Rolf Pohmann1, Sathiya Raju1,2, and Klaus Scheffler1,3
1Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 2Furtwangen University, Schwenningen, Germany, 3Biomedical Magnetic Resonance, University Tübingen, Tübingen, Germany
Synopsis
Ultra-high field has the potential to improve the
quality of 31P spectroscopic applications due to the increased SNR
and spectral dispersion. In addition, previous studies have reported a decreasing
T1 with increasing field strength, which would further contribute to
an improved quantification quality. In this study, we have measured the
longitudinal relaxation time T1 in the visual cortex of healthy
human subjects at 9.4 T, using an inversion-recovery technique. The values obtained were consistently lower than published data at 7 T, confirming observations of a decreasing T1 with field strength.
Introduction
31P spectroscopy offers the unique possibility
to directly observe the major metabolites involved in the energy metabolism in
vivo and non-invasively. However, the low NMR sensitivity of the 31P
nucleus and the low concentration of the phosphorus metabolites, together with
their short T2 and long T1 values, make 31P
spectroscopy challenging, requiring large voxel sizes and long scan durations
for often still noisy measurements. Ultra-high field has the potential to
improve the quality of 31P spectroscopic applications due to the
increased SNR and spectral dispersion. In addition, previous studies have reported
a decreasing T1 with increasing field strength, which would further
contribute to an improved quantification quality. In this study, we have
measured the longitudinal relaxation time T1 in the visual cortex of
healthy human subjects at 9.4 T to verify the T1 reduction at higher
fields and to enable sequence optimization and metabolite quantification.Methods
Five healthy
subjects were examined at a field strength of 9.4 T, using a home-built 31P
array coil, consisting of four 1H dipole antennae for shimming and
localizer images, four 31P transmit loops and 27 31P
receive coil elements positioned on a tight fitting helmet1. Since we are only interested in the visual
cortex, only signals from the two receive coil elements just below the visual
cortex were used. T1 is measured using an inversion-recovery technique
consisting of a simple pulse-acquire sequence, preceded by an inversion pulse.
Seven saturation slices around the visual cortex are used to further define a
region of interest and suppress signal contributions from outside. The
positions of the saturation slices and the ROI are shown in Fig. 1. For inversion, a 15 ms TR-FOCI pulse was
used, while the FID was acquired after a 0.3 ms nonselective block pulse. 512
samples were recorded within 51 ms with a bandwidth of 10000 Hz. Seven
inversion times TI between 61 ms and 14 s were applied, each
measured with 20 averages (except for the TI = 14 s measurement, for
which 40 averages were acquired) and a TR of 15 s.
For analysis, the
signals from the two coil elements were phased and added manually and the AMARES
algorithm in the jMRUI software package was used to obtain signal amplitudes.
The TI = 14s scan, which was acquired with a larger number of
averages and thus has a high SNR, was used to determine the prior knowledge for
optimal fitting. The amplitudes found in jMRUI are then used to fit an
exponential relaxation curve, taking into account that the relaxation delay is
not always larger than 5 T1.
Results
For all subjects,
all spectra had excellent quality and could be fitted in jMRUI with high
precision. Spectra from one subject and all seven inversion times are shown in
Fig. 2. The resulting values for the T1 times of 9 metabolite peaks
in comparison to previously published 7T data2 are shown and listed in Fig. 3.
In addition, the linewidths obtained from the TI = 14 s measurements
can be used to determine the T2*-values of the 31P
metabolites (Fig. 4).Discussion
Due to the high SNR, the increased spectral
dispersion of 31P metabolite peaks at 9.4 T and the correspondingly
high spectral quality, T1 times could be determined with high
accuracy in all subjects. The T1 values found here were lower than
previously published data from 7 T, which might be attributed to a strong contribution
of chemical shift anisotropy to the relaxation process. A literature review
over a larger range of field strengths, including also data from rat brain and muscle
spectroscopy, also points towards a decrease of the T1 of PCr with
increasing field strength (Fig. 5).Conclusion
At 9.4 T, 31P
spectra with excellent quality can be acquired within reasonable time. The lower
T1 values at higher field strength are another advantage of ultra
high field for 31P spectroscopy.Acknowledgements
No acknowledgement found.References
1. Mirkes
C, Shajan G, Chadzynski G, Buckenmaier K, Bender B, Scheffler K. (31)P CSI of
the human brain in healthy subjects and tumor patients at 9.4 T with a
three-layered multi-nuclear coil: initial results. Magn Reson Mater Phy
2016;29(3):579-589.
2. Ren J, Sherry AD, Malloy CR.
(31)P-MRS of healthy human brain: ATP synthesis, metabolite concentrations, pH,
and T1 relaxation times. NMR Biomed 2015;28(11):1455-1462.