Arnold Julian Vinoj Benjamin1,2, Wajiha Bano1,2, Grant Mair2, Michael Davies1, and Ian Marshall2
1School of Engineering, Institute for Digital Communications, University of Edinburgh, Edinburgh, United Kingdom, 2Centre for Clinical Brain Sciences, University of Edinburgh, Edinburgh, United Kingdom
Synopsis
This study shows the importance of sampling order optimization for the contrast preservation of accelerated prospective 3D MRI leading to the improvement in clinical diagnostic utility of accelerated scans using compressed sensing and parallel imaging reconstructions.
Purpose
The main purpose of this study is to validate
the clinical diagnostic utility of accelerated sampling order optimized
prospective MRI in hospitals by radiological assessment of image quality and
artefacts for reducing the overall scan duration. Sampling order optimization
is important because, in certain clinical sequences like Inversion Recovery
(IR) prepared 3D Gradient Echo (GRE), each readout starts at the effective TE
that has optimum contrast. For each line of ky space, a train of 80 kz samples
(with TR=9.6 ms) is acquired after each inversion pulse (with 500 ms delay). Therefore,
the samples collected at the beginning of the readout have better contrast than
those towards the end. Hence, it is important to ensure that the central
kz-space data which contains most of the contrast information in the image is
acquired towards the beginning of each readout for contrast preservation1. In this study, we compare the
radiological scores of sampling order optimized accelerated prospective 3D
scans with the fully sampled scans to determine whether accelerated scans can
be used for clinical diagnosis.Methods
The
scanning was performed on a 1.5T GE Signa Horizon HDX scanner using the
manufacturer’s 3D IR-prepared GRE sequence on 8 subjects who were recruited
under the healthy volunteer ethics protocol. Informed consent was received from
the subjects prior to the scanning. Sequence parameters were
TR/TE/TI=10/4/500 ms; flip angle = 8°; matrix 192×192×160 slices; isotropic 1.3
mm voxels. Fully sampled and accelerated prospective scans (by using patterns without
sampling order optimization – Fig. 1 and with sampling order optimization –
Fig. 2) were performed on each volunteer and the images were reconstructed
using the compressed sensing-parallel imaging (CS-PI) based NESTA algorithm2. The acquired fully
sampled and subsampled brain datasets were then randomised and given to a neuroradiologist
for assessment of image quality and artefacts. The radiological scoring key is
shown in table 1 in which the brain was divided into 4 separate regions that
were scored (i.e. basal nuclei, brainstem, temporal gyri and precentral gyri)
with 0 being the lowest score (i.e. non-diagnostic) and 4 being the highest
(i.e. excellent quality). Artefact scoring of the image datasets was also done
and it ranged from 0 (i.e. severe artefact) to 3 (i.e. no artefact)3.Results
Fig. 1
shows the fully sampled (1a) and three times accelerated subsampling pattern
(1c) without sampling order optimization along with its corresponding CS-PI
reconstructions (1b and 1d). Fig. 2 shows the sampling order optimized fully
sampled (2a) and subsampled patterns (2b, 2c and 2d) at various acceleration factors (R).
The colourbars in Fig. 1 and Fig. 2 illustrate the time instant at which the
samples are collected during each readout train which starts after an
initial inversion pulse of 500 ms duration. Fig. 3 shows the corresponding
CS-PI reconstructions of the sampling order optimized fully sampled (3a) and
subsampled patterns (3b, 3c and 3d) for a single slice of the 3D brain scan. Fig.
4 shows the mean
radiological scores (S) along with standard error (SE) of the fully sampled
and subsampled datasets in which each column was further subdivided into the
mean radiological scores of the four different brain regions and the mean artefact
scores. The CS-PI reconstructions of accelerated scans (up to R= 3) produced
images that were fully diagnostic. The scan time was reduced from 8:08 minutes
to 2:42 minutes for R= 3.Discussion and Conclusion
Although the fully sampled images routinely had
higher radiological scores compared to the accelerated images, it can be seen
from Fig. 4 that the accelerated scans were still fully diagnostic (S > 2)
except for one or two cases in different brain regions that were partly
non-diagnostic (S < 2). This variability may be attributed to motion during
the scanning and is more likely to improve with more volunteer scans because each
S was closer to being fully diagnostic (i.e. S = 2) than partly non-diagnostic
(i.e. S = 1). These results show that sampling order optimization of
subsampling patterns improved the clinical utility of accelerated scans by
producing clinically usable images up to R = 3. The deep brain grey matter
could also be easily identified in sampling order optimized accelerated scans
shown in Fig. 3 due to contrast preservation while it was not the case when
sampling order optimization was not performed as shown in Fig. 14. Further acceleration (R > 3) produced reconstruction artefacts that
made the images clinically unusable. However, these artefacts could be
potentially supressed by the use of more sophisticated compressed sensing
algorithms based on non-local means5 and a current study is ongoing to validate it.Acknowledgements
The research leading to these results has
received funding from the European Union's H2020 Framework Programme
(H2020-MSCA-ITN-2014) under grant agreement no 642685 MacSeNetReferences
[1] A.
J. V. Benjamin, W. Bano, M. Davies, and I. Marshall, "Sampling Order
Optimization for contrast preservation in accelerated prospective 3D MRI,"
in Proceedings of ESMRMB, 2017, p.
#256.
[2] S. Becker, #233, #244, m. Bobin, E. J.
Cand, and #232, "NESTA: A Fast and Accurate First-Order Method for Sparse
Recovery," SIAM J. Img. Sci., vol.
4, pp. 1-39, 2011.
[3] S. Sirin, S. L. Goericke, B. M. Huening,
A. Stein, S. Kinner, U. Felderhoff-Mueser,
et al., "Evaluation of 100 brain examinations using a 3 Tesla
MR-compatible incubator—safety, handling, and image quality," Neuroradiology, vol. 55, pp. 1241-1249,
2013.
[4] I. Marshall, G. Rilling, Y. H. Tao, C.
Du, S. Varma, D. Job, et al., "Radiological
and quantitative assessment of Compressed Sensing reconstruction of
undersampled 3D brain images," in Proceedings
of ISMRM, 2015, p. #2497.
[5] C. Ulas, P. A. Gomez, J. I. Sperl, C.
Preibisch, M. I. Menzel, A. Haase, et al.,
"A Robust Reconstruction Method for Quantitative Perfusion MRI:
Application to Brain Dynamic Susceptibility Contrast (DSC) Imaging," in Proceedings of ISMRM, 2017, p. #209.