4065

Image-Based MR Safety Evaluation of Heating in Implanted Devices
Samuel Rob Guyer1, Ken Sakaie2, Se-Hong Oh2,3, and Stephen Jones2

1School of Medicine, Case Western Reserve University, Cleveland, OH, United States, 2Imaging Institute, The Cleveland Clinic, Cleveland, OH, United States, 3Department of Biomedical Engineering, Hankuk University of Foreign Studies, Yongin, Republic of Korea

Synopsis

A thermal contrast agent with high sensitivity, thulium 1,4,7,10-tetramethyl-1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetate (TmDOTMA-), was investigated for use in measuring the three-dimensional heating profile surrounding implantable devices with a goal of simplifying the safety assessment process. The technique was first demonstrated by imaging a vertical temperature gradient and was subsequently used to measure the heating profile surrounding a copper wire. We found gross agreement of this technique with concurrent fiber-optic temperature probe measurements, with the TmDOTMA- measurement complicated by a number of technical challenges and imaging artifacts.

Introduction

Heating of implanted devices under magnetic resonance imaging (MRI) carries a risk of damage to surrounding tissue, especially within the brain, which necessitates rigorous safety testing before such devices can be used in the MRI setting1,2. The dearth of MRI-approved devices negatively affects research as well as clinical diagnostic and treatment options. One reason devices remain unapproved for MRI use is that the safety validation process is time consuming and expensive, requiring many point temperature measurements with fiber-optic temperature probes (FOPs) which are ill-suited to evaluate the three-dimensional (3D) nature of the heating1,3. We aim to simplify the safety assessment of implanted devices by directly measuring the 3D heating profile of a brain implant using thulium 1,4,7,10-tetramethyl-1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetate (TmDOTMA-), a thermally sensitive contrast agent, in an extension of previous work on an external device4.

Methods

MRI images were acquired on a Siemens 7 tesla whole-body scanner (Magnetom; Siemens Healthcare, Erlangen, Germany). TmDOTMA- was obtained from Macrocyclics (Plano, TX) and prepared in two 40 mL vials to a concentration of 4 mM using normal saline. TmDOTMA- signal was acquired by a manual shift of the center frequency -106 ppm (-31.5 kHz) from water using a spoiled gradient echo (GRE) pulse sequence. A time bandwidth product of ≥5.4 was necessary in order to excite the sample over the experimental temperature range due to the relatively large temperature coefficient. Relative temperatures were calculated using the proton resonance frequency shift (PRF) shift equation:

$$\Delta T=\frac{\Delta\phi}{2\pi\gamma{B_{0}}cTE}$$

where $$$\Delta\phi$$$ is the difference in phase between two voxels, $$$\gamma$$$ is the gyromagnetic ratio, $$$B_{0}$$$ is the magnetic field strength, $$$c$$$ is the temperature coefficient of the frequency shift (water: -0.01 ppm / °C, TmDOTMA-: 0.6 ppm / °C), and $$$TE$$$ is the echo time in milliseconds5,6. Analysis performed in MATLAB (MathWorks, Natick, MA, USA).

As a preliminary demonstration, a vertical temperature gradient was created in a single vial of TmDOTMA- by placing the vial on a block of chilled agar and making repeated scans while allowing the vial-agar system to thermally equilibrate. FOPs were placed at the top and bottom of the vial.

Subsequently, heating was induced in a 65 cm 22 AWG solid-core copper wire placed in a TmDOTMA- vial while simultaneously measuring the resulting 3D temperature profile. Heating was accomplished using the radiofrequency (RF) pulse (500 ms pulse, 7.0 ms repetition time, 22° flip angle, 128 averages, 30.6 min acquisition). Heating was limited to the tip of the wire, located in the center of the solution, by leaving the original insulation on all but the final 2 mm. Another TmDOTMA- vial lacking a wire was also present as a control. In the experimental vial 4 FOPs were placed: side-top, side-middle, side-bottom, and wire tip. A single FOP was placed in the center of the control vial.

Results

Using TmDOTMA- in aqueous solution we have successfully imaged both a thermal gradient (Figure 1) and the 3D heating profile induced by the MRI scan in a copper wire (Figure 2). Temperature readings obtained by concurrent measurement with FOPs are on the whole consistent, but not identical, to those calculated from the PRF shift equation.

Discussion

While we successfully imaged gross and localized thermal gradients using TmDOTMA- there are a number of limitations to the data obtained thus far. First and foremost is a lack of quantitative agreement of the calculated temperature values with the FOPs. Second is the model system used, which is not suitable for robust characterization of the technique in human tissue. Finally, we did not use an implant with clinical utility. Ideally, a realistic head phantom and agar gel would be used along with a deep-brain stimulation electrode.

There are a number of factors which make this measurement challenging and require further investigation before the technique can be used for precision safety measurements. Most prominent are a number of artifacts, such as pi bounces (from changes in magnetic susceptibility) and signal pileup (known to occur with the GRE pulse sequence) at material interfaces, which convolute the TmDOTMA- signal. Additionally, the expense of TmDOTMA- necessitated the dilute and small-volume samples used here and contributes to a weak signal that limits flexibility in scanning parameters. The combination of complicating factors interferes with a quantitative analysis of the phase signal and makes correlation with FOPs challenging.

Conclusion

While TmDOTMA- shows promise as a temperature-sensitive contrast agent to investigate the 3D heating profile surrounding implantable devices there are a number of technical challenges to the measurement that need to be overcome before it can be used in the assessment of device safety.

Acknowledgements

This work was supported by the Imaging Institute, Cleveland Clinic. Authors acknowledge technical support by Siemens Medical Solutions.

References

  1. Finelli DA, Rezai AR, Ruggieri PM, et al. MR imaging-related heating of deep brain stimulation electrodes: in vitro study. AJNR Am J Neuroradiol. 2002;23:1795-802.
  2. Henderson JM, Tkach J, Phillips M, Baker K, Shellock FG, and Rezai AR. Permanent neurological deficit related to magnetic resonance imaging in a patient with implanted deep brain stimulation electrodes for Parkinson's disease: case report. Neurosurgery. 2005;57:E1063.
  3. Kahan J, Papadaki A, White M, et al. The safety of using body-transmit MRI in patients with implanted deep brain stimulation devices. PLoS One. 2015;10:e0129077.
  4. Graedel NN, Polimeni JR, Guerin B, Gagoski B, Wald LL. An anatomically realistic temperature phantom for radiofrequency heating measurements. Magn Reson Med. 2015;73:442-50.
  5. Ishihara Y, Calderon A, Watanabe H, et al. A Precise and Fast Temperature Mapping Using Water Proton Chemical Shift. Magn Reson Med. 1995;34:814-23.
  6. Pakin SK, Hekmatyar SK, Hopewell P, Babsky A, Bansal N. Non-invasive temperature imaging with thulium 1,4,7,10-tetraazacyclododecane-1,4,7,10-tetramethyl-1,4,7,10-tetraacetic acid (TmDOTMA-). NMR Biomed. 2006;19:116-24.

Figures

Vertical temperature gradient with time progressing from left to right. Vial signal is skewed in the frequency-encoded horizontal axis due to the pixel bandwidth being less than the change in TmDOTMA- frequency with temperature. Vial appears to lengthen as the bottom warms and enters the bandwidth of the excitation pulse. (Top) Magnitude. (Second from top) Phase. (Second from bottom) Phase masked to magnitude; a subset of pi bounces are marked with arrows. (Bottom) Phase-unwrapped and false-colored images demonstrate an evolving thermal gradient.

3D heating profile resulting from heat induced at the tip of an insulated wire using the RF pulse of the MRI scan. A single representative slice is shown for clarity. (Image Left) Control vial, no wire. (Image Right) Experimental vial with wire. Red dashed boxes indicate regions used for the PRF shift calculation and correspond approximately to the location of the associated temperature probe. A temperature difference was calculated for each probe/region relative to the side-top probe/region of the experimental vial. There is a lack of quantitative agreement between FOP and PRF shift calculated differences.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
4065