James Smith1, Olivier Mougin1, Kingkarn Aphiwatthanasumet1, Matthew Clemence2, Andrew Peters1, Richard Bowtell1, Paul Glover1, and Penny Gowland1
1Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom, 2Philips Healthcare, Nottingham, United Kingdom
Synopsis
We demonstrate a combination of prospective motion
correction and retrospective B0 unwarping using predictive field
maps to improve the registration of EPI images. Prospective motion correction
is performed using an optical tracking camera and custom scanner code. Field
maps were generated using a multilinear fit of subject orientation using data
gathered at known orientations. Results show a higher correlation when
registering EPI data to a high-resolution anatomical, post-correction.
Purpose
Prospective motion correction can increase image quality by
locking the image geometry to the subject’s head [1].
Volume-by-volume correction is particularly useful during fMRI acquisitions in which
acquisition time is relatively short, so that inter- rather than intra- scan
movement effects dominate. Prospective motion correction maintains the phase
encoding direction relative to the head, allowing straightforward correction of
the image warping which is prevalent in EPI. However, susceptibility induced field variation within the
head is also dependent on head orientation and cannot be corrected for
prospectively yet. Here, we use prospective motion correction using an optical
tracking camera to correct for head motion during the scan and perform
additional B0 unwarping using predicted field maps for various head movementsMethods
EPI data were acquired at 5 different head positions with
prospective motion correction applied to align all volumes. Motion tracking was
performed using an optical tracking camera (Metria
Innovation, Milwaukee, WI) with a MPT marker fixed to a bite-bar. Prospective
motion correction was performed using code for geometry adjustment running on
the scanner computer. Data were acquired on a
Philips 7T Achieva scanner (Philips
Healthcare, Netherlands) with TE/TR=26/2000ms; FOV=210x210x45 mm3,
RL phase encoding and SENSE factor 2. Subjects were asked to hold 5 poses; resting, half nod, half shake, full nod and full shake
position for 5 volume acquisitions.
B0 maps were also acquired independently at 10
different head positions using a dual-echo sequence with TE/ΔTE = 5.6/1.0ms; TR 25ms and a FOV of 256x160x192 mm3. The change of head position was measured using the MPT system. Magnitude image data acquired during each
field mapping experiment were registered to the resting position in MATLAB and
the resulting transform was applied to the corresponding B0 maps.
Each voxel of the B0 map was fitted as a function of pitch, roll, yaw
and a constant, using a multilinear fit [2].
A predicted B0 map for each position in the EPI scan was then
created. The distorted EPI images were corrected using the predicted B0
map with FSL [3] [4]. Both the corrected
and uncorrected EPI images were then registered using a rigid body transform to
the anatomical image and the correlation coefficient was measured. Results
Figure 1 shows the EPI data acquired with prospective motion
correction applied. A delay in the PMC geometry update allows for the volume directly
after movement to be imaged before the geometry update is applied.
A B0 map acquired in the half nod position is
shown in Figure 2 along with the predicted field map. A total of 10 B0
maps were acquired over a range of approximately 15o of nod, 10o
of shake and 10o head roll.
Figure 3 shows the improved registration to the high-resolution
anatomical after B0 unwarping using boundary based registration.
A graph of the correlation coefficients between the
distorted and undistorted EPI images to the anatomical are shown in Figure 4.
Discussion
Figure 1 demonstrates the prospective motion correction for a
large head shake. Signal loss in Figure 1(C) is most likely due to the change in
coil sensitivity profile affecting SENSE reconstruction. Brain asymmetry as a
result of B0 inhomogeneity within the head is also shown.
The predictive field map matches closely with the true field
map in the half nod position, with typical differences in the order of 10Hz.
B0 unwarping shown in Figure 3 increases the accuracy
of the registration of the anatomical to the EPI as measured by the correlation
coefficient. This was shown to be the case in all positions with the exception
of the resting position where there was a small decrease in correlation (Figure 4).
As the size of the movement increases the correlation between the EPI and
anatomical decreases probably due to the larger warping due to gradient non
linearity, however predictive field maps are still able to offer a notable
improvement. It is likely that for large movements the linear approximation
between measured angle and actual B0 begins to break down. With
further analysis it may be possible to extend the viable range of the
predictive field maps by characterising fields at high angles of rotation and
to predict the minimum number of required field maps. Conclusion
We have demonstrated the ability to restore image quality of
EPI data using a combination of prospective motion correction and B0
unwarping. Only a limited number of B0 maps at the start of the
scanning session are required to create predictive field maps, which in turn
allows for correction of fMRI data corrupted by motion without increasing fMRI
scan time. Acknowledgements
This work was supported by funding from
the Engineering and Physical Sciences Research Council (EPSRC)
and Medical Research Council (MRC) [grant number EP/L016052/1].References
[1] J. Maclaren, B. S.
Armstrong, R. T. Barrows, K. Danishad, T. Ernst, C. L. Foster, K. Gumus, M.
Herbst, I. Y. Kadashevich and T. P. Kusik, “Measurement and correction of
microscopic head motion during magnetic resonance imaging of the brain,” PLoS
One, vol. 7, no. 11, p. e48088, 2012.
[2] A. Sulikowska, Motion
correction in high-field MRI, PhD Thesis: University of Nottingham, 2016.
[3] M. Woolrich, S.
Jbabdi, B. Patenaude, M. Chappell, S. Makni, T. Behrens, C. Beckmann, M.
Jenkinson and S. Smith, “Bayesian Analysis of Neuroimaging Data in FSL,” NeuroImage,
vol. 45, no. 1, pp. S173 - S186, 2009.
[4] P. Jezzard and R. S.
Balaban, “Correction for Geometric Distortion in Echo Planar Images from b0
Field Variations,” Magnetic resonance in medicine, vol. 34, no. 1, pp.
65 - 73, 1995.