4144

Asymmetric-echo-based Ultrashort Echo-time (ASE-UTE) Imaging
Jinil Park1, Jeongtaek Lee1, and Jang-Yeon Park1,2

1Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Republic of Korea, 2Center for Neuroscience Imaging Research, Institute for Basic Science, Suwon, Republic of Korea

Synopsis

The first 3D-UTE sequence used FID acquisition techniques using ramp sampling to minimize TE, but defining the center of k-space precisely was a difficult problem (true-k0). In case of echo acquisition containing echo peaks, it is better to define true-k0 than FID acquisition although true-k0 problem still exists because echo peaks often do not agree with true-k0 due to the echo shifts caused by system imperfections such as gradient delays and eddy currents. Here, we propose an asymmetric-echo-based 3D-UTE (ASE-UTE) sequence where echo peaks are acquired on the ramp.

Introduction

Ultrashort echo-time (UTE) imaging with radial acquisition is gaining more popularity since it can visualize very short T2/T2* species with less sensitivity to motion artifacts.1-3 Since the first 3D-UTE sequence took a FID-acquisition scheme with ramp sampling to minimize TE, it has been a challenging issue to correctly define the center of k-space (true-k0). In FID acquisition, a first couple of points are probable to be missing (Fig.1A), and incorrect registration of true-k0 may lead to some contrast-varying artifacts. In case of echo acquisition containing echo peaks, it is better to define true-k0 than FID acquisition although true-k0 problem still exists because echo peaks often do not agree with true-k0 due to the echo shifts caused by system imperfections such as gradient delays and eddy currents. While several methods have been proposed to solve the off-centering issues of k-space trajectory4-6, practically, the mean value of echo-peak positions is commonly used for defining true-k0. While this mean-peak method is approximate but working well to some extent, sometimes it also might cause a true-k0 problem when the position of true-k0 or echo peak is located within one sampling interval (Fig.1B). Here, we propose an asymmetric-echo-based 3D-UTE (ASE-UTE) sequence where echo peaks are acquired on the ramp. In terms of true-k0, it not only shares the benefit of echo acquisition, but also minimizes potential errors that can occur in the mean-value method due to a narrower k-space sampling interval on the ramp. Its performance was demonstrated in phantom imaging in comparison to original FID-based 3D-UTE at 3T.

Methods

Pulse Sequence Diagram: Sequence diagrams of original FID-based 3D-UTE (FID-UTE) and ASE-UTE are shown in Figure1. While FID-UTE starts FID acquisition on the ramp (Fig.2A), ASE-UTE starts asymmetric-echo acquisition on the ramp to include the echo peak (Fig.2B). A delay time (or TE) is ~10μs longer in ASE-UTE.

Experiments: All experiments were performed on a clinical 3-T (Magnetom Trio, Siemens, Erlangen, Germany) scanner with a two-channel birdcage head coil using a ACR phantom. Common scan parameters were: TR = 3ms, FOV = 400mm3, spatial-resolution = 1.25mm3, bandwidth = 121.6kHz, flip-angle = 5°, pulse-duration = 20μs. TE was 70μs and 80μs for FID-UTE and ASE-UTE, respectively.

Image Reconstruction and Data Analysis:All reconstructions were performed offline with a home-built MATLAB(ver.8.2.0;R2013b) program using fast-Fourier-transform (FFT) with gridding. For ASE-UTE reconstruction, the mean value of echo-peak positions was used to define true-k0. Phantom images of FID-UTE and ASE-UTE were compared in terms of contrast-varying artifacts and signal-to-noise ratio (SNR). For SNR comparison, main signal was calculated from five different ROIs inside the phantom. Noise standard-deviation was calculated from four ROIs in the background of the subtracted image where the reconstruction-related noise pattern was effectively removed.

Results and Discussion

Figure3 shows representative axial slices of the ACR phantom acquired using FID-UTE (A) and ASE-UTE (B). Contrast-varying artifact is seen at the periphery of the phantom in FID-UTE (A), whereas it is much reduced in ASE-UTE due to less sensitivity to true-k0 registration (B). Figure4 shows the results of SNR comparison between FID-UTE and ASE-UTE. ROIs for main-signal calculation are represented by red square boxes and ROIs for noise calculation by yellow square boxes. SNR of FID-UTE and ASE-UTE was 491.39±17.01 and 562.15±31.16 (E), respectively, and ASE-UTE showed ~14% better SNR than FID-UTE despite a slightly longer TE(10μs). Possible explanations for better SNR of ASE-UTE are: 1) Main signals inside the phantom seem to be reduced in FID-UTE due to signal pile-up artifact. 2) In ASE-UTE, more samples are acquired around the k-space center, which might contribute to the increase of signal energy.

Conclusion

In this study, we proposed a new UTE sequence (ASE-UTE) which starts asymmetric-echo acquisition on the ramp. When compared to FID-UTE, it is less sensitive to incorrect registration of true-k0 with better SNR. We expect that the proposed sequence can be more promising in UTE imaging when it is combined with up-to-date UTE imaging techniques, e.g., limiting the FOV in one direction using a slab-selective RF pulse.7

Acknowledgements

This work was supported by NRF-2017R1A2B2004944.

References

1. Irarrazabal P, Nishimura DG. Fast three dimensional magnetic resonance imaging. Magn Reson Med 1995;33:656-662.

2. Bergin CJ, Pauly JM, Macovski A. Lung parenchyma: Projection reconstruction MR imaging. Radiology 1991;179:777-781.

3. Glover GH, Pauly JM. Projection Reconstruction Techniques for Reduction of Motion Effects in MRI. Magnetic Resonance in Medicine 1992;28:275–89.

4. Brodsky EK, Samsonov AA, Block WF. Characterizing and correcting gradient errors in non-cartesian imging: Are gradient errors linear-time-invariant. Magn Reson Med 2009;62(6):1466-1476.

5. Peters DC, Derbyshire JA, McVeigh ER. Centering the projection reconstruction trajectory: Reducing gradient delay errors. Magn Reson Med 2003;50(1):1-6 .

6. Block KT, Uecker M. Simple method for adaptive gradient-delay compensation in radial MRI. In Proceedings of the 19th Annual Meeting of ISMRM, Montreal, Canada, 2011. p. 2816 .

7. Johnson, Kevin M., et al. Optimized 3D ultrashort echo time pulmonary MRI. Magnetic resonance in medicine 70.5 (2013): 1241-1250.

Figures

Figure 1. Illustration of the scheme of FID-UTE and ASE-UTE acquisition. (A) A first couple of points are probable to be missing in FID-UTE acquisition resulting in incorrect registration of true-k0. (B) In case of ASE-UTE containing echo peaks, it is better to define true-k0 compared to FID-UTE acquisition although true-k0 problem still exists.

Figure 2. Schematic pulse sequence diagram of original FID-based 3D-UTE (FID-UTE) and asymmetric-echo-based UTE (ASE-UTE). (A) For original FID-based 3D-UTE, gradient encoding begins on the gradient ramp acquiring FID after RF excitation. On the other hand, (B) ASE-UTE acquires asymmetric echo applying dephasing gradient during delay time to include echo peak on the ramp. A delay time (or TE) is ~10μs longer in ASE-UTE compared to FID-UTE .

Figure 3. Axial slices of the ACR phantom acquired using FID-UTE (A) and ASE-UTE (B). While FID-UTE image showed the typical image artifacts such as contrast variations at the periphery of the phantom (white arrow head), ASE-UTE provided a good image quality with reduced artifacts showing its tolerance to true-k0 problem.

Figure 4. The results of signal-to-noise-ratio (SNR) comparison between FID-UTE and ASE-UTE. ROIs for main-signal (red square boxes) and noise standard-deviation (yellow square boxes) were selected from each magnitude image and approximately 14% better SNR was observed in ASE-UTE in comparison with FID-UTE.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
4144