Jonathan Vappou1, Paolo Cabras1, Kisoo Kim1, Pramod Rao1, Afshin Gangi1,2, and Elodie Breton1
1ICube CNRS / University of Strasbourg, Strasbourg, France, 2Department of Interventional Imaging, Strasbourg university hospital, Strasbourg, France
Synopsis
This study introduces the use of interventional
Magnetic Resonance Elastography (iMRE) for monitoring HIFU ablations, using the
acoustic radiation force as a means for generating the shear waves necessary
for MRE directly from the HIFU focus. This method allows for monitoring tissue
elasticity and temperature in real time during the ablation. Its feasibility is
illustrated in vivo on porcine muscle. Tissue was found to stiffen
significantly (+125%) when the temperature increased, and changes in tissue
stiffness were found to be irreversible. These findings suggest that tissue
stiffness may be an interesting biomarker, complementary to thermal dose, to
monitor HIFU ablations.
Introduction
Tissue elasticity has
been shown to be a promising biomarker for monitoring thermal ablations 1-3.
Interventional MR Elastography (iMRE) has been developed recently in order to address
this particular challenge 3. iMRE allows computing both tissue elasticity and
temperature in real time, and includes specific features such as: (1)
Interventional-friendly, compact mechanical exciters that use interventional
needles as source for shear waves, (2) Interactive, real-time MRE pulse
sequence that allow simultaneous MR Thermometry and MRE measurements, and (3)
Online reconstruction of elastograms and temperature maps in real time (~ 1
Hz). The capability of the method for monitoring MR-guided laser ablations has
already been demonstrated in vivo in swine liver 3-4. In this study, the iMRE
method is adapted to the monitoring of High Intensity Focused Ultrasound (HIFU)
ablations. The main originality of this study is the use of the HIFU transducer
as a mechanical exciter through the use of the acoustic radiation force, which generates
shear waves propagating directly from the focal spot while the ablation is
being performed. Feasibility is demonstrated in vivo in porcine muscle tissue.Methods
Experiments are performed using a MR-compatible HIFU
system (Image Guided Therapy, France) with a 256-element transducer driven at 1
MHz. The signal transmitted to the transducer is modulated by a square wave at 50
Hz (duty cycle 50%), allowing for simultaneous heating and mechanical
excitation through an oscillating pseudo-harmonic acoustic radiation force. This
modulation process is depicted in Figure 1 together with the general ablation
protocol used in this study. The HIFU transducer is positioned using an
orientable MR-compatible structure (Figure 2). Acoustic coupling is realized
through the use of a degassed water-filled balloon. MRI experiments are
conducted in a 1.5T MRI system (MAGNETOM Aera, Siemens). A spoiled, interactive,
interventional MRE-GRE sequence is used for encoding displacements resulting
from this excitation. Main MRI parameters are: TR/TE 20ms/13ms, bipolar
motion-sensitizing gradients at a frequency of 90 Hz, FOV 400 mm × 400 mm, 128
px × 128 px, 3 phase shifts, total acquisition time per slice 2.5 s. Both
elasticity and temperature maps are reconstructed simultaneously using a Local Frequency
Estimation-based approach and the Proton Resonance Frequency (PRF) method,
respectively4. A HIFU heating phase of 150 s at an acoustic power of 285W was followed by a relaxation phase of 250 s. During the relaxation
phase, isolated, short MRE measurements (2 acquisitions with opposite bipolar
gradients, total acquisition time = 5 s) were performed every 60 s at lower
acoustic power (200W), to avoid any additional heating. Experiments were
performed in vivo in the thigh muscles of a swine, under the approval of the
local ethics committee.Results
Shear waves could be clearly observed emerging from
the focal region (Figure 3). Tissue shear modulus was found to increase
consistently from ~4 kPa up to ~9 kPa at the end of the ablation. As expected,
tissue temperature decreased back to 37°C after HIFU was turned off, while
tissue elasticity remained high, with an average value of 9.5 kPa after
ablation (Figure 4).Discussion
This study introduces the use of interventional MRE
for monitoring HIFU ablations, using a dedicated MRE protocol that relies on
the acoustic radiation force for generating shear waves. Tissue elasticity was
found to increase significantly during HIFU ablation, and such increase was
found to be irreversible. This study suggests that simultaneous MRE/MR
Thermometry may be an interesting tool for monitoring HIFU ablations in real
time. However, further investigation is needed to assess the value of tissue
elasticity as a biomarker. Ongoing work is focused on comparing tissue
elasticity and thermal dose to histology in order to establish the sensitivity
of such biomarkers. Beyond
its potential for HIFU monitoring, the method presented in this study may also be
interesting for generating localized shear waves in deep-lying organs for MRE,
as an alternative to external, conventional excitation systems that are
associated with limitations in terms of shear wave penetration. However,
further studies are needed to ensure that this excitation method complies with
regulatory requirements in terms of acoustic energy that is deposited.Conclusion
These preliminary results suggest that tissue
elasticity is an interesting biomarker directly related to permanent,
irreversible tissue damage, as opposed to temperature. This study illustrates a
new method that allows for simultaneous HIFU ablation and MRE/MR Thermometry
measurements without the use of additional excitation devices. Acknowledgements
This study was partly funded by French excellence program Labex CAMI
(ANR-11-LABX-0004)References
1. Mariani A, Kwiecinski W, Pernot M et al. Real
time shear waves elastography monitoring of thermal ablation: in vivo
evaluation in pig livers. J. Surg. Res. 2014;188:37-43.
2. Chen J, Woodrum
D, Glaser K et al. Assessment of in vivo laser ablation using MR elastography
with an inertial driver. Magn. Reson. Med. 2014;1:59-67.
3. Corbin N,
Vappou J, Breton E et al. Interventional Magnetic Resonance Elastography for
MRI-guided percutaneous procedures. Magn. Reson. Med. 2015;75(3):1110-1118.
4.
Corbin N, Vappou J, Breton E et al. Interventional Magnetic Resonance
Elastography for MRI-guided percutaneous procedures. ISMRM, Toronto, Canada,
2015.