4117

CS+M: A Simultaneous Reconstruction and Motion Estimation Approach for Improving Undersampled MRI Reconstruction.
Angelica I. Aviles-Rivero1, Guy B. Williams2, Martin J. Graves3, and Carola-Bibiane Schönlieb4

1Department of Pure Mathematics and Mathematical Statistics (DPMMS), University of Cambridge, Cambridge, United Kingdom, 2Wolfson Brain Imaging Centre, Department of Clinical Neurosciences, University of Cambridge, Cambridge, United Kingdom, 3Department of Radiology, University of Cambridge, Cambridge, United Kingdom, 4Department of Applied Mathematics and Theoretical Physics (DAMTP), University of Cambridge, Cambridge, United Kingdom

Synopsis

Current research in MRI is based on using CS implications to reconstruct high-quality images from a subset of $$$k$$$-space data acquired in an incoherent manner. In this work, we introduce a mathematical framework for improving undersampled MRI data reconstruction, which we call CS+M, where M stands for motion. The significance here, and unlike existing solutions is that by modeling explicitly and simultaneously the inherent complex motion patterns, given by physiological or involuntary motion, in a CS setting, synergies in a complex variational problem are created. These synergies have clinical potentials in terms of improving image quality while reducing motion artifacts.

INTRODUCTION.

The advent of Compressed Sensing (CS) in Magnetic Resonance Imaging (MRI)$$$^1$$$ has allowed tackling the constraints associated with slow data acquisition, which is still considered a challenge in clinical practice. Current research in MRI is based on using CS implications to reconstruct high-quality images from a subset of $$$k$$$-space data acquired in an incoherent manner. In this work, we introduce a mathematical framework for improving undersampled MRI data reconstruction, which we call CS+M, where M stands for motion. The significance here, and unlike existing solutions such as in$$$^{2,3,4}$$$ is that by modeling explicitly and simultaneously the inherent complex motion patterns, given by physiological or involuntary motion, in a CS setting, synergies in a complex variational problem are created. These synergies have positive clinical potentials in terms of improving image quality while reducing motion artifacts. We demonstrate these benefits through careful visual and quantitative evaluation of a range of numerical experiments.

METHODS.

Consider an MRI acquisition in which signal measurements of target object $$$u$$$ is collected in spatial-frequency space (i.e. $$$k$$$-space) composed of $$$\mathcal{M}$$$ time samples expressed as $$$s(\mathbf{k},t_{i})=\int_{\mathbf{x}} u(\mathbf{x},t_{i})\exp^{-j\mathbf{k_{i}}^{\intercal}\mathbf{x}} d\mathbf{x} + n(\mathbf{k},t_{i}) \text{ $$$for$$$ } i=1,...,\mathcal{M}$$$ where $$$\mathbf{x}=[x,y]^\intercal$$$ and $$$t$$$ denote the spatial and temporal locations and $$$n$$$ denotes inherent noise during the acquisition. Previous expression can be rewritten in the matrix form $$$y=\mathcal{A}u+n$$$ where $$$y$$$ refers to the $$$\mathbf{k},t-$$$space measurements and $$$\mathcal{A}$$$ is the Fourier operator. Then, reconstruction of the given samples within a standard CS setting can be defined as:

$$[\text{CS}] - arg\min_{u} \frac{1}{2}\parallel\mathcal{A}u-y\parallel_{2}^{2} + \gamma\parallel\mathcal{T}u\parallel_{1}$$

where $$$\mathcal{T}$$$ is the operator promoting sparse representation and $$$\gamma$$$ the regularization parameter. Although CS offers a good reconstruction of the data, it lacks modelling of the inherent motion in MRI settings which may result in a source of artifacts during the reconstruction. One method for modelling physical motion is to use the optical flow constraint which for an image sequence $$$u(x,t)$$$ and velocity field $$$\mathbf{\boldsymbol\upsilon}(x,t)$$$ reads:

$$[\text{M}]- \nabla u \cdot\mathbf{\boldsymbol\upsilon} +\frac{\partial u}{\partial t} =\nabla u \cdot\mathbf{\boldsymbol\upsilon}+u_{t} = 0$$

encoding the intensity invariance assumption. Linking the computation of $$$\mathbf{\boldsymbol\upsilon}$$$ and $$$u$$$ in a single model by variational regularization, our CS+M model is defined as:

$$[\text{CS+M}]-arg\min_{u,\boldsymbol\upsilon} \int_{0}^{T} \bigg( \frac{1}{2}\parallel\mathcal{A}u(\cdot,t)-y(\cdot,t)\parallel_{2}^{2} + \gamma\parallel\mathcal{T}u(\cdot,t)\parallel_{1} + \delta \parallel \nabla\boldsymbol\upsilon(\cdot,t) \parallel_{1} \bigg ) dt \\\text{ subject to } \nabla u \cdot\mathbf{\boldsymbol\upsilon}+u_{t} = 0$$

The benefits of computing motion simultaneously while recovering an image sequence was first proved in$$$^5$$$ for hidden texture datasets and in the medical domain for photoacoustic imaging and X-ray tomography in$$$^{6,7}$$$ respectively.

EXPERIMENTS.

Data description. We used four datasets, described at the left side of Figure 4, for evaluating our approach. Fully sampled data were acquired and then retrospectively undersampled by 8x along ky.

Results. We started evaluating our CS+M against CS approach increasing the undersampling factor for reconstruction by different values ($$$\Phi=10,20,30$$$) and using as reference the fully sampled reconstruction. As shown in Figure 1, our CS+M significantly removed motion artifacts (see white arrows). This, together with the signal intensity profile plots from Figure 2, in which our approach is closer to the reference signal intensity, shows the positive effect of modelling simultaneously the inherent motion during the reconstruction. This is further supported by a quantitative analysis using two gold-standard image quality metrics: the structural similarity (SSIM) index and the Peak signal-to-noise ratio (PSNR). Figure 3 shows the results derived from computing the SSIM index applied on Dataset I, while the table on the right side of Figure 4 shows the complete quantitative analysis for all datasets.

Discussion. Our approach shows higher SSIM, mean SSIM and PSNR than CS, yielding to higher-quality images in comparison with CS while removing artifacts derived from the physiological motion.

CONCLUSION.

Our CS+M approach resulted on having reconstructed images with the highest quality w.r.t. CS and being the closest to the gold-standard reconstruction. Beyond a CS setting, our framework has potentials on improving well-known CS-schemes such as k-t FOCUSS or k-t SPARSE.

Acknowledgements

Support from the Centre for Mathematical Imaging in Healthcare, University of Cambridge is greatly acknowledged.

References

1. Lustig M, Donoho D and Pauly JM. Sparse MRI: The Application of Compressed Sensing for Rapid MR Imaging. Magnetic Resonance in Medicine 2007; 58:1182-1195.\noindent

2. Jung H., Sung K, Nayak K S, Kim E Y and Ye J C. k-t FOCUSS: A general compressed sensing framework for high resolution dynamic MRI. Magnetic Resonance in Medicine 2009; 61: 103–116.

3. Lustig M, Santos J M, Donoho D L and Pauly J M. k-t SPARSE: High frame rate dynamic MRI exploiting spatio-temporal sparsity. In Proceedings of the 13th Annual Meeting of ISMRM, Seattle 2006; Vol. 2420.

4. Otazo R, Candes E and Sodickson DK. Low-rank plus sparse matrix decomposition for accelerated dynamic MRI with separation of background and dynamic components. Magnetic Resonance in Medicine, 2015; 73(3), 1125-1136.

5. Burger M, Dirks H and Schönlieb C-B. A Variational Model for Joint Motion Estimation and Image Reconstruction, arXiv:1607.03255, 2016.

6. Huynh N, Lucka F., Zhang E, Betcke M, Arridge S, Beard P and Cox B. Sub-sampled Fabry-Perot photoacoustic scanner for fast 3D imaging, Proc.SPIE , 2017; Vol.10064.

7. Burger M, Dirks H , Frerking L, Hauptmann A, Helin T and Siltanen S. A Variational Reconstruction Method for Undersampled Dynamic X-ray Tomography based on Physical Motion Models, arXiv:1705.06079, 2017.

8. Wissmann, L., Santelli, C., Segars, W. P. and Kozerke, S. MRXCAT: Realistic numerical phantoms for cardiovascular magnetic resonance. Journal of Cardiovascular Magnetic Resonance, 2014; 16(1), 63.

Figures

Figure 1. Resulting reconstructions, with different undersampling factors ($$$\Phi$$$), using CS (left) and our approach CS+M (right). Zoom-in views in the middle part show more details in whichthe left (red) and right (green) ventricular endocardial borders have been outlined. White arrows show artifacts created during the CS reconstruction, these are sidestepped by our approach.

Figure 2. Signal intensity profiles retrieved from one frame of the cardiac cine dataset. Left side shows the segment (yellow line) used for generating plots on the right side. Red lines in the plots show ideal signal intensity while blue and green lines represent our CS+M and CS approaches. The results show that measurements generated from our approach are closer to the ideal one. The zoom-in views show a detailed comparison, in which we can see the positive effect of modeling simultaneously motion during the reconstruction (CS+M).

Figure 3. First part of our quantitative image quality evaluation is given by the SSIM index, taking as reference the fully sampled reconstruction and for different undersampling factors ($$$\Phi$$$). Left side shows the SSIM maps for CS and CS+M, in which white circles show a region of interest on the heart. Right side plots show the mean SSIM index to evaluate the overall image quality at each frame of the sequence. For all cases, our CS+M offered the highest SSIM and mean SSIM values which is translated in a better quality image.

Figure 4. Left side shows the description of the four datasets used to evaluate our approach. Right side shows numerical results derived from the quantitative analysis using SSIM and PSNR image quality metrics. The results show that our approach, CS+M, reported the highest values for all datasets. This is translated in clinical potentials in terms of better image quality resulted from our CS+M reconstruction.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
4117