Mikhail Kozlov1 and Wolfgang Kainz2
1Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany, 2U.S. FDA, CDRH, Office of Science and Engineering Laboratories, Division of Biomedical Physics, Silver Spring, MD, United States
Synopsis
This case study investigated RF-induced heating
of straight and helix leads at 127.7 MHz obtained with the lead electromagnetic
model (LEM) and direct 3D electromagnetic and thermal co-simulations. A large
set of incident electric fields was generated in a phantom by an array of four
antennas with varying spatial positions and sources. LEM was a suitable
approach for predicting temperature
in close proximity to the end face of lead tip. However the variance of the
fitted values and observed values was rather high for temperature at other locations around the lead tip.
Introduction:
One
approach to evaluate RF-induced heating near an active implantable medical device (AIMD) lead is the lead
electromagnetic model (LEM) prescribed in Clause#8 of ISO/TS_10974_Ed2 [1]. Using
the transfer function (S(l)) [2] the LEM relates the incident tangential electric
field (Etan) along lead trajectory to the RF power deposition (P) and temperature rise at a given point in space (ΔTp) due to presence of the lead. In the LEM
$$P = A\cdot\left[\int_{0}^{L} S(l)\cdot E_{tan}\cdot dl \right]^{2}$$
$$\triangle T_{p} = A_{T_{p}}\cdot\left[\int_{0}^{L} S(l)\cdot E_{tan}\cdot dl \right]^{2}$$
where: A and
ATp are the calibration
factors of the LEM for P and ΔTp, respectively, and L is the lead length.
Both calibration
factors can be assessed using a linear regression analysis
of P and ΔTp versus $$$P_{lem} = \left[\int_{0}^{L} S(l)\cdot E_{tan}\cdot dl \right]^{2}$$$ obtained for a set of
diverse non-uniform Etan(l). The associated regression coefficients R2A
and R2ATp are the quotient of the variances of the fitted
values and the observed values of the dependent variable. Etan(l) disturbance depends on the proximity between the
implant segments and the construction of the implant. The smallest disturbance
of Etan(l) occurs for the straight lead trajectory allowed the most reliable estimation of A and ATp.
Methods:
Our test objects were titanium alloy helix and
straight wire insulated lead (Fig. 1).
The leads were
positioned in the middle of a box filled with a medium with εr=78 and σ=1.2S/m. 120 Etan(l) along the lead trajectory was generated
by four antennas located along one side of the box by varying the relative antenna positions as
well as the amplitude and the phase of each antenna source. P was calculated by integrating the
volume loss density (VLD) around the lead tip. Additionally point VLD values (VLD3D_EM) were obtained at axial location 1mm from the
tip. S(l) for the tip
electrode was calculated using the reciprocity
approach described in [3]. 3D-EM simulations were performed at 127.7MHz using
ANSYS HFSS-2014 (ANSYS, Canonsburg, USA). 3D temperature distributions
after 1, 10, 60, 360 and 900 seconds continuous excitation were calculated using
the ANSYS-15. ΔTp was obtained at 14 locations. The variable mesh was generated in each solver
independently. The mesh size in the area of the hot spot was less than 0.1
mm. The 3D-EM mesh
adaptation procedure with 30% increase of the mesh elements was stopped if the
variation of P or ‖S‖max between two consecutive meshes was less than 3%. The convergence of the thermal simulations was
obtained by the manual reduction of the mesh size between two consecutive runs
of the thermal solver for a given 3D-EM simulation. The mesh reduction was
stopped when the difference between maximum induced temperature rise (ΔTmax) of two consecutive runs was less than 3%.Results and Discussion:
Variation of antenna location, source amplitudes, and
phases is a robust approach to generate different amplitude and phase
distributions. R2A was 0.9714 and 0.5759 for straight and helix lead, respectively. The latter means that this linear regression was
rather poor. However the linear regression for VLD3D_EM was very good: 0.9876. Thus the LEM can predict VLD at axial locations 1mm
from the helix lead tip. This was strong evidence of the reliable
calculation of TF for straight and helix leads. For helix lead diversity of the pathways influenced
on both VLD and temperature distribution. For the helix lead R2ATp depended substantially on the duration of the RF-induced
heating and also significantly on the location of temperature
probe. For the straight lead the influence of R2(ATp) on the location of temperature probe was significantly
smaller but still noticeable. The ratio AVLD/A was equal 5.98 and 12.33 for straight
and helix leads, respectively. It provided strong evidence that power deposition distribution depends not only on lead tip geometry
and the surrounding medium, but also on lead wire geometry.Conclusion:
LEM was a suitable
approach for predicting VLD or temperature in close proximity to the end
face of lead tip. However for the helix lead the variance of
the fitted values and observed values was rather high for ATp at other locations around the lead tip as well as for A. The variation of the power
deposition and the RF-induced heating profiles in the proximity of the lead tip due to
Etan(l)
cannot be readily
generalized to the wider range of lead geometries and tissue dielectric
properties, because we investigated only one set of Etan(l),
two lead geometries, and only one set of tissue dielectric properties.Disclaimer:
The mention of commercial products, their
sources, or their use in connection with material reported herein is not to be
construed as either an actual or suggested endorsement of such products by the
Department of Health and Human Services.Acknowledgements
No acknowledgement found.References
[1] Technical specification ISO/TS 10974, “Assessment of the safety of
magnetic resonance imaging for patients with an active implantable medical
device”, 1st edition 2012.
[2] S-M. Park, K.
Kamondetdacha, and J. A. Nyenhuis, “Calculation of MRI-induced heating of an
implanted medical lead wire with an electric field transfer function”, J. Magn.
Reson. Imaging, 26(5), 2007, 1278–1285.
[3] S. Feng, R. Qiang, W. Kainz, and J. Chen, “A technique to evaluate
MRI-Induced electric fields at the ends of practical implanted lead,” IEEE Transactions on Microwave Theory and
Techniques, 63(1), 2015, 305-313.