Jordan Michael Slagowski1, Yao Ding2, Clifton David Fuller2,3, Caroline Chung2, Mo Kadbi4, Zhifei Wen1, and Jihong Wang1,3
1Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States, 2Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, TX, United States, 3The University of Texas Graduate School of Biomedical Sciences at Houston, Houston, TX, United States, 4MR Therapy, Philips HealthTech, Cleveland, OH, United States
Synopsis
The
integration of a MR scanner with a linear accelerator (MR-Linac) may improve image
guidance in radiation therapy (RT). Spatial fidelity is an important consideration
in RT and should be evaluated for MR-Linac imaging. This work presents a phantom
and software developed in-house to measure geometric distortion across a large MR
imaging field-of-view. A comparative assessment of geometric distortion within
a 1.5T MR scanner integrated with a 7 MV linear accelerator is performed versus
several commercial scanners with field strengths ranging from 1.5T-3.0T. The geometric distortion within the MR-Linac
was comparable or less than that measured with the diagnostic scanners.
Purpose
The
recent realization of a high-field (1.5T) MR scanner integrated with a linear
accelerator (MR-Linac) may improve image-guidance in radiation therapy (RT). The
excellent soft tissue contrast offered by MRI could potentially enable online
or real-time adaptive treatment planning, improve target localization, and facilitate
markerless real-time tumor tracking during treatment. However, geometric
distortion attributed to main magnetic field inhomogeneities and gradient
nonlinearities may reduce the spatial localization accuracy of target and
normal tissue structures. High spatial fidelity over a large imaging field-of-view
(FOV) that encompasses the patient skin surface is paramount for RT treatment
planning. The purpose of this study was to assess the geometric distortion for
an MR-guided linear accelerator (MR-Linac, Elekta AB, Stockholm, Sweden) and
compare to several diagnostic scanners with field strengths ranging from
1.5T-3.0T. Methods
A
geometric distortion phantom (Fig. 1A) was fabricated in house and used to
assess geometric distortion for an MR-Linac and several commercial scanners by different
vendors. The phantom consists of 828 high-contrast fiducials suspended in a
non-MR visible material. The fiducials are spaced 5 mm in each of the x-, y-,
and z- dimensions corresponding to the medial-lateral, anterior-posterior, and cranial-caudal
axes, respectively. The phantom was designed to span a large volume-of-interest
(50 cm lateral, 40 cm anterior-posterior, 40 cm cranial-caudal) in order to
assess distortion that may occur near the skin surface. For each scanner
considered, the phantom was centered near MR iso-center and imaged with 3D T1
and T2 weighted pulse sequences (e.g. Fig. 1B) with the vendor provided 3D
geometric distortion correction applied. Similar receiver bandwidth and spatial
resolution were selected during imaging for each scanner. Fiducials were
segmented from the acquired images using a semi-automatic approach. First,
image processing is performed to improve uniformity across the imaging FOV and
enhance fiducial contrast. Second, fiducials are detected using convolution
based template matching and thresholding. The threshold level was set to 10% of
the maximum value of the filtered image. Finally, spatial coordinates (x, y, z)
corresponding to the grayscale weighted center-of-mass of each fiducial are
determined using connected component analysis. Each image is manually inspected
to ensure fiducials in low-contrast regions, often near the edges of the
imaging volume where distortion is greatest, are detected. A manual refinement
procedure, with automated center-of-mass computation, ensures all fiducials are
properly detected. After segmentation, a rigid registration is performed over a
limited FOV near the iso-center to align the measured fiducial positions with a
reference template that specifies coordinates derived from the phantom
manufacturing specifications (Fig. 1C). Distortion vectors are then computed
that specify the deviation of each measured fiducial position relative to the
reference position in each of the x-, y-, and z-dimensions (Fig. 1D). The mean,
standard deviation, and range of the absolute distortion are reported for the
MR-Linac and several commercial scanners. Distortion vectors superimposed on
heat maps facilitate visualization of distortion within the 3D FOV. Results
The
mean geometric distortion measured from T1 weighted MR-Linac images was 1.0 mm
+/- 0.6 mm for fiducials within a spherical volume of diameter 30 cm defined about
iso-center. Geometric distortion for the MR-Linac was less than or equal to
results obtained with 1.5T (1.4 mm +/- 0.8 mm) and 3.0T (1.0 mm +/- 0.4 mm)
diagnostic scanners (Fig. 2). Mean geometric distortion measured 0.8 mm +/- 0.4
mm (MR-Linac), 1.5 mm +/- 1.4 mm (1.5T scanner), and 0.9 mm +/- 0.4 mm (3.0T
scanner) for T2 weighted imaging. Distortion increased with distance from
iso-center, as expected. Heat maps (Fig. 3) and three-dimensional distortion
maps (Fig. 4) present results throughout the imaging volume. A slight
distortion was observed near the center of the MR-Linac imaging volume,
relative to slices +/- 7.5 cm cranial/caudal. The distortion at the center may
be caused by the 15 cm gap between the passive shim coils of the MR-Linac to
permit radiation beam transmission. Figure 5 demonstrates an example
application of the developed distortion phantom and software. Fusion of the
color-coded distortion maps with anatomic images could provide useful
information during contouring, margin determination, or patient set-up. Conclusion
The
geometric distortion measured for an MR-Linac was comparable or less than that
observed for diagnostic scanners. A novel visualization technique was demonstrated
to display geometric distortion maps relative to anatomic structures. This
display feature could be utilized during contour delineation or margin determination.
Efforts are currently ongoing to quantify the geometric distortion of several
additional scanners at our institution. Acknowledgements
No acknowledgement found.References
No reference found.