Edward DiBella1, Jason Mendes1, Mark Ibrahim2, Ye Tian3, Brent Wilson2, and Ganesh Adluru1
1Radiology and Imaging Sciences, University of Utah, Salt Lake City, UT, United States, 2Cardiology, University of Utah, Salt Lake City, UT, United States, 3Physics, University of Utah, Salt Lake City, UT, United States
Synopsis
We
propose a unique perfusion acquisition that offers improved coverage and confidence
of detecting true ischemia and artifacts in cardiac perfusion dynamic
acquisitions. Three slices are acquired simultaneously after each saturation
pulse, and there is time to acquire 3 sets of such slices even at high
heartrates. The ability to
simultaneously acquire multiple slices opens up many new possibilities. The
approach proposed here can acquire for example 6 short axis slices and 3 long
axis slices each heartbeat, which allows detection of small areas of ischemia
and can provide additional volume coverage and confidence. Preliminary results
show the promise of this multi-plane SMS approach.
Introduction
Dynamic
contrast enhanced (DCE) MRI of the heart for detecting and characterizing
coronary artery disease is becoming a more widespread tool. In such studies, stress
and rest acquisitions are performed. The first pass of gadolinium contrast is
tracked by imaging a number of slices sequentially each heartbeat. The number
of slices is limited by the time in one heartbeat. Typically 2-4 short axis
slices are acquired per beat, depending on heartrate.
New simultaneous multi-slice (SMS) methods with controlled
aliasing (CAIPI) allow for greater coverage. Such methods have been used with
Cartesian
1 and radial
2-4 readouts. The additional noise or leakage from
this approach is small when multi-band (MB) factors of 2-3 are used. Here we
propose to acquire 3 sets or “blocks” of SMS slices, each with 3 simultaneous slices.
Thus 9 slices are acquired per beat, at heartrates up to 120 beats per minute.
The 9 slices can be acquired all short axis, or, it may be preferable to
acquire one of the sets of slices as long axis slices, which could possibly
increase the coverage or the confidence of ischemia that is found in
intersecting regions that are common to the short and long axis views. Others have acquired a mixture of short and
long axis slices previously, by reducing the number of short axis slices or by
alternating the slice orientation between even and odd heartbeats (reduced
temporal resolution). Including SMS allows new and beneficial trade-offs for
myocardial perfusion coverage, which we formulate here.
Methods
A
saturation recovery radial turboFLASH sequence was used in 4 subjects to study
the potential of the multi-plane SMS cardiac perfusion imaging. TR/TE=2.7/1.6msec,
FOV=260mm, ~1.8x1.8x8mm pixel size on a 3T Prisma (Siemens) scanner. Three
slices were acquired simultaneously with the radial CAIPI SMS method5, 6. 30 rays (golden
ratio ordering) were acquired after each saturation pulse, with a delay of
~80msec.
With
this delay the total acquisition time for 9 slices was < 500msec, so
heartrates up to 120 could acquire the full 9 slices every beat. Scans were
done at regadenoson stress, and ~20min. later at rest, both with ~0.75mmol/kg
gadoteridol. Free-breathing scans were acquired.
Image
reconstruction: For each set of the
simultaneously acquired slices, the 3 slices were reconstructed jointly with an
iterative compressed sensing SMS method. The reconstruction included the
sensitivity maps created from each coil by summing all of the time frames and
using adaptive combine7. Temporal and
spatial total variation constraints were included.
In
addition, in one subject single slices (SMS with multi-band (MB)=1) were
acquired separately, ~10 minutes after contrast was administered. The SMS with MB=3
sequence was also repeated at that time, in order to compare the image quality
directly. Results/Discussion
Fig. 1 shows the multi-planar SMS acquisition
timings. In two subjects, 9 short axis slices were acquired rather than 6 short
axis and 3 long axis slices. Two chamber long axis views were used since with
free-breathing these slices typically have much less out-of-plane motion than
four chamber slices.
The perfusion acquisition is very
similar to standard methods, but more slices are acquired, essentially for
free.
Fig. 2 shows an example from a
subject that had X-ray angiography a year previous (total LAD occlusion,
partial right coronary artery occlusion, bypasses of both vessels were done). Fig. 3, subject had distal occlusions of a
diagonal and apical LAD.
SMS with 3 slices is compared to
single slice acquisitions in Fig. 4. SMS shows essentially no visible degradation. Four simultaneous slices is possible but
would likely work better with breath-holding, and can be more dependent on
coils and slice orientations.
The temporal resolution of the SMS
readouts is better than 3D acquisitions but as in 3D, the slices in each set are
acquired simultaneously so have similar respiratory motion and cardiac cycle.
The
studies here show that the multi-plane SMS approach has promise for characterizing
coronary artery disease and
has unique advantages. More studies are needed to further evaluate the
technique.Acknowledgements
This research was supported by the National Heart, Lung, And Blood Institute of the
National Institutes of Health under Award Number R01HL138082. The content is solely the
responsibility of the authors and does not necessarily represent the official views of the National
Institutes of Health. References
1. Stab D, Wech T, Breuer FA, Weng AM,
Ritter CO, Hahn D, Kostler H. High resolution myocardial first-pass perfusion
imaging with extended anatomic coverage. J
Magn Reson Imaging. 2014;39:1575-1587
2. Adluru
G, Chen L, DiBella EVR. Radial CAIPI-CS for simultaneous multi-slice cardiac
perfusion imaging. Proc Int Soc Magn
Reson. 2014:1532
3. Wang
H, Adluru G, Chen L, Kholmovski EG, Bangerter NK, DiBella EV. Radial
simultaneous multi-slice CAIPI for ungated myocardial perfusion. Magn Reson Imaging. 2016;34:1329-1336
4. Rosenzweig
S, Holme HCM, Wilke RN, Voit D, Frahm J, Uecker M. Simultaneous multi-slice MRI
using cartesian and radial FLASH and regularized nonlinear inversion:
SMS-NLINV. Magn Reson Med. 2017
5. Tian
Y, Adluru G, Mendes J, DiBella E. Optimized 2D radial CAIPIRINHA for cardiac
perfusion MRI. ISMRM. 2017;44
6. Chen
W, Gillett E, Khoo MCK, Davidson Ward SL, Nayak KS. Real-time multislice MRI
during continuous positive airway pressure reveals upper airway response to
pressure change. J Magn Reson Imaging.
2017;46:1400-1408
7. Walsh
DO, Gmitro AF, Marcellin MW. Adaptive reconstruction of phased array MR imagery.
Magn Reson Med. 2000;43:682-690