Chu-Yu Lee1, In-Young Choi1,2,3, Jean C Dinh4, William M Brooks1,2, Steven J Leeder4,5, and Phil Lee1,3
1Hoglund Brain Imaging Center, University of Kansas Medical Center, Kansas City, KS, United States, 2Department of Neurology, University of Kansas Medical Center, Kansas City, KS, United States, 3Department of Molecular & Integrative Physiology, University of Kansas Medical Center, Kansas City, KS, United States, 4Division of Clinical Pharmacology, Toxicology, and Therapeutic Innovation, Children's Mercy Hospital, Kansas City, MO, United States, 5Department of Pediatrics, Children's Mercy Hospital, Kansas City, MO, United States
Synopsis
Fluorine MR Spectroscopy (19F MRS) allows in vivo quantification of fluorine-containing antipsychotic and antidepressant drug concentrations in the brain. To detect the low concentration of the drugs (~ 5-30 µM) in the brain, it requires multiple repeated acquisitions to increase SNR and the scan time is relatively long. Therefore, it is important to ensure consistent frequency alignment across the repeated acquisitions. However, MR system instability induces drifts of the scanner frequency, particularly following MR scans with a high gradient duty cycle. Previous frequency correction methods for 1H MRS require internal reference signals, such as under- or un-suppressed water, and cannot be applied to 19F MRS, where the SNR of 19F signals is low at each repeated acquisition. The purpose of this study is to investigate the feasibility of using an external reference for retrospective frequency correction in 19F MRS.
INTRODUCTION
Fluorine MR Spectroscopy (19F MRS) allows in vivo
quantification of fluorine-containing antipsychotic and antidepressant drug
concentrations in the brain. 1-10 Due to the low drug concentration in the brain (~ 5-30 µM), it is necessary to acquire multiple repeated acquisitions
to increase SNR and the scan time is relatively long. Therefore, it is
important to ensure consistent frequency alignment across the repeated acquisitions.
However, MR system instability induces drifts of the scanner frequency,
particularly following MR scans with a high gradient duty cycle. 11,12
Previously proposed frequency correction methods for 1H MRS utilize internal
reference signals, such as under- or un-suppressed water. 11-14 However, the same approach of using internal reference signals cannot be applied to 19F MRS, where the SNR of 19F
signals is too low to reliably determine the frequency. The purpose of this study is to
investigate the feasibility of using an external reference for retrospective frequency
correction in 19F MRS.METHODS
All 19F MRS experiments were performed on a Skyra
3T scanner (Siemens, Erlangen, Germany) using a dual-tuned (1H and 19F)
quadrature volume head coil. The calibration of the transmitter power and
shimming were performed using 1H measurements and their values were
applied to 19F measurements. 19F MRS measurements were
performed on a phantom with a
similar size (1.5 L) and shape as human head. The phantom contained 0.1
mM pimozide dissolved in dimethylsulfoxide
(DMSO). Additionally, a 5 ml cylindrical vial of trifluoroethanol (TFE) was attached to the
head coil and was used as an external reference. A non-localized RF pulse of 500 µs
was used to excite the 19F signals. The flip angle was adjusted to
be the Ernst angle: 60° based on the measured T1 relaxation time of
pimozide: 1422 ms on a phantom and TR of 1000 ms. Sequence parameters were TR = 1000 ms, number of
samples = 2048, and spectral width = 12 kHz. A
30-min diffusion weighted imaging (DWI) experiment was performed before 19F
MRS measurements to create gradient heating-induced frequency drifts. To investigate the behavior of frequency
drifts on 19F and 1H, 19F MRS
measurements were interleaved with 1H MRS measurements with a total
number of averages of 3000 (Fig. 1). Frequency drifts of 19F were determined
from the peak position of the spectrum and were measured from the external reference (TFE). Frequency drifts of 1H were measured from the solution phantom (DMSO
solvent). Frequency correction for 19F MRS was performed by multiplying the FID signal
by a linear phase term. The linewidth and area under the curve (AUC) of the
pimozide spectrum were evaluated.RESULTS
Frequency drifts
of 19F and 1H MRS measurements exhibited a similar
temporal variation (Fig. 1), but frequency drifts of 19F were
smaller (19F: 6.2 Hz/min versus 1H: 6.5 Hz/min). The
differences resulted from the difference in their gyromagnetic ratios: γ19F
= 40.05 MHz/T versus γ1H = 42.58 MHz/T. Without frequency corrections, the spectrum from the
averaged pimozide signals showed a broad linewidth of 0.4 ppm and a frequency
drift of 1.8 ppm (Fig. 2b). The AUC was 7.8 × 10-5. With the
frequency correction using the external reference, the linewidth was reduced to
0.2 ppm (Fig. 2c). The AUC was increased to be 1.9 × 10-4. DISCUSSION
Detection
of the low concentration fluorine-containing drugs in the brain using 19F MRS requires a
long scan time. This study demonstrated that gradient heating-induced frequency
drifts during the long scan time resulted in degraded spectral quality and
reduced AUC, which are associated with reduced detection sensitivity and
increased errors in quantification of drug compounds. The external reference has
been used in previous 19F MRS studies as a reference for chemical
shifts and as a standard for normalization to reduce the effects of different
coil loading. 2,4,7,8,10 This
study demonstrated that the external reference can also be used to correct frequency
drifts retrospectively, and that those effects of frequency drifts were notably
reduced (Fig. 2). Gradient-induced frequency drifts are spatially invariant
(Fig. 1). However, frequency drifts resulting from subject motion are spatially
varying, and the correction would require an internal reference. For a
dual-tuned coil, 1H signals may be used as an internal reference to
correct frequency drifts induced by gradient heating and subject motion in
19F MRS, although many practical issues need to be resolved. In
conclusion, this study demonstrated the feasibility of using external reference
for retrospective frequency correction in 19F MRS that is essential
for consistent detection sensitivity and accurate quantification of fluorine-containing
drug compounds. Acknowledgements
This study was
partly supported by NIH (UL1TR000001, P20GM103418).
The Hoglund Brain Imaging Center is supported by the NIH (S10RR029577) and the
Hoglund Family Foundation.References
[1]. Bartels et al.,
Psychiatry Res. (18), 197-201, 1986. [2]. Karson et al., Psychiatry Res. (45),
95-104, 1992. [3]. Renshaw et al., Am J Psychiatry (149), 1592-4, 1992. [4]. Komoroski et al., MRM (31), 204-11, 1994.
[5]. Bartels et al., J Neural Transm Gen Sect. (99), 1-6, 1995. [6]. Strauss
et al., Am J Psychiatry (154), 516-22, 1997. [7]. Henry et al., Am J Psychiatry (157), 1506-8, 2000.
[8]. Bolo et al., Neuropsychopharmacology (23), 428-38, 2000. [9]. Strauss
et al., Biol Psychiatry (49), 798-802, 2001. [10]. Henry et al., Neuropsychopharmacology (30),
1076-83, 2005. [11]. Henry et al., MRM (42), 636-42, 1999. [12]. Lange
et al., JMRI (33), 748-54, 2011. [13]. Helms et al., MRM (46), 395-400,
2001. [14]. Ernst et al., MRM (65),
13-7, 2011.