Ilaria Liorni1, Esra Neufeld1, Sven Kuehn1, Manuel Murbach1, and Niels Kuster1,2
1IT'IS Foundation, Zuerich, Switzerland, 2ETHZ, Zuerich, Switzerland
Synopsis
Safety assessment of
active implants is defined in the technical specification ISO/TS 10974.
However, the safety of small passive implants must be assessed individually for
each device, e.g., via the ASTM F2182-11a standard test method. Computational
studies have demonstrated that the mechanism postulated for elongated implants
at 64 MHz is not applicable. We propose a new mechanistic theory of
general validity applicable across a large frequency range to describe exposure
enhancement at critical lead locations. Simulation-based assessment was successfully
compared to predictions from the mechanistic model on the impact of several
parameters (tissue dielectric properties, insulation thickness, and tip shape).
Introduction
Radiofrequency
(RF) fields generated by magnetic resonance imaging (MRI) scanners can generate
implant-related tissue heating that induces potentially irreversible damage. The
results of extensive research devoted to the study of this problem demonstrate
that elongated implants could pose a risk to wearers during MRI1-4. In
2012, the technical specification ISO/TS 109745 was established for testing
the safety of implantable parts of active implantable medical devices (AIMD),
intended to be used in patients who undergo 1.5T (64 MHz) MRI scans. However,
the safety of passive implants, (rods, clips, stents) and small abandoned leads
has not yet been systematically assessed. Each device must be tested
individually with standard testing methods such as ASTM F2182-11a6, even when a small
extension is not likely to pose a risk in patients undergoing MRI scans.
The
mechanism describing exposure enhancement for elongated leads at MRI
frequencies is based on energy pick-up by the lead and deposition at critical
implant locations, according to a measurable transfer function. It predicts a small
dependence of power deposition on the dielectric properties of tissue at the tip
of a small abandoned lead, while simulations demonstrate a near-linear
dependence on conductivity. In this study, we propose a new mechanistic theory to
support a predictive model valid for any small implant that is scalable for frequencies
of 3 kHz – 128 MHz. The mechanistic model has been used to derive a
standardizable safety assessment approach and to make verifiable predictions
about the impact of several parameters.Methods
We propose the following mechanistic picture for
partially insulated implants significantly shorter than the half-wavelength or
resonant length: current is driven through the conducting implant by either the
potential difference (electro-quasistatic (QS) regime) or the electromotoric
force (magneto-QS). Both are related to the integrated tangential field
component along the implant trajectory. Current is forced into the implant
ends, requiring that enhanced currents are generated through small cross-sections
of tissues with finite conductivity. The overall resistance can be approximated
as two series, each inversely dependent on the local tissue conductivity and
weakly affected by tip shape. Tip sharpness increases the magnitude of the locally
induced fields, but volumetric averaging or thermal diffusion mostly removes
that effect. In the absence of insulation, resistance is reduced, but current
is distributed over a larger area – hence, the insulated implant constitutes
the worst-case. Increasing the frequency results in capacitive coupling through
the insulation, reducing its effectiveness and introducing phase
considerations, limiting the applicability of the model for implants of length approaching
the tissue-specific resonance length. Validation of the mechanism was performed
within i) homogeneous setups (different dielectric properties) featuring
insulated and bare straight wires of varying length at frequencies from <10 –
128 MHz, ii) with additional local inhomogeneity near the tip, and iii)
within complex anatomical models (Figure 1). The parameters investigated
include local and spatially averaged electric (E-) fields and current density, peak
specific absorption rate (pSAR) averaged on 0.1, 1, and 10 g tissue masses, and
temperature increase (dT), as well as SAR to temperature (SAR2T) factors. The dependence
on tip shape (Figure 2) and insulation thickness was also simulated.Results
For
insulated implants and lengths smaller than the half-wavelength in tissue, the
mechanistic model predicts pSAR and SAR2T behavior and parameter dependence
(frequency, length, tissue properties, trajectory, shape) with less than 10%
variability. For longer insulated leads, a universally scalable behavior-curve
across frequencies could be extracted (Figure 3). The expected impact of insulation
and its frequency dependence could be qualitatively confirmed (Figure 4). Tip
shape not actively considered in the resistance model introduces increased
variability requiring safety margins on the order of 20% (pSAR) and 35% (SAR2T).Discussion and Conclusions
We
have developed a quantitative generalized electromagnetic safety model for small
implants at MRI frequencies that could be useful for entire passive-implant categories
or for corresponding surrogate modelling. For (partially) insulated leads shorter than
the half-wavelength in tissue, the deposited power at implant tips depends on
the voltage difference between the two tips and is a simple function of the local
tissue resistivity at the tips, weakly dependent on tip shape. A universal
curve of deposited power and temperature increase as a function of the
electrical length is obtained for insulated implants, which extends the
applicability of the model to frequencies corresponding to implant lengths
comparable to wavelength. The mechanistic model naturally translates to a
safety assessment approach (worst-case integrated tangential field trajectory
and dielectric properties can be easily identified) that is applicable to MRI
safety and at lower frequencies, e.g., for wireless power transfer (WPT) safety.Acknowledgements
No acknowledgement found.References
1.
Nyenhuis
JA, Park SM, Kamondetdacha R et al. MRI and implanted medical
devices: basic interactions with an emphasis on heating, IEEE Transactions on
Device and Materials Reliability 2005; 5(3): 467-480.
2.
Park SM, Kamondetdacha R, Amjad
A, Nyenhuis JA. MRI Safety: RF-induced
heating near straight wires. IEEE Transactions on Magnetics 2005; 41(10):
4197-4199.
3.
Yeung CJ, Susil RC and Atalar
E. RF heating due to conductive wires during MRI depends on the phase
distribution of the transmit field. Magnetic Resonance in Medicine 2002; 48:
1096-1098.
4.
Mattei E, Triventi M, Calcagnini
G, et al. Complexity of MRI induced heatingon metallic leads: Experimental
measurements of 374 configurations. BioMedical Engineering Online 2008; 7(1):
11.
5.
ISO-TS
10974:2012(E). 2012. Assessment of the safety of magnetic resonance imaging for
patients with an active implantable medical device.
6.
ASTM, “F2182-11a.
Standard Test Method for Measurement of Radio Frequency Induced Heating Near
Passive Implants During Magnetic Resonance Imaging,” 2011; pp. 1–8.