Mikhail Kozlov1, Roland Müller 1, Andre Pampel1, Benjamin Kalloch1, Nikolaus Weiskopf1, and Harald E. Möller1
1Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
Synopsis
We calculated the RF-induced power deposition at
3T due to the presence of tDCS equipment using an approximation of the human
body by a single tissue numerical model. Influences of the tDCS lead including
a built-in serial resistor on the power deposition were evaluated.
Our case study provided strong
evidence that the tDCS setup has a relevant impact on power deposition in
proximity of the electrode edges at 3T. These particular conditions are not
explicitly considered in the procedures that are used for SAR monitoring on the
MRI scanner. The peak voltage across the serial resistor has to be carefully
considered in terms of sufficient electrical strength to safely prevent sparks.
Introduction:
An increasing number of MRI investigations to study
the human brain employ multi-modal setups, where additional devices are used to record complementary information or to manipulate brain
states. In the case of transcranial direct current stimulation (tDCS) this requires
additional sensor setups that entail two external wires with an electrode that is
in contact with the human skin. The wire enters the radio-frequency (RF) coil’s
effective exposure volume and operates as an antenna whose performance depends
on the relative positioning of the wire and human body as well as the electrical
contact of the electrode with the skin. In general, RF-induced power deposition
(p)
may heat up the area where the tDCS electrode contacts the skin. In current
commercial tDCS devices, a serial resistor is used to reduce this unwanted (and
potentially harmful) temperature rise.
Although computational power is increasing constantly,
it remains a challenge to perform 3D electromagnetic (EM) simulations of tDCS
equipment with high resolution human body models located inside the MRI whole-body
coil. Goals of the current study
were: (a) to calculate the RF-induced power deposition at 3T (123.2 MHz) due to
presence of tDCS equipment labeled MR “conditional” using an
approximation of the human body by a single tissue numerical model; (b) to
evaluate the influences of the serial resistor as well as the tDCS lead
positioning on the power deposition.Methods:
The tDCS setup consisted of two electrodes, two leads, and a metal connection box located 410mm away from the coil shield. The serial resistors integrated in the leads were located 100 mm away from the electrodes. Three resistor values were simulated: 1 mΩ, 5kΩ, and 1GΩ to simulate conditions of a short (potential manufacturing fault), normal operation, and an open connection (resistor failure after long-term operation), respectively. To ensure proper electrical contact between the tDCS electrode and the skin as well as high quality numerical mesh generation, the triangular faces of the skin object in areas around both electrodes were merged into a single face (see Figs. 1a and b). The skin object of the Visible Human Project (VHP) model [1] was used with a scaling factor 0.9 and εr=66.5 and σ=0.518S/m. The model head was positioned in the coil’s isocenter. The whole-body coil was a 3T (123.2 MHz) 16-rung high-pass
birdcage with inner diameter
615mm; total length 480mm (Fig.2). Two lead positions were simulated: on axis of the scanner bore and shifted towards the edge of the patient table.
Results and Discussion:
Results obtained without consideration of tDCS
equipment were consistent with the literature [2]: B1+ was rather homogeneously
distributed across the head and the power deposition showed a maximum in the
neck region (Fig. 3). Upon adding the tDCS setup with the lead directed along the
magnet axis, substantial power deposition was observed in close proximity of
the tDCS electrode edges for all values of the serial resistor (Fig. 3). Unsurprisingly,
the B1+ disturbance was highest for a short (R=1mΩ). For normal operation (R=5kΩ),
repositioning of the tDCS lead resulted in
a small variation of the power deposition in proximity of the electrode edges
(Fig. 5). Assuming a (pulsed) peak value of 30kW of the total transmit
power (which can be generated by the scanner’s transmitter) yielded voltages
across the serial resistor up to 850 V for normal tDCS operation (R=5kΩ) and up to 1.4 kV for an open
connection (R=1GΩ). For a whole-body
SAR level of 4 W/kg, average voltages across the serial resistor were 130V and
225 V for R=5kΩ and R=1GΩ, respectively. The results indicated that the 1206-sized
resistors with a voltage rating of 200V do not provide sufficient electrical
strength for safe operation under all possible conditions. Due to the similarity of the power
deposition in proximity of the electrode edges for all investigated conditions,
we conclude that the tDCS electrodes and the straight segments of the leads
between them and the serial resistor predominantly determine the power
deposition.Conclusion:
Our simulations provide strong evidence that the
tDCS setup has a relevant impact on power deposition in proximity of the
electrode edges at 3T. These particular conditions are not explicitly
considered in the procedures that are used for SAR monitoring on the MRI
scanner. The peak voltage across the serial resistor has to be carefully
considered in terms of sufficient electrical strength to safely prevent sparks.
Future work should include investigations of RF-induced heating and experimental
validation of simulation results. While our results were derived assuming
realistic conditions that may occur with commercially available equipment, it
is to note that they cannot be readily
generalized because only a single coil geometry and a single tDCS device were
included in our investigation.Acknowledgements
No acknowledgement found.References
[1] J. Yanamadala, G. M. Noetscher, V. K. Rathi, S. Maliye,
H. A. Win, A. L. Tran, X. J. Jackson, A. T. Htet, M. Kozlov, A. Nazarian, S.
Louie, S. N. Makarov, “New VHP-Female v. 2.0 Full-Body Computational Phantom
and Its Performance Metrics Using FEM Simulator ANSYS HFSS,” 37th Annual Int.
Conf. of the IEEE Eng. in Medicine and Biology Society, Milano, Italy, Aug.
25-29, 2015, pp. 3237-3241.
[2] Z. Yu, X. Xin, and C. M. Collins, “Potential for
High-Permittivity Materials to Reduce Local SAR at a Pacemaker Lead Tip During
MRI of the Head With a Body Transmit Coil at 3 T,” Magnetic Resonance in
Medicine 78:383–386 (2017)