Thijs Schoots1, Berit Wassenaar1, Anita Kuiper2, Hareld Kemps1, Jeroen Jeneson2, and Remco Renken2
1Flow, Maxima Medical Center, Eindhoven, Netherlands, 2Medical University Center Groningen, Groningen, Netherlands
Synopsis
Patients with chronic heart
failure suffer from diminished leg blood flow (LBF). Question remains to what
extent the distribution or the cardiac output (CO) is responsible. This study
investigates whether CO and LBF could be measured reliably using phase contrast
MRI during supine exercise. 10 healthy subjects performed a supine exercise
test in the MRI at two days at different exercise intensities. Comparison
between both days showed promising reproducibility of measuring CO and LBF
during supine cycling in the MRI although LBF measurements proved more
challenging.
Introduction
Patients with chronic heart
failure suffer from exercise intolerance. Exercise performance is (partly) limited
because of reduced leg blood flow (LBF)1-5. LBF is depended on
cardiac output (CO) and the distribution factor (DF), that is: the fraction of
the CO that is distributed to the legs. Question remains to what extent the DF
or CO is causing the reduced LBF. Current measurements were set up to
investigate whether CO (blood flow in ascending aorta) and LBF (blood flow in
the abdominal aorta, distal from the arteries to the upper body and vital
organs) could be measured reliably using phase contrast MRI (PC-MRI) while
subjects performed supine cycling exercise inside an MR environment. Methods
10 healthy subjects performed a similar exercise protocol (figure 1) in
the MRI (Philips Achieva 3T, Eindhoven, The Netherlands) on a supine cycling
ergometer (Lode, Groningen, The Netherland) at two days, with a minimal resting
period of one day. Prior, subject performed a maximal cardio-pulmonary exercise
test (CPET) on a stationary bike (Lode, Groningen, The Netherlands). During
this CPET the workload at anaerobic threshold (PAT) was assessed to
determine intensities for exercise in the MRI. Exercise protocol is depicted in
figure 2. After a measurement at rest, three exercise intensities where
performed at respectively 30%, 50%, and 70% of PAT, with 1 minute
rest in between. Measurement of blood flow was performed after cycling three
minutes at each intensity, to ensure steady state exercise. Cardiac triggering for both CO and LBF was performed
using a peripheral pulse unit (PPU). A breathing navigator was used to ensure
measurement of blood flow at the same level in the ascending aorta. Velocity
encoding was adjusted for every exercise intensity and entity that was measured
(see table 1 for the MRI settings). Data acquisition was stored in DICOM-format
for offline analysis using matlab (Mathworks, Matlab R2014a). The differences
in blood flow between both days per exercise intensity are expressed in
percentages. Reproducibility in blood flow where considered good when
difference was <10%, reasonable when between 10-20%, moderate when between
20-30% and bad when >30%.Results
All 10 subjects completed
both the exercise tests (see table 2 for basic characteristics). However, due
to technical issues (e.g. PPU triggering) at heart rates of 30%, 50%, and 70% PAT,
measurements of respectively 1, 4, and 7 subjects was not achieved. Therefore
only the results of the measurements during rest and 30% PAT is
presented. Table 3 shows general results. Mean difference of CO during rest and
30% PAT between day 1 and 2 was for both intensities 15%. 5 out of
10 subjects had a good reproducibility, 3 where reasonable, 1 was moderate and
1 had a bad reproducibility. At 30% PAT, 5 are reviewed as good, 3
as reasonable and 2 as bad. Mean difference of LBF during rest and 30% PAT
was respectively 43% and 24%. Of the LBF in rest there was 1 reviewed as good,
2 as reasonable, 3 as moderate and 4 as bad. During LBF
measurement at 30% PAT there was 1 subject rated as good reproducibility, 2 as reasonable,
3 as moderate and 3 as bad. Discussion
The aim of this project was
to investigate reproducibility of measuring LBF and CO during exercise. CO
measurements during rest and 30% PAT had reasonable reproducibility,
especially when taking into account the physiological daily CO fluctuations of
about 15% in rest6. The measurements with a bad reproducibility
where due to movement artifacts. The LBF measurements showed less
reproducibility although it improved at 30% PAT. It is likely that
the daily fluctuation of LBF is larger since it is more sensitive to factors as
physical activity prior to the test. Furthermore, the blood flow in rest is
much less and so more sensitive to absolute measurement errors. This also
explains the improvement of reproducibility during the 30% PAT where
the difference in LBF as compared to CO diminishes. The movement artifacts of
the LBF were bigger as compared to CO measurement, partly due to physical
constraints of the MRI bore.
Future research will aim at
1) reducing motion artifacts by synchronizing data acquisition to peddle
position, 2) improving measurements at higher intensity rates, and 3) focus on
dynamic measurements when heart rate hasn’t reached steady state yet.
Conclusion
This study shows promising
reproducibility of measuring CO and LBF during supine cycling in the MRI
although LBF measurements proved more challenging. Future improvements are
anticipated by using wider MRI bore and improved cardiac triggering which
should eventually lead to medical applications. Acknowledgements
References
1. Esposito,
F., Mathieu-Costello, O., Shabetai, R., Wagner, P. D. & Richardson, R. S.
Limited maximal exercise capacity in patients with chronic heart failure:
partitioning the contributors. J. Am. Coll. Cardiol. 55, 1945–54
(2010).
2. Kemps,
H. M. C., Prompers, JJ., Wessels, B., De Vries, W.R., Zonderland, M.L., Thijssen, E.J.M., Nicolay, K., Schep, G., & Doevendans, P.A.F.M. Skeletal muscle metabolic recovery following submaximal
exercise in chronic heart failure is limited more by O 2 delivery than O 2
utilization. Clin. Sci. 118, 203–210 (2010).
3. Miyazaki,
A., Adachi H., Oshima, S., Taniguchi, K., Hasegawa, A. & Kurabayashi, M. Blood flow redistribution during exercise contributes to
exercise tolerance in patients with chronic heart failure. Circ. J. 71,
465–470 (2007).
4. Poole,
D. C., Hirai, D. M., Copp, S. W. & Musch, T. I. Muscle oxygen transport and
utilization in heart failure: implications for exercise (in)tolerance. Am.
J. Physiol. Heart Circ. Physiol. 302, H1050–63 (2012).
5. Sullivan,
M. J., Knight, J. D., Higginbotham, M. B. & Cobb, F. R. Relation between
central and peripheral hemodynamics during exercise in patients with chronic
heart failure. Muscle blood flow is reduced with maintenance of arterial
perfusion pressure. Circulation 80, 769–781 (1989).
6.
Grollman, A. Variation in the cardiac output of man VIII, the constancy of the
Cardiac Output form Day to Day throughout the Year. Journal of Clinical Investigation, vol. 8, pp. 25-32 (1929).