Jacob A Macdonald1, Grant S Roberts1, and Oliver Wieben1,2
1Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Radiology, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
MRI
during exercise stress can be a powerful tool in discerning abnormal cardiac
behavior not apparent at rest. As a result of increased cardiac and respiratory
motion, robust gating is essential for high-quality acquisitions during
exercise. Due to increased patient motion, however, missed ECG triggers are
more likely during exercise than at rest. For reconstructions with
retrospective gating, such missed triggers can result in data attributed to the
wrong portion of the cardiac cycle. In this work, we present an algorithm to
identify and correct missed ECG triggers, allowing for exercise scans otherwise
compromised by poor gating to be salvaged.
Introduction
A
major challenge for cardiovascular MRI during exercise remains proper
synchronization of the acquisition and reconstruction with the cardiac cycle.
ECG gating is increasingly prone to missed R-wave triggers during exercise,
likely due to subject motion, motion of the ECG electrodes, and decreased
electrode contact from sweat. In this work, we characterize ECG gating
performance during lengthy 4D flow acquisitions at rest and exercise and introduce
an algorithm for retrospective gating to correct for missed ECG triggers
without the need for prolonging scan time.Methods
Nine
healthy controls (26±1 years; 6 male, 3 female) were imaged on a 3.0T scanner
(Discovery 750, GE Healthcare). 4D flow imaging of the chest was performed at
rest and during exercise with a radially undersampled trajectory (PC VIPR1,2;
TR/TE=6.2/2.0ms; FA=10°; VENC=200cm/s; FOV=32x32x32cm; spatial
resolution=1.25mm isotropic; scan time=9.25 min). Bellow position and R-wave
trigger locations as detected by 4-lead vector ECG were recorded for retrospective
respiratory and ECG gating in a custom offline reconstruction. Supine exercise
was conducted at 70% of each subject’s VO2,max in the magnet bore
with an MR-compatible stepper (Ergospect GmbH).
Exercise imaging began when each subject had achieved a steady-state
heart rate. ECG gating for scans at rest and during exercise were analyzed in
MATLAB. A custom algorithm was used to count the number of missed ECG triggers
by analyzing the time intervals between successive trigger points and estimate
the locations of missed ECG triggers for an improved reconstruction. Missed
trigger locations were estimated by dividing artificially extended intervals
from missed heart beats into multiple intervals on the order of a moving median
RR interval. Images were reconstructed with the original and corrected gating information.
A custom tool3 was used to measure flow waveforms in the aorta pre-
and post-gating correction. Following gating corrections, composite Poincaré
plots4 were created from normalized RR measurements for the cases at
rest and stress to characterize short-term and long-term heart rate variability.Results
Figure
1 summarizes the missed heart beats across all subjects. At rest, an average of
2% of R-wave triggers were missed. During exercise, gating performance degraded
with an average 7% of triggers missed. Figure 2A shows a representative plot of
each recorded RR interval length during exercise. While a clear baseline RR
interval could be noted at approximately 500 ms, frequent RR intervals approximately
integer multiples of the baseline RR interval were observed where one or more
consecutive ECG triggers had been missed. Figure 2B shows effective correction
of the RR interval lengths for the same subject after the ECG correction
algorithm was applied. Figure 3 shows the effect the ECG correction had on total
flow and peak flow measurements. On average, total flow changed 29±42%
(max:121%) following correction, while peak flow only varied 4±2% (max:8%). These
changes can be visualized on a representative corrected flow waveform in Figure
4. Figure 5 shows the composite Poincaré plots generated at rest and stress.
Similar long-term heart rate variability was observed (σ(x2,rest)=0.095,
σ(x2,stress)=0.105), while exercise acquisitions showed noticeably higher
short-term variability (σ(x1,rest)=0.054, σ(x1,stress)=0.076).Discussion
Missed
ECG triggers were only a minor concern at rest, as most subjects had none
missed or few enough to not noticeably affect flow waveforms when acquired over
a 10 min scan. However, during exercise, it was not uncommon for 10-15% of
triggers to be missed, which caused significant inaccuracies in flow waveforms
and measurements. The simple ECG trigger correction algorithm proved to be effective
for corrections. Minimal changes in peak flow measurements following correction
suggested preservation of the systolic flow curve, while larger reductions in
total flow corresponded to the elimination of false flow measurements during
diastole. This algorithm allows for original image SNR to be preserved during
exercise imaging, as no recorded data has to be discarded. With trigger
corrections in place, the Poincaré plots showed similar long-term variations in
heart rate at both rest and stress, suggesting that a steady-state exercise heart
rate was achieved for the duration of the scan. Future work will seek to
improve temporal binning of corrected data by accounting for variable relative
systole and diastole lengths as a function of RR interval length.Conclusion
In
this study, we characterized ECG performance at rest and stress and implemented
an algorithm to retrospectively correct for missed ECG triggers. Missed
triggers were less common at rest, but a noticeable problem during exercise.
The algorithm we implemented proved to be effective at correcting the ECG
gating, allowing for better flow characterization without the need to discard acquired
data associated with missed triggers or prolong scan time.Acknowledgements
We gratefully acknowledge research support from GE Healthcare.References
1.
Gu T, Korosec FR, Block WF, Fain SB, et al. PC VIPR: a high-speed 3D
phase-contrast method for flow quantification and high-resolution angiography.
AJNR. 2005; 26(4): 743-749.
2.
Johnson KM, Lum DP, Turski PA, Block WF, et al. Improved 3D phase contrast MRI
with off-resonance corrected dual echo VIPR. MRM. 2008;60(6):1329-1336.
3.
Stalder AF, Russe MF, Frydrychowicz A, Bock J, et al. Quantitative 2D and 3D
phase contrast MRI: Optimized analysis of blood flow and vessel wall
parameters. MRM. 2008; 60(5): 1218-1231.
4.
Singh B, Singh M. ECG artifacts and Poincare plot based heart rate variability.
IJLTET. 2015; 5(4).