Fluid velocity was measured in vitro using the difference between spatial phase responses of array coils in 1993 by Famili, Wright and Porter. Aiming at application to cardiovascular stenotic jets, this abstract re-investigates their method with newer coil arrays and adds a multi-echo approach enabling aortic velocity measurement in normal subjects, with moderate results.
Phantom tests in a straight 15mm OD tube compared steady velocities up to 4.27m/s against phase-contrast. Spoiled gradient-echo (SGE) imaging obtained the PAF along the tube (TE 5.23ms, TR 100ms, FA 20deg, 4 averages). Flow on/off SGE single-echo SPD imaging (10) was obtained, but (see below) was flawed in vivo; only multiecho (ME) imaging is presented. ME SPD scans acquired each velocity at TE=[3.15,7.07,10.99ms] and phase images gave velocity using PAF and TOF for 3 echo pairs, i.e. 3 measurements per scan. Mean tube velocity was taken over the central 70% by diameter.
Four normal subjects were scanned supine with anterior mid-sagittal linear array, adjusting using separate coil images to ensure a pair of coils gave strong magnitude in the left-ventricular outflow tract (LVOT). Phase-contrast cines were the comparison standard.
The PAF was acquired through the valve (Figure 3) using turbo-SGE (16 cycle breathhold TE=3.2ms TR=88.2ms, FA 20deg. 1 average, 8 rawdata lines per cycle). Single-echo SPD was not evaluated in-vivo because it requires diastolic repetition assuming zero flow, further confounded by inplane LVOT movement from systole. Cross-sectional SPD valve slices were acquired using the following three ME sequences. Breathholds (BH) were 16 cycles and free-breathing (FB) scans approximately 2 minutes. To estimate intra-session reproducibility, each method was repeated as shown below, with phase-contrast BH cines for comparison.
1. Multi-echo BH single peak systolic image (timed from PC velocity cine) x 4. TE=[2.81,6.49,10.17ms], TR=47ms, FA 20deg, NEX 1, FOV 300x225mm(FExPE)
2. Multi-echo BH cine x 4. TE=[3.01,6.69,10.37ms], TR=54ms, FA 20deg, NEX 1, FOV 360x214mm(FExPE)
3. Multi-echo FB cine x 2. TE=[3.01,6.69,10.37ms], TR=54ms, FA 20deg, NEX 4, FOV 360x214mm(FExPE)
1. Hegrenaes L, Hatle L. Aortic stenosis in adults. Non-invasive estimation of pressure differences by continuous wave Doppler echocardiography. British Heart Journal 1985;54(4):396-404.
2. Sawaya F, Liff D, Stewart J, et al. Aortic Stenosis: A Contemporary Review. The American journal of the medical sciences 2012;343(6):490–496.
3. Otto, CM. Valvular Aortic Stenosis Disease Severity and Timing of
Intervention. Journal of the
American College of Cardiology, Jan 2006;47(11):2141-2151.
4. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics 2002;3:651-671.
5. Nayak K, Nielsen J, Bernstein MA et al. Cardiovascular magnetic resonance phase contrast imaging. Journal of Cardiovascular Magnetic Resonance 2015;17:71
6. Kilner P, Gatehouse P, Firmin DN. Flow Measurement by Magnetic Resonance: A Unique Asset Worth Optimising. Journal of Cardiovascular Magnetic Resonance 2007;9(4):723–728.
7. Myerson SG. Heart valve disease: investigation by cardiovascular magnetic resonance. Journal of Cardiovascular Magnetic Resonance 2012;14(1):7.
8. O'Brien KR, Cowan BR, et al. MRI phase contrast velocity and flow errors in turbulent stenotic jets. J. Magn. Reson. Imaging, 2008;28:210–8.
9. Doyle M, Kortright E, Anayiotos AS et al. Correction of temporal misregistration artifacts in jet flow by conventional phase-contrast MRI. J Magn Reson Imaging. 2007;25(6):1256-62.
10. Famili N, Wright SM, Porter JR. MR flow measurement using RF phase gradients in receiver coil arrays. IEEE Transactions on Biomedical Engineering 1993;40:1243-9.
11. McVeigh ER, Bronskill MJ, Henkelman RM. Phase and sensitivity of receiver coils in magnetic resonance imaging. Medical physics 1986;13(6):806-14.
12. Lorrain and Corson, Electromagnetic Fields and Waves, 2nd Edition. ISBN 0-7167-0331-9 Section 14.5, 623-6.