Yang Yang1, Christopher M Kramer1,2, and Michael Salerno1,2,3
1Medicine, University of Virginia, Charlottesville, VA, United States, 2Radiology, University of Virginia, Charlottesville, VA, United States, 3Biomedical Engineering, University of Virginia, Charlottesville, VA, United States
Synopsis
First-pass
contrast-enhanced myocardial perfusion imaging is a useful noninvasive tool to
evaluate patients with coronary artery disease, but current techniques are
still limited in spatial-temporal resolution, and ventricular coverage which
reduces the sensitivity to detect perfusion differences between the endocardium
and epicardium and quantify ischemic burden. Outer-volume suppression (OVS) can
achieve good signal suppression around the heart, but may have SAR limitations
at 3T. In this study, we designed a spiral pulse sequence with slice-interleaved
or simultaneous multi-slice (SMS) acquisition without OVS to achieve comparable
high quality ultra-high 1.25mm resolution perfusion imaging. The sequences were
tested in heathy volunteers and demonstrated high image quality.
Introduction
First-pass
contrast-enhanced myocardial perfusion imaging is a useful noninvasive tool to
evaluate patients with coronary artery disease (CAD)1. However, current first-pass perfusion imaging
techniques still have limited spatial resolution (~2 mm) which may reduce the sensitivity
to detect perfusion differences between the endocardium and epicardium. We recently
demonstrated perfusion imaging with an ultra-high spatial resolution of 1.25 mm
by using outer-volume suppression (OVS)2, simultaneous multi-slice (SMS) imaging3, and dual density spiral trajectories at 3T4.
However, the OVS module requires a high B1 field to achieve good
suppression performance which will increase SAR and limit the application of
the technique for adenosine stress perfusion imaging. Thus, the goal of this
study is to further develop these spiral pulse sequences to achieve ultra-high
spatial resolution perfusion imaging at 3T without OVS.
Methods
Three
different pulse sequences were evaluated in this study, including an
interleaved acquisition of two slices per saturation recovery (SR) block with
OVS (Fig 1a) and without OVS (Fig 1b), and SMS with a multi band factor of 2
per SR without OVS (Fig 1c). To further reduce the SAR, an optimized 5
hard-pulse SR was utilized5. Detailed parameters of three sequences are
listed in Table 1. Resting first-pass perfusion using the proposed sequences
were performed with a 0.075 mmol/kg Dotarem bolus, separated by 20 min contrast
washout time, in 5 healthy subjects on a 3T Prisma Siemens scanner. One subject
was imaged using interleaved acquisition with and without OVS, while all other
subjects were imaged by interleaved acquisition without OVS and SMS without
OVS. The images were reconstructed by L1-SPIRiT or SMS-L1-SPIRiT using finite
temporal difference as the sparsity transform6. The image reconstruction was
formulated as the following optimization problem: $$ argmin_x \| \Phi DFx - y \|^2 + \lambda_1 \|(G-I)x\|^2 + \lambda_2\|\Psi x\|_1 $$ Where $$$F$$$ transfers
the data from image domain to k-space domain, $$$D$$$ is the inverse gridding
operator that transfers the Cartesian grid to spiral trajectory, $$$G$$$ is an
image-space SPIRiT operator that represents the k-space self-consistency
convolutions in the image domain, $$$\Psi$$$ is the finite time difference transform
that operates on each individual coil separately to achieve sparsity in the
temporal domain of image time series, $$$\Phi$$$
is the operator to
combine multiple phase modulation slices to single SMS slice for SMS acquisition
or $$$\Phi=I$$$ or interleaved
acquisition, $$$\lambda_1$$$ and
$$$\lambda_2$$$ are parameters that balance the data acquisition consistency with
calibration consistency and sparsity.
An upfront scan without SMS is used to
derive the calibration kernel. The overall image quality was graded by a single
cardiologist based on 5 point scale (5-excellent, 1-poor).Results
Figure
2 shows the direct comparison of whole heart perfusion imaging at a middle time
frame during first pass using interleaved acquisition with OVS (a) and without
OVS (b) from the same healthy subject. The OVS can achieve good signal
suppression outside the heart but requires higher SAR to achieve the necessary B1
field. Good image quality is achieved at 1.25mm spatial resolution with or
without OVS. Reduction of the
acquisition time per interleaf from 5ms to 4ms improved image quality and
reduced drop-out artifacts as compared to our prior study. Figure 3 presents another
example case with the same resolution at a similar time frame during first-pass
of contrast using an interleaved acquisition pulse sequence without OVS. Figure
4 shows perfusion images from the same subject acquired using the SMS MB=2
pulse sequence without OVS. Image
quality scores were 4.6±0.2 and 4.0±0.7 for the interleaved and SMS techniques without
OVS respectively. Either an interleaved
acquisition or SMS strategy can produce high quality perfusion images with
whole heart coverage and ultra-high spatial resolution without requiring the OVS
module, which may improve applicability for stress perfusion imaging.Conclusion
We
demonstrated the successful application of ultra-high resolution spiral
perfusion sequence using interleaved acquisition and SMS techniques either with
or without OVS. High-resolution whole-heart perfusion will potentially serve as
a tool to quantify regional differences in perfusion of the subendo and subepi
myocardium. Further validation will be required in patients undergoing
adenosine stress CMR.Acknowledgements
This
work was supported by NIH K23 HL112910 and R01 HL079110.References
1. Nagel E, Klein C, Paetsch I, Hettwer S, Schnackenburg B, Wegscheider K,
Fleck E. 2003. Magnetic resonance perfusion measurements for the noninvasive
detection of coronary artery disease. Circulation. 108(4):432-437.
2. Yang Y, Zhao L, Chen X, Shaw PW, Gonzalez JA, Epstein FH, Meyer CH,
Kramer CM, Salerno M. 2017. Reduced field of view single-shot spiral perfusion
imaging. Magnet Reson Med.
3. Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. 2016. Simultaneous
multislice (sms) imaging techniques. Magn Reson Med. 75(1):63-81.
4. Yang Y, Van Houten M, Kramer CM, Salerno M. 2017. Ultra-high
spatial resolution spiral myocardial perfusion imaging with whole heart
coverage at 3T. Accept for 2018 SCMR/ISMRM workshop.
5. Chow K, Kellman P, Spottiswoode BS, Nielles-Vallespin S, Arai AE,
Salerno M, Thompson RB. 2015. Saturation pulse design for quantitative
myocardial t1 mapping. Journal of Cardiovascular Magnetic Resonance 17:84.
6. Yang Y, Van Houten M, Norton P, Hagspiel K, Kramer
CM, Mugler JP, Salerno M. Spiral Simultaneous Multi-Slice First-Pass Myocardial
Perfusion Imaging. Proc Intl Soc Mag Reson Med 2017;25:3242