the
reciprocal effect of the conductor on the receiving coil. We hypothesize that the
hyperintense pixels at the wire location in a 2D image excited with a vLFA RF
pulse will provide a measure of the relative coupling at that specific slice
location. Each slice location is sampled once per transmit coil, such that, with
an N channel Tx-array, the application of N measurement cycles (one per
channel) is sufficient to measure a full coupling matrix. The measurement
cycles can be highly accelerated by acquiring very small numbers of phase
encode steps.Methods
All measurements were
performed on a 3T Philips Achieva with an 8-channel Tx/Rx loop coil (Rapid
Biomedical GmbH, Germany) using a torso sized phantom filled with 40 liters of
poly acrylic acid gel 6 with inserted 90cm long nitinol
guidewire (Terumo Corporation, Japan) instrumented by an optically coupled toroidal
current sensor 7 placed over its proximal end outside
the scanner bore that was connected to the scanner’s spectrometer (Phillips)
for digitization. The proposed coupling measurement method was implemented with
5 slices using a 2D spoiled gradient echo sequence (FOV=157mmx400mm, Acq
Matrix=132x132, Slice Gap=20mm, Slice Thinkness=10mm, TR=10ms, TE=1.51ms, and nominal
FA=7 degrees). In each acquisition a single coil element was used to transmit
with all coils used to receive.
Coupling/decoupled modes
were computed on coupling measurement data (i) from the current sensor and (ii)
the proposed method using SVD as in ref 4. The coil array was then driven either
in the maximum coupled mode or a pure decoupled mode at the same drive level for
a high SAR scan, while temperature at the end of the wire was monitored using a
fibre-optic temperature probe (LumaSense Technologies, Inc. USA).Results and Discussion
Figure 1 shows signal from pixels
on the wire, determined by full imaging, and the measurement from the current
sensor. The polar plots are of the magnitude and phase of the coupling matrices.
Figure 2 shows the derived coupling modes from each method and their corresponding
singular values. It can be seen from the singular values derived from the multislice
measurements that there are at least two modes with significantly non-zero
values, which would be identified as high coupling modes. A single external
sensor cannot identify all of these.
Figure 3 shows the heating
results for maximum coupling mode (CM) for each method. A higher degree of heating was
observed when using the image-based method, suggesting that this method is able
to better identify the worst-case coupling mode for the wire than the current
sensor. In addition, using 5 slices rather than one slice led to a larger
amount of heating, again suggesting there is a need for multiple measurements
of coupling at different spatial positions. This is consistent with previous
work 4.
By retrospectively down-sampling
the acquired image data it was found that only 3 lines of k-space were required
to measure the coupling matrix without loss of accuracy.
Conclusion
Active control of guide wire
safety will require an ability to rapidly measure coupling, which may change as
the wire is manipulated. Although these preliminary experiments used fully
sampled 2D imaging, highly reduced sampling would potentially enable full
detection at multiple spatial locations in around one second, paving the way
for real time monitoring and control. Acknowledgements
This work was supported by the Wellcome EPSRC Centre for
Medical Engineering at Kings College London [WT 203148/Z/16/Z], MRC strategic
grant [MR/K006355/1] and developmental
pathway funding [MR/N027949/1] and by the National Institute for Health
Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS
Foundation Trust and King’s College London. The views expressed are those of
the authors and not necessarily those of the NHS, the NIHR or the Department of
Health. References
1. Nitz WR,
Oppelt A, Renz W, Manke C, Lenhart M, Link J. On the heating of linear
conductive structures as guide wires and catheters in interventional MRI. J
Magn Reson Imaging. 2001;13(1):105-114.
doi:10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO;2-0.
2. Park SM,
Kamondetdacha R, Nyenhuis JA. Calculation of MRI-induced heating of an
implanted medical lead wire with an electric field transfer function. J Magn
Reson Imaging. 2007;26(5):1278-1285. doi:10.1002/jmri.21159.
3. Griffin GH,
Anderson KJT, Celik H, Wright GA. Safely assessing radiofrequency heating
potential of conductive devices using image-based current measurements. Magn
Reson Med. 2015;73(1):427-441. doi:10.1002/mrm.25103.
4. Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Controlling radiofrequency-induced
currents in guidewires using parallel transmit. Magn Reson Med.
2015;74(6):1790-1802. doi:10.1002/mrm.25543.
5. Godinez F,
Hajnal J, Scott G, Mooiweer R, Malik S. “Pulse-acquire” method for obtaining
the guidewire coupling modes of a PTx transmit array. In: Proc. Intl. Soc.
Mag. Reson. Med. Honolulu: ISMRM; 2017.
6. ASTM standard
F 2182-2002a. Standard test method for measurement of radio frequency induced
heating near passive implants during magnetic resonance imaging. ASTM Int.
2002;(December):1-14. doi:10.1520/F2182-11A.1.7.
7. Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Interventional device visualization with
toroidal transceiver and optically coupled current sensor for radiofrequency
safety monitoring. Magn Reson Med. 2015;73(3):1315-1327.
doi:10.1002/mrm.25187.