Chris Hanstock1 and Christian Beaulieu1
1University of Alberta, Edmonton, AB, Canada
Synopsis
Gradients used for diffusion spectroscopy
require minimization of the eddy currents. Eddy currents impact phase alignment, spectral
distortion, and signal amplitude, contributing to error in estimating diffusion
coefficients. Strategies have been employed to minimize their effects, including
using bipolar gradient pulses around the MRS volume selective pulses. This
concept has been further refined to allow multiple eddy current time constants
to be minimized by using asymmetric bipolar gradient pairs. An additional
oscillatory eddy current may arise as the result of vibration and acoustic
noise. We present a simple method for assessing and minimizing the oscillatory eddy
currents.
Introduction
The use of strong diffusion sensitizing
gradient fields for spectroscopic studies of metabolite diffusion require
strategies to minimize the effects of induced eddy currents. These effects
impact the accuracy of phase alignment of target peaks, cause spectra
distortion, and therefore affect the resulting measured signal amplitude, which
in turn increases the potential error in the calculated values for diffusion
coefficients. A number of strategies have been employed to minimize the eddy
current effects, which include the use of bipolar gradient pulses around each
of the latter two MRS volume selective pulses (1-3). This concept has been
further refined to allow multiple eddy current time constants to be minimized
by using asymmetric bipolar gradient pairs (4,5). An additional oscillatory
eddy current may arise as the result of vibration and acoustic noise, which
have long time constants in the order of several seconds (6,7).
Here we present a simple method for
assessing and subsequently minimizing the oscillatory eddy currents present
when using typical diffusion gradient schemes for diffusion tensor spectroscopy
(DTS).Methods
All data were acquired using a 4.7T
magnet, a Varian/Agilent spectrometer, and a single channel transmit/receive
head coil. A dual concentric sphere containing n-Octanol in the inner chamber,
and 50mM saline in the outer chamber was used for these studies. A PRESS
sequence was used with the following modifications: (i) A gradient pre-pulse
array for eddy current assessment. (ii) Diffusion gradient pairs around each of
the PRESS 180 pulses.
Eddy current assessment used the PRESS
scheme we described previously (8), with 8 averages acquired for each summed spectrum, and where the time interval following a gradient pre-pulse of 5 Gauss/cm were adjusted from
10ms up to 2s. All 3 gradient directions were assessed with single gradient
pulse (unipolar), two 30ms gradient pulses with either the same phase or opposite
phase, or 15ms gradient pulse lengths for the 4 gradient pulse schemes. The timing between the leading edge of the gradient pairs was adjusted to be consistent with
diffusion sensitization gradient timings in the PRESS portion of the sequence.
All spectra were phase aligned, using
the n-Octanol methylene peak, and the mean and SD of the phase correction
required for each set of 8 averages computed. Once each set of 8 averages was
summed the methylene peak amplitude was measured. For DTS experiments all pre-pulse amplitudes set to zero.Results
The effect on the signal phase following a single X, Y or Z gradient pre-pulse is illustrated in
Figure 1. The effect of the Z-gradient shows a simple single exponential decay
of the eddy current, with a time constant of ~45ms. While the X and Y gradients
show a very long-lasting sinusoidal decay, and where the period of oscillation
is ~150ms. Figure 2 illustrates the effect of applying none 2(a), one 2(b) or two
30ms gradient pre-pulses separated by a time interval of 45ms. The
gradient pulses were either opposite phase, Figure 2(c), or same phase,
Figure 2(d). Extending the pre-pulses to two pairs either alternating phase, Figure 2(e), or same phase, Figure 2(f), show further improvements in reduced phase error accumulation.Discussion
While the typical schemes to minimize
eddy current effects for DTS applications will be effective for the Z gradient
characteristics observed here, such methodology will not permit minimization of
the oscillatory eddy current effects observed for the X and Y gradients.
Furthermore, using the normal reversed phase bipolar scheme as used in most
studies about each 180 pulse, and with typical sequence timings, results in constructive interference, illustrated
in Figure 2(c), and thereby degrading the quality of the DTS data. On the other
hand, using a same phase bipolar pair, Figure 2(d), with the inter-pulse timing
adjusted to match half the period of the oscillatory eddy current, destructive
interference results and the effect on the signal phase was greatly reduced, and thereby providing a more stable
signal phase and amplitude.
Using a hardware approach compensation
for an oscillatory eddy current has been shown to be effective, and is the
ideal method to resolve this (7). However, implementation of this hardware upgrade is not readily feasible for the
majority of scanners, since it requires the installation non-standard
equipment, and we present here a simple method for achieving reasonable
suppression.
In summary, using same phase pairs of diffusion gradients around each 180
pulses in the PRESS sequence provides an effective and efficient means of suppressing oscillatory eddy currents.
Acknowledgements
The authors acknowledge the funding
support from the Canadian Institute for Heath Research.References
1. Alexander
AL, Tsuruda JS, Parker DL. Magn Reson Med 1997,38,1016-1021.
2. Koch
M, Norris DG. Phys Med Biol 200,45,3821-3832.
3. Steidle
G, Schlick F. Magn Reson Med 2006,55,541-548.
4. Reese
TG, Heid O, Weisskoff RM, Wedeen VJ. Magn Reson Med 2003,49,177-182.
5. Finsterbusch
J. Magn Reson Imag 2010,28,434-440.
6. Ryner
LN, Stroman P, Wessel T, Hoult DI, Saunders JK. ISMRM 1998, p1903.
7. Nixon
TW, McIntyre S, Rothman DL, de Graaf RA. J Magn Reson 2008,192,209-217.
8. Hanstock
C, Cobzas D, Beaulieu C. ISMRM 2017, 3019